Twin Robotic X-Ray System for 3D Cone-Beam CT of the Wrist: An Evaluation of Image Quality and Radiation Dose.

Jan-Peter Grunz1, Carsten H Gietzen1, Andreas S Kunz1, Andreas M Weng1, Maike Veyhl-Wichmann2, Süleyman Ergün2, Thorsten A Bley1, Rainer Schmitt1,3, Tobias Gassenmaier1.   

Abstract

OBJECTIVE. The purpose of this study was to assess image quality and radiation dose of a novel twin robotic x-ray system's 3D cone-beam CT (CBCT) function for the depiction of cadaveric wrists. MATERIALS AND METHODS. Sixteen cadaveric wrists were scanned using dedicated low-dose and standard-dose CBCT protocols as well as clinical MDCT for comparison. Three readers assessed overall image quality, noise, and artifacts in bone and soft tissue on 5-point Likert scales. For radiation dose analysis, volume CT dose indexes (CTDIvol) were compared. RESULTS. Overall image quality of most studies was very good or excellent in MDCT (for readers 1, 2, and 3: 100%, 100%, and 88%, respectively), standard-dose CBCT (100%, 100%, and 94%), and low dose CBCT (100%, 94%, and 88%) with two readers favoring standard-dose CBCT over MDCT image quality (readers 1 and 2; p ≤ 0.046). In soft tissue, standard-dose (readers 1, 2, and 3; p ≤ 0.021) and low-dose (all p ≤ 0.001) CBCT images had more noise than MDCT in all cases. Standard-dose (all p ≤ 0.003) and low-dose (all p < 0.001) CBCT images also displayed more artifacts. In osseous tissue, one reader observed more noise (p < 0.001) and artifacts (p = 0.020) for low-dose CBCT than for MDCT, whereas no difference was found between standard-dose CBCT and MDCT. Mean CTDIvol was significantly lower for standard-dose (5.2 ± 0.6 mGy; p < 0.001) and low-dose CBCT (1.8 ± 0.2 mGy; p < 0.001) than for clinical MDCT without automatic dose modulation (15.0 ± 0.0 mGy). CONCLUSION. The tested CBCT function delivers suitable image quality for clinical wrist imaging at significantly lower radiation levels than conventional MDCT. In combination with comfortable positioning options and the ability to perform additional radiographic and fluoroscopic examinations, the twin robotic x-ray system may hold the potential to be a one-stop shop device for trauma-associated wrist imaging.

Entities:  

Keywords:  artifacts; cone-beam CT; patient positioning; radiation dosage; wrist

Mesh:

Year:  2019        PMID: 31799871     DOI: 10.2214/AJR.19.21911

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  2 in total

1.  Generating patient-matched 3D-printed pedicle screw and laminectomy drill guides from Cone Beam CT images: Studies in ovine and porcine cadavers.

Authors:  Andrew Kanawati; Alex Constantinidis; Zoe Williams; Ricky O'Brien; Tess Reynolds
Journal:  Med Phys       Date:  2022-05-06       Impact factor: 4.506

2.  3D cone-beam CT with a twin robotic x-ray system in elbow imaging: comparison of image quality to high-resolution multidetector CT.

Authors:  Jan-Peter Grunz; Andreas Max Weng; Andreas Steven Kunz; Maike Veyhl-Wichmann; Rainer Schmitt; Carsten Herbert Gietzen; Lenhard Pennig; Stefan Herz; Süleyman Ergün; Thorsten Alexander Bley; Tobias Gassenmaier
Journal:  Eur Radiol Exp       Date:  2020-09-08
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.