Improved reconstruction stability for chemical shift encoded hyperpolarized 13 C magnetic resonance spectroscopic imaging using k-t spiral acquisitions.

Erin B Macdonald1, Gregory P Barton1, Benjamin L Cox1,2,3, Kevin M Johnson1,4, Roberta M Strigel1,4,5, Sean B Fain1,4,6.   

Abstract

PURPOSE: A multiecho, field of view (FOV)-oversampled k-t spiral acquisition and direct iterative decomposition of water and fat with echo asymmetry and least-squares estimation reconstruction is demonstrated to improve the stability of hyperpolarized 13 C magnetic resonance spectroscopic imaging (MRSI) in the presence of signal ambiguities attributed to low-SNR (signal-to-noise-ratio) species, local uncertainties in metabolite peaks, and echo-to-echo signal inconsistencies. THEORY: k-t spiral acquisitions redistribute readout points to be more densely spaced radially in k-space by acquiring an FOV and matrix that are oversampled by η. These more densely spaced spiral turns constitute effective intraspiral echoes and can supplement conventional interspiral echoes to improve spectral separation and reduce spectral cross-talk to better resolve 13 C-labeled species for spectroscopic imaging.
METHODS: Digital simulations and imaging phantom experiments were performed for a range of interspiral echo spacings and η using multiecho, k-t spiral acquisitions. Image spectral cross-talk artifacts were evaluated both qualitatively and quantitatively as the percent error in measured metabolite ratios. In vivo murine experiments evaluated the feasibility of multiecho, k-t spiral [1-13 C]pyruvate MRSI to reduce spectral cross-talk for 3 scenarios of different expected reconstruction stability.
RESULTS: Digital simulations and imaging phantom experiments both demonstrated reduced or comparable image spectral cross-talk and percent errors in measured metabolite ratios with increasing η and better choices of echo spacings. In vivo images displayed markedly reduced spectral cross-talk in lactate images acquired with η = 7 versus η = 1.
CONCLUSION: The precision of hyperpolarized 13 C metabolic imaging and quantification in the presence of low-SNR species, local uncertainties in metabolite resonances, and echo-to-echo signal inconsistencies can be improved with the use of FOV-oversampled k-t spiral acquisitions.
© 2019 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  MRSI; carbon-13; hyperpolarized; k-t spiral; metabolism; off-resonance

Mesh:

Substances:

Year:  2019        PMID: 31814173      PMCID: PMC7083691          DOI: 10.1002/mrm.28122

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  22 in total

1.  Least-squares chemical shift separation for (13)C metabolic imaging.

Authors:  Scott B Reeder; Jean H Brittain; Thomas M Grist; Yi-Fen Yen
Journal:  J Magn Reson Imaging       Date:  2007-10       Impact factor: 4.813

2.  Dynamic nuclear polarization polarizer for sterile use intent.

Authors:  Jan H Ardenkjaer-Larsen; Andrew M Leach; Neil Clarke; John Urbahn; Denise Anderson; Timothy W Skloss
Journal:  NMR Biomed       Date:  2011-03-18       Impact factor: 4.044

3.  Spectral-spatial excitation for rapid imaging of DNP compounds.

Authors:  Angus Z Lau; Albert P Chen; Ralph E Hurd; Charles H Cunningham
Journal:  NMR Biomed       Date:  2011-07-12       Impact factor: 4.044

4.  Dual-Echo EPI sequence for integrated distortion correction in 3D time-resolved hyperpolarized 13 C MRI.

Authors:  Benjamin J Geraghty; Justin Y C Lau; Albert P Chen; Charles H Cunningham
Journal:  Magn Reson Med       Date:  2017-04-10       Impact factor: 4.668

5.  Point spread functions of the T2 decay in k-space trajectories with long echo train.

Authors:  Qin Qin
Journal:  Magn Reson Imaging       Date:  2012-07-18       Impact factor: 2.546

6.  Improved tolerance to off-resonance in spectral-spatial EPI of hyperpolarized [1-13 C]pyruvate and metabolites.

Authors:  Justin Y C Lau; Benjamin J Geraghty; Albert P Chen; Charles H Cunningham
Journal:  Magn Reson Med       Date:  2018-01-29       Impact factor: 4.668

7.  IDEAL spiral CSI for dynamic metabolic MR imaging of hyperpolarized [1-13C]pyruvate.

Authors:  Florian Wiesinger; Eliane Weidl; Marion I Menzel; Martin A Janich; Oleksandr Khegai; Steffen J Glaser; Axel Haase; Markus Schwaiger; Rolf F Schulte
Journal:  Magn Reson Med       Date:  2011-11-29       Impact factor: 4.668

Review 8.  Magnetic resonance imaging with hyperpolarized agents: methods and applications.

Authors:  Erin B Adamson; Kai D Ludwig; David G Mummy; Sean B Fain
Journal:  Phys Med Biol       Date:  2017-04-06       Impact factor: 3.609

9.  Increase in signal-to-noise ratio of > 10,000 times in liquid-state NMR.

Authors:  Jan H Ardenkjaer-Larsen; Björn Fridlund; Andreas Gram; Georg Hansson; Lennart Hansson; Mathilde H Lerche; Rolf Servin; Mikkel Thaning; Klaes Golman
Journal:  Proc Natl Acad Sci U S A       Date:  2003-08-20       Impact factor: 11.205

Review 10.  Understanding the Warburg effect: the metabolic requirements of cell proliferation.

Authors:  Matthew G Vander Heiden; Lewis C Cantley; Craig B Thompson
Journal:  Science       Date:  2009-05-22       Impact factor: 47.728

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.