Literature DB >> 31793767

Long-Term Accurate Continuous Glucose Biosensors via Extended Nitric Oxide Release.

Maggie J Malone-Povolny, Elizabeth P Merricks, Lauren E Wimsey, Timothy C Nichols, Mark H Schoenfisch.   

Abstract

Analytical performance and tissue interactions of nitric oxide (NO)-releasing continuous glucose sensors were evaluated over a 28 d study in a diabetic swine model. Interstitial glucose was detected using an implanted needle-type amperometric glucose sensor. Two NO-release durations from the sensor surface were achieved by doping the membranes with nonporous (14 d release) or porous (30 d release) S-nitrosothiol-functionalized silica nanoparticles. Numerical and clinical accuracy of the sensors were assessed at time points (1, 7, 14, 21, and 28 d) following implantation. Nitric oxide-releasing sensors demonstrated accurate glucose detection over a time period directly correlated with the active release of NO. Silica particle-doped sensors that released NO for 30 d showed standard-compliant accuracy (i.e., mean absolute relative difference ≤ 15%) for >3 weeks post-implantation. Histological staining for inflammatory biomarkers suggested that the observed performance improvement was the result of decreased inflammatory cell count and a lower density collagen capsule.

Entities:  

Keywords:  continuous glucose monitor; diabetes; foreign body response; nitric oxide; swine model

Mesh:

Substances:

Year:  2019        PMID: 31793767     DOI: 10.1021/acssensors.9b01779

Source DB:  PubMed          Journal:  ACS Sens        ISSN: 2379-3694            Impact factor:   7.711


  1 in total

1.  Mechanisms of Foreign Body Response Mitigation by Nitric Oxide Release.

Authors:  James B Taylor; Maggie J Malone-Povolny; Elizabeth P Merricks; Lauren E Wimsey; Daniel Soliman; Timothy C Nichols; Shannon M Wallet; Robert Maile; Mark H Schoenfisch
Journal:  Int J Mol Sci       Date:  2022-10-01       Impact factor: 6.208

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.