Predictors of successful ultrasound-guided lead implantation.

Asim S Ahmed1, Jasen L Gilge1, Brad A Clark1, Ankur Shah2, Shiv Bagga1, Mathew S Padanilam1, Eric N Prystowsky1, Sandeep A Joshi1, Girish V Nair1, Parin J Patel1.   

Abstract

BACKGROUND: Technical advances have improved the safety of cardiac implantable electronic device (CIED) insertion, but periprocedural complications persist. Despite ultrasound (US) guidance for vascular access being feasible and exhibiting shorter fluoroscopy times, it is not widely adopted for insertion of CIEDs. Thus, we studied the use of US for CIED insertion to (1) quantify the success rate of venous cannulation, (2) identify predictors of failed cannulation, and (3) quantify the rate of complications using US guidance.
METHODS: We studied 166 consecutive patients who underwent US-guided CIED implantation. Anatomic parameters of the axillary vein were measured. The primary outcome was success (group 1) or failure (group 2) to obtain vascular access utilizing US guidance. Secondary outcomes included pneumothorax and hematoma.
RESULTS: Successful US-guided cannulation occurred in 154 of 166 patients (93%). No patient had a pneumothorax. Hematoma occurred in 1 of 166 patients (0.01%). Group 2 exhibited higher male proportion at 11 of 12 (92%) compared with 94 of 154 (61%) in group 1 (P = .03), increased vein depth at 3.84 versus 2.85 cm (P = .003), more right-sided implants (P = .03), higher weight at 104.6 versus 85.3 kg (P = .017), higher body mass index at 35.6 versus 29.2 kg/m2 (P = .049), and higher body surface area at 2.24 versus 1.99 m2 (P = .013). Other parameters were statistically nonsignificant. In multivariate analysis, vein depth remained significantly associated with failure.
CONCLUSION: Using US guidance for CIED implantation is successful in the vast majority (93%) of patients. Rare cases of unsuccessful cannulation were associated with right-sided implants and increased venous depth.
© 2019 Wiley Periodicals, Inc.

Entities:  

Keywords:  axillary vein; cardiac implantable electronic devices; implantable cardioverter defibrillator; pacemaker; pneumothorax; ultrasound access

Mesh:

Year:  2020        PMID: 31808167     DOI: 10.1111/pace.13855

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  A Practical Guide to Ultrasound-guided Venous Access During Implantation of Pacemakers and Defibrillators.

Authors:  Dingxin Qin; Leon M Ptaszek
Journal:  J Innov Card Rhythm Manag       Date:  2022-02-15

2.  Case report: A rare complication after the implantation of a cardiac implantable electronic device: Contralateral pneumothorax with pneumopericardium and pneumomediastinum.

Authors:  Shao-Wei Lo; Ju-Yi Chen
Journal:  Front Cardiovasc Med       Date:  2022-08-18

3.  Feasibility study for echocardiography-guided lead insertion for permanent cardiac implantable electronic devices.

Authors:  Omolade O Sogade; Rieta N Aben; Harry Eyituoyo; Nkechi C Arinze; Felix O Sogade
Journal:  Pacing Clin Electrophysiol       Date:  2021-07-18       Impact factor: 1.976

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.