Comparative evaluation of operating room terminal cleaning by two methods: Focused multivector ultraviolet (FMUV) versus manual-chemical disinfection.

Donna Armellino1, Kristine Goldstein2, Linti Thomas2, Thomas J Walsh3, Vidmantas Petraitis3.   

Abstract

BACKGROUND: This non-randomized comparative observational study evaluated the performance of a standard manual-chemical disinfection process with an automated process employing focused multivector ultraviolet (FMUV) light technology during operating room (OR) terminal cleaning.
METHODS: An Association of periOperative Registered Nurses terminal cleaning protocol was modified to incorporate the use of automated FMUV technology that allows workers to occupy the room during operation. This modified protocol was compared with a standard manual-chemical cleaning and disinfection protocol. Equipment surfaces were pre-sampled before and after terminal cleaning. A total of 165 objects were sampled in each process using a 5-point multisided sampling method.
RESULTS: The parallel process employing FMUV reduced the active microbial burden by 96.5% from baseline (P < .0001), which was over 2.5 times better than the standard process. The standard terminal manual-chemical disinfection process reduced the active microbial burden on sampled objects by 38.4% from baseline (P < .0001).
CONCLUSIONS: The data demonstrates that the performance of standard manual-chemical disinfection alone is variable in a live clinical setting even under the most ideal conditions. By comparison, automated FMUV treatment incorporated in a parallel process consistently produced thorough and significant reductions of microbial contamination levels on all visibly clean patient care equipment.
Copyright © 2019 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  5-point sampling; Hospital quality; Infection prevention nurse; Light based disinfection; Parallel process; Shadowless delivery

Mesh:

Substances:

Year:  2019        PMID: 31796341     DOI: 10.1016/j.ajic.2019.10.009

Source DB:  PubMed          Journal:  Am J Infect Control        ISSN: 0196-6553            Impact factor:   2.918


  4 in total

1.  Mathematical Modeling for Evaluating Inherent Parameters Affecting UVC Decontamination of Indicator Bacteria.

Authors:  Phetcharat Jaiaue; Jirabhorn Piluk; Kanokon Sawattrakool; Jesnipit Thammakes; Chacriya Malasuk; Sitanan Thitiprasert; Nuttha Thongchul; Siwaruk Siwamogsatham
Journal:  Appl Environ Microbiol       Date:  2022-03-15       Impact factor: 5.005

Review 2.  Hospital surface disinfection using ultraviolet germicidal irradiation technology: A review.

Authors:  Robert Scott; Lovleen Tina Joshi; Conor McGinn
Journal:  Healthc Technol Lett       Date:  2022-05-28

3.  Focused multivector ultraviolet (FMUV) technology rapidly eradicates SARS-CoV-2 in-vitro: Implications for hospital disinfection of COVID-19 environments.

Authors:  Steven Park; David S Perlin; Sean Fitzgerald; Vidmantas Petraitis; Thomas J Walsh
Journal:  Am J Infect Control       Date:  2022-02-11       Impact factor: 4.303

4.  Patients With COVID-19 Undergoing Cesarean Deliveries: Adapting the OR Suite and Perioperative Care to Prevent Transmission.

Authors:  Kang Zou; Hong Chen; Yang Liu
Journal:  AORN J       Date:  2020-09       Impact factor: 0.676

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.