Analytical validation of new ELISAs for the quantitation of polyclonal free light chains and comparison to existing assays for healthy and patient samples.

Jennifer L J Heaney1, John P Campbell2, Margaret Goodall3, Tim Plant3, Meena Shemar2, Christopher Hand4, Mark T Drayson3.   

Abstract

BACKGROUND: Polyclonal FLCs can be used as a biomarker of inflammation and immune activation in a range of diseases. This study evaluated the performance of new FLC ELISAs (Seralite FLC ELISA) for the quantitation of polyclonal κ and λ FLC, including comparisons to existing assays.
METHODS: Technical performance was assessed for the ELISA and reference ranges were generated using healthy donor serum (N = 91). Patients with a range of conditions associated with polyclonal FLC dysregulation (N = 164) were measured across platforms.
RESULTS: The ELISAs generated references ranges of: 8.72-23.0 mg/L κ FLC, and 8.52-25.24 mg/L for λ FLC. ELISAs demonstrated linearity across the calibration range and intra-assay (≤ 8.7%) and inter-assay (≤ 12.3%) imprecision was low. The limit of detection was 0.63 mg/L for κ and 0.57 mg/L for λ FLC. Minimal cross-reactivity was observed for interference agents, alternate FLC and whole immunoglobulin (median change ≤3.6 mg/L). Assays showed good batch-to-batch consistency. For patient samples, methods generated different κ and λ FLC concentrations and differences were seen between methods for the number of patients classified as below, with and above references ranges for κ and λ FLC. There was no significant difference in the FLC sum between the different techniques.
CONCLUSIONS: The ELISAs displayed good analytical and technical performance. The quantification of individual κ and λ FLC appears inherently different between platforms. These differences are attenuated if using the FLC sum, which was similar between methods and provided agreement in relation to patients having normal or elevated FLCs.
Copyright © 2019 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Assay; ELISA; Free light chain; Methods; Polyclonal; Quantitation

Mesh:

Substances:

Year:  2019        PMID: 31783021     DOI: 10.1016/j.jim.2019.112713

Source DB:  PubMed          Journal:  J Immunol Methods        ISSN: 0022-1759            Impact factor:   2.303