A daily end-to-end quality assurance workflow for MR-guided online adaptive radiation therapy on MR-Linac.

Xinfeng Chen1, Ergun Ahunbay1, Eric S Paulson1, Guangpei Chen1, X Allen Li1.   

Abstract

PURPOSE: Magnetic Resonance (MR)-guided online adaptive radiation therapy (MRgOART), enabled with MR-Linac, has potential to revolutionize radiation therapy. MRgOART is a complex process. This work is to introduce a comprehensive end-to-end quality assurance (QA) workflow in routine clinic for MRgOART with a high-magnetic-field MR-Linac. MATERIALS AND
METHOD: The major components in MRgOART with a high-magnetic field MR-Linac (Unity, Elekta) include: (1) a patient record and verification (R&V) system (e.g., Mosaiq, Elekta), (2) a treatment session manager, (3) an offline treatment planning system (TPS), (4) an online adaptive TPS, (5) a 1.5T MRI scanner, (6) an 7MV Linac, (7) an MV imaging controller (MVIC), and (8) ArtQA: software for plan data consistency checking and secondary dose calculation. Our end-to-end QA workflow was designed to test the performance and connectivity of all these components by transferring, adapting and delivering a specifically designed five-beam plan on a phantom. Beams 1-4 were designed to check Multi-Leaves Collimator (MLC) position shift based on rigid image registration in TPS, while beam 5 was used to check daily radiation output based on image pixel factor of MV image of the field. The workflow is initiated in the R&V system and followed by acquiring and registering daily MRI of the phantom, checking isocenter shift, performing online adaptive replanning, checking plan integrity and secondary 3D dose calculation, delivering the plan while acquiring MV imaging using MVIC, acquiring real-time images of the phantom, and checking the delivering parameters with ArtQA.
RESULTS: It takes 10 min to finish the entire end-to-end QA workflow. The workflow has detected communication problems, permitted resolution prior to setting up patients for MRgOART. Up to 0.9 mm discrepancies in isocenter shift based on the image registration were detected. ArtQA performed the secondary 3D dose calculation, verified the plan integrity as well as the MR-MV isocenter alignment values in TPS. The MLC shapes of beam 1-4 in all adaptive plans were conformal to the target and agreed with MV images. The variation of daily output was within ±2.0%.
CONCLUSIONS: The comprehensive end-to-end QA workflow can efficiently check the performance and communication between different components in MRgOART and has been successfully implemented for daily clinical practice.
© 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  MR-Linac; adaptive RT; daily end-to-end QA workflow

Year:  2019        PMID: 31799753     DOI: 10.1002/acm2.12786

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


  6 in total

1.  Boosting radiotherapy dose calculation accuracy with deep learning.

Authors:  Yixun Xing; You Zhang; Dan Nguyen; Mu-Han Lin; Weiguo Lu; Steve Jiang
Journal:  J Appl Clin Med Phys       Date:  2020-06-19       Impact factor: 2.102

Review 2.  MR-Guided Radiotherapy for Liver Malignancies.

Authors:  Luca Boldrini; Stefanie Corradini; Cihan Gani; Lauren Henke; Ali Hosni; Angela Romano; Laura Dawson
Journal:  Front Oncol       Date:  2021-04-01       Impact factor: 6.244

3.  Longitudinal assessment of quality assurance measurements in a 1.5T MR-linac: Part I-Linear accelerator.

Authors:  Ergys Subashi; Seng Boh Lim; Xesus Gonzalez; Neelam Tyagi
Journal:  J Appl Clin Med Phys       Date:  2021-09-10       Impact factor: 2.102

4.  Quality assurance for on-table adaptive magnetic resonance guided radiation therapy: A software tool to complement secondary dose calculation and failure modes discovered in clinical routine.

Authors:  Carolin Rippke; Oliver Schrenk; C Katharina Renkamp; Carolin Buchele; Juliane Hörner-Rieber; Jürgen Debus; Markus Alber; Sebastian Klüter
Journal:  J Appl Clin Med Phys       Date:  2022-01-12       Impact factor: 2.102

5.  Integration of an Independent Monitor Unit Check for High-Magnetic-Field MR-Guided Radiation Therapy System.

Authors:  Jie Yang; Pengpeng Zhang; Neelam Tyagi; Paola Godoy Scripes; Ergys Subashi; Jiayi Liang; Dale Lovelock; James Mechalakos; Anyi Li; Seng B Lim
Journal:  Front Oncol       Date:  2022-03-11       Impact factor: 5.738

Review 6.  Medical physics challenges in clinical MR-guided radiotherapy.

Authors:  Christopher Kurz; Giulia Buizza; Guillaume Landry; Florian Kamp; Moritz Rabe; Chiara Paganelli; Guido Baroni; Michael Reiner; Paul J Keall; Cornelis A T van den Berg; Marco Riboldi
Journal:  Radiat Oncol       Date:  2020-05-05       Impact factor: 3.481

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.