Optimized electrical bioimpedance measurements of abdominal wall on a porcine model for the continuous non-invasive assessment of intra-abdominal pressure.

Marcelo David1, Omer Amran2, Aviad Peretz2, Aviad Raviv2, Francisco Pracca3,4.   

Abstract

This work describes the optimization of electrical bioimpedance measurements for indirect intra-abdominal pressure (IAP) assessment. The experimental run was performed on a female Sus scrofa domesticus (domestic pig). Different values of IAP were induced by inflation of the abdominal cavity, using a trocar placed near the umbilicus over the linea alba. The whole experiment was run within 1 h of the subject being sacrificed. The abdominal wall thickness was measured at an IAP of 5 mmHg. An exponential trend linking between the bioimpedance values at 99.8 kHz and the IAP was found. Non-optimized electrode placement presented a strongly reduced sensitivity to IAP changes above 7 mmHg. Upon optimization and placing the electrodes with a separation of about 3.6 times the measured abdominal wall thickness, the sensitivity for high IAP drastically increased, allowing continuous non-invasive monitoring of IAP, confirming the optimization method proposed in this work.

Entities:  

Keywords:  Bioimpedance; Intra-abdominal pressure; Non-invasive

Year:  2019        PMID: 31802321     DOI: 10.1007/s10877-019-00441-6

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  1 in total

1.  Pre-Clinical Validation of A Novel Continuous Intra-Abdominal Pressure Measurement Equipment (SERENNO).

Authors:  Salar Tayebi; Robert Wise; Ali Pourkazemi; Johan Stiens; Manu L N G Malbrain
Journal:  Life (Basel)       Date:  2022-07-30
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.