Abstract
Healthcare-associated infections (HAIs) can be caused by microorganisms present in common practice instruments generating major health problems in the hospital environment. The aim of this work was to evaluate the disinfection capacity of a portable ultraviolet C equipment (UV Sanitizer Corvent® -UVSC-) developed to disinfect different objects. For this purpose, six pathogens causing HAIs: Acinetobacter baumannii, Bacillus subtilis, Escherichia coli, Pseudomonas aeruginosa, Staphylococcus aureus and Candida albicans, were inoculated on slides and discs of different biomaterials (borosilicate, polycarbonate, polyurethane, silicone, Teflon and titanium) and exposed to ultraviolet C radiation. UVSC disinfection was compared with ethanol and chlorhexidine antimicrobial activities following the standards EN14561 and EN14562. Disinfection, established as a reduction of five logarithms from the initial inoculum, was achieved with the UVSC at 120 s of exposure time, with and without the presence of organic matter. The disinfectant effect was observed against S. aureus, P. aeruginosa, E. coli, B. subtilis and C. albicans (reduction >99.999%). Disinfection was also achieved with 70% ethanol and 2% chlorhexidine. As conclusion, UVSC was effective disinfecting the most contaminated surfaces assayed, being a promising alternative for disinfecting hospital materials and inanimate objects that cannot be immersed in liquid biocides, reducing the risk of pathogen transmission.Entities:
Keywords: UVC light disinfection; disinfection; healthcare-associated infections; portable equipment
Mesh:
Substances:
Year: 2019 PMID: 31783593 PMCID: PMC6926820 DOI: 10.3390/ijerph16234747
Source DB: PubMed Journal: Int J Environ Res Public Health ISSN: 1660-4601 Impact factor: 3.390
Figure 1Corvent® UV Sanitizer equipment. (A). Equipment layout. (B). Images of the equipment.
Figure 2Microbial burden values obtained after the exposure to ultraviolet radiation in UVSC (A: clean conditions, B: dirty contidions) and the immersion in 70% ethanol (C: clean conditions, D: dirty contidions) and 2% chlorhexidine (E: clean conditions, F: dirty contidions). The lines that do not have a term indicate values that are inside the excluded range of the y axis and are not close to the reduction of 5 logarithms established by the standard as a disinfection value.
Microbial burden values expressed in CFU/mL obtained after exposure of the six materials to 30 and 120 s to the different disinfection methods, UVSC, 70% ethanol and 2% chlorhexidine, in both clean and dirty conditions. Lower detection limit: ≤0.5.
| Strain | Initial Inoculum (CFU/mL) | Disinfection Conditions | Materials | |||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Borosilicate | Polycarbonate | Polyurethane | Silicone | Teflon | Titanium | |||||
|
| 1.7 × 108 | Clean | UVSC | 30″ | 4.6 × 103 ± 41.7 | ≥ | ≥ | ≥ | 4.2 × 103 ± 141.4 | ≥ |
| 120″ | ≤0.5 | 3.3 × 102 ± 128 | 5.5 × 103 ± 50.3 | 1.6 × 103 ± 136.6 | 6.5 × 103 ± 40 | 3.0 × 101 ± 38.3 | ||||
| Ethanol | 30″ | 4.5 × 101 ± 50 | 0.5 × 101 ± 10 | ≥ | 3.0 × 102 ± 31 | 4.5 × 102 ± 137.9 | 1.4 × 102 ± 16.3 | |||
| 120″ | ≤0.5 | 0.5 × 101 ± 10 | 1.5 × 103 ± 244.9 | 3.0 × 101 ± 11.5 | 7.5 × 101 ± 34.1 | ≤0.5 | ||||
| Chlorhexidine | 30″ | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | |||
| 120″ | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ||||
| Dirty | UVSC | 30″ | 5.4 × 102 ± 80 | 3.5 × 103 ± 200.1 | ≥ | ≥ | ≥ | 5.1 × 103 ± 50.4 | ||
| 120″ | ≤0.5 | 6.0 × 101 ± 95.2 | 5.8 × 103 ± 81.6 | ≥ | 3.8 × 103 ± 141.4 | 3.3 × 102 ± 40.8 | ||||
| Ethanol | 30″ | 4.9 × 103 ± 210.2 | 3.6 × 102 ± 91.4 | 4.9 × 103 ± 206.1 | 4.9 × 103 ± 290.6 | ≥ | 1.3 × 103 ± 200.0 | |||
| 120″ | 0.5 × 101 ± 10 | 0.5 × 101 ± 10 | ≥ | 6.5 × 103 ± 150.2 | 3.2 × 102 ± 162.4 | 1.0 × 101 ± 11.5 | ||||
| Chlorhexidine | 30″ | ≤0.5 | ≤0.5 | 2.5 × 101 ± 50 | ≤0.5 | ≤0.5 | ≤0.5 | |||
| 120″ | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ||||
|
| 1.2 × 108 | Clean | UVSC | 30″ | ≥ | ≥ | ≥ | ≥ | ≥ | ≥ |
| 120″ | 2.8 × 102 ± 78.3 | ≥ | ≥ | 6.0 × 102 ± 66.5 | ≤0.5 | 5.1 × 102 ± 91.4 | ||||
| Ethanol | 30″ | 6.3 × 103 ± 76.7 | ≥ | 3.8 × 103 ± 50.3 | ≥ | ≥ | 3.9 × 103 ± 180.3 | |||
| 120″ | ≤0.5 | ≥ | 1.5 × 101 ± 19.1 | 5.5 × 102 ± 92.9 | 2.6 × 103 ± 91.0 | ≤0.5 | ||||
| Chlorhexidine | 30″ | ≤0.5 | 1 × 101 ± 20 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | |||
| 120″ | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ||||
| Dirty | UVSC | 30″ | ≥ | ≥ | ≥ | ≥ | ≥ | ≥ | ||
| 120″ | 9.5 × 102 ± 17 | ≥ | ≥ | ≥ | 3.7 × 102 ± 26.7 | 3.7 × 102 ± 26.7 | ||||
| Ethanol | 30″ | ≤0.5 | ≥ | ≥ | ≥ | ≥ | ≥ | |||
| 120″ | ≤0.5 | ≥ | 9.8 × 102 ± 42.4 | ≥ | ≥ | ≤0.5 | ||||
| Chlorhexidine | 30″ | ≤0.5 | 0.5 × 101 ± 10 | 4.5 × 101 ± 10 | ≤0.5 | ≤0.5 | ≤0.5 | |||
| 120″ | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ||||
|
| 1.4 × 107 | Clean | UVSC | 30″ | 3.5 × 102 ± 131 | 9.9 × 102 ± 158.7 | ≥ | 4.5 × 103 ± 120.7 | 5.0 × 103 ± 170.0 | 3.1 × 103 ± 140.4 |
| 120″ | ≤0.5 | 1.3 × 102 ± 44.3 | 2.8 × 103 ± 150.5 | 5.0 × 102 ± 48.9 | 2.5 × 101 ± 5.7 | 0.5 × 101 ± 10 | ||||
| Ethanol | 30″ | ≥ | 3.5 × 101 ± 30 | ≤0.5 | 2.8 × 102 ± 50 | 2.8 × 103 ± 228.2 | 7.5 × 101 ± 19.1 | |||
| 120″ | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | 0.5 × 101 ± 10 | ≤0.5 | ||||
| Chlorhexidine | 30″ | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | |||
| 120″ | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ||||
| Dirty | UVSC | 30″ | 2.0 × 101 ± 9.1 | 2.0 × 102 ± 82.2 | ≥ | ≥ | ≥ | 1.3 × 103 ± 98.9 | ||
| 120″ | ≤0.5 | 1.1 × 102 ± 10 | 3.8 × 103 ± 280.5 | 5.7 × 103 ± 90.8 | 1.3 × 102 ± 5 | 7.0 × 101 ± 25.8 | ||||
| Ethanol | 30″ | 4.4 × 103 ± 140.0 | 6.4 × 103 ± 136.6 | ≥ | ≥ | ≥ | 2.6 × 103 ± 160.7 | |||
| 120″ | ≤0.5 | ≤0.5 | ≤0.5 | 3.5 × 101 ± 10 | 4.1 × 102 ± 131.2 | ≤0.5 | ||||
| Chlorhexidine | 30″ | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | |||
| 120″ | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ≤0.5 | ||||
Mean log10 CFU/plate ± SD; ≥: colony counting ≥ 6.6 × 103 as specified in the standard.