Susanne R Kerscher1, Louise L Schweizer1, Thomas Nägele2, Andrea Bevot3, Michael Alber3, Martin U Schuhmann1. 1. Department of Neurosurgery, Division of Pediatric Neurosurgery, University Hospital of Tuebingen, Tuebingen, Germany. 2. Department of Neuroradiology, University Hospital of Tuebingen, Tuebingen, Germany. 3. Department of Pediatric Neurology and Developmental Medicine, University Children' s Hospital of Tuebingen, Tuebingen, Germany.
Abstract
INTRODUCTION: In a retrospective magnetic resonance imaging (MRI)-based study, we showed that changes of the third ventricle diameter (TVD) are a reliable mirror of changes of the entire ventricular system. The third ventricle is easily accessible in more than 90% of children and adults using ultrasound (US) via the transtemporal bone-window; thus it can be assessed quickly and free of radiation. In order to use transtemporal US determination of TVD instead of MRI/CT in clinical practice, it is important to know if there is a correlation and bias between both methods, which is addressed in this study. MATERIALS AND METHOD: This prospective study investigates 122 children (newborn-18 years). Diagnoses encompassed hydrocephalus (50%), tumors (14.8%), and other intracranial pathologies (35.2%). US-based TVD was measured via the transtemporal bone-window using a phased array 1 to 4MHz transducer. Results were compared with TVD measured on simultaneously acquired axial T1-weighted axial MRI or computed tomography (CT) scans. RESULTS: Overall mean values for TVD were 6.56 ± 5.84 and 6.47 ± 5.64 mm for US and MRI, respectively. There was an outstanding correlation between TVD measured by MRI and US (r = 0.991, p < 0.01). Bland-Altman analysis showed a mean bias of 0.096 mm with limits of agreement of -0.99 and 1.18 mm. CONCLUSION: US- and MRI-based TVD measurements correlate excellently and measure almost identical TVD values. US-based TVD is in mean ∼0.096 mm larger than MRI-based TVD due to a more angulated measurement plane. US is equal to the gold-standard MRI, a fact, opening new avenues for US-based TVD as a first-line assessment tool of ventricular width. Georg Thieme Verlag KG Stuttgart · New York.
INTRODUCTION: In a retrospective magnetic resonance imaging (MRI)-based study, we showed that changes of the third ventricle diameter (TVD) are a reliable mirror of changes of the entire ventricular system. The third ventricle is easily accessible in more than 90% of children and adults using ultrasound (US) via the transtemporal bone-window; thus it can be assessed quickly and free of radiation. In order to use transtemporal US determination of TVD instead of MRI/CT in clinical practice, it is important to know if there is a correlation and bias between both methods, which is addressed in this study. MATERIALS AND METHOD: This prospective study investigates 122 children (newborn-18 years). Diagnoses encompassed hydrocephalus (50%), tumors (14.8%), and other intracranial pathologies (35.2%). US-based TVD was measured via the transtemporal bone-window using a phased array 1 to 4MHz transducer. Results were compared with TVD measured on simultaneously acquired axial T1-weighted axial MRI or computed tomography (CT) scans. RESULTS: Overall mean values for TVD were 6.56 ± 5.84 and 6.47 ± 5.64 mm for US and MRI, respectively. There was an outstanding correlation between TVD measured by MRI and US (r = 0.991, p < 0.01). Bland-Altman analysis showed a mean bias of 0.096 mm with limits of agreement of -0.99 and 1.18 mm. CONCLUSION: US- and MRI-based TVD measurements correlate excellently and measure almost identical TVD values. US-based TVD is in mean ∼0.096 mm larger than MRI-based TVD due to a more angulated measurement plane. US is equal to the gold-standard MRI, a fact, opening new avenues for US-based TVD as a first-line assessment tool of ventricular width. Georg Thieme Verlag KG Stuttgart · New York.