Accuracy of left ventricular ejection fraction determined by automated analysis of handheld echocardiograms: A comparison of experienced and novice examiners.

Omar M Aldaas1, Sachiyo Igata1, Ajit Raisinghani1, Megan Kraushaar1, Anthony N DeMaria1.   

Abstract

BACKGROUND: Handheld ultrasound devices have been developed that facilitate imaging in new clinical settings. However, quantitative assessment has been difficult. Software algorithms have recently been developed with the aim of providing rapid measurements of left ventricular ejection fraction (LVEF) with minimal operator input.
METHODS: We prospectively enrolled a cohort of 70 patients scheduled to undergo echocardiography at the University of California, San Diego. Each patient underwent a standard echocardiography examination by an experienced sonographer as well as a handheld ultrasound with automated software by both the sonographer and an inexperienced resident.
RESULTS: There was a positive correlation between the LVEFs obtained from the standard transthoracic echocardiogram and handheld device in the hands of a novice (r = 0.62; 95% CI 0.45-0.75) and experienced sonographer (r = 0.69; 95% CI 0.54-0.80). The sensitivity and specificity to detect a reduced LVEF (<50%) were 69% and 96% for the novice and 64% and 98% for the experienced sonographer. The sensitivity and specificity to detect a severely reduced LVEF (<35%) were 67% and 97% for the novice and 56% and 93% for the experienced sonographer, but when limited to recordings of at least adequate quality, improved to 100% and 100% for the novice and 100% and 98% for the experienced sonographer, respectively.
CONCLUSION: These data demonstrate that the handheld ultrasound device paired with novel software can provide a clinically useful estimate of LVEF when the images are of adequate quality and yield results by novice examiners that are similar to experienced sonographers.
© 2019 Wiley Periodicals, Inc.

Entities:  

Keywords:  diagnostic imaging tools; echocardiography; hand-carried ultrasound; heart failure; left ventricular ejection fraction

Year:  2019        PMID: 31786824     DOI: 10.1111/echo.14546

Source DB:  PubMed          Journal:  Echocardiography        ISSN: 0742-2822            Impact factor:   1.724


  3 in total

1.  Machine learning algorithm using publicly available echo database for simplified "visual estimation" of left ventricular ejection fraction.

Authors:  Michael Blaivas; Laura Blaivas
Journal:  World J Exp Med       Date:  2022-03-20

Review 2.  Cardiovascular examination using hand-held cardiac ultrasound.

Authors:  Sam Jenkins; Mohamed G Shiha; Eron Yones; James Wardley; Alisdair Ryding; Chris Sawh; Marcus Flather; Paul Morris; Andrew J Swift; Vassilios S Vassiliou; Pankaj Garg
Journal:  J Echocardiogr       Date:  2021-08-02

3.  Diagnostic accuracy of handheld cardiac ultrasound device for assessment of left ventricular structure and function: systematic review and meta-analysis.

Authors:  Sam Jenkins; Samer Alabed; Andrew Swift; Gabriel Marques; Alisdair Ryding; Chris Sawh; James Wardley; Benoy Nalin Shah; Peter Swoboda; Roxy Senior; Robin Nijveldt; Vassilios S Vassiliou; Pankaj Garg
Journal:  Heart       Date:  2021-08-06       Impact factor: 5.994