Performance of Makerless Tracking for Gimbaled Dynamic Tumor Tracking.

Marc Ziegler1, Sebastian Lettmaier1, Rainer Fietkau1, Christoph Bert2.   

Abstract

BACKGROUND AND
PURPOSE: The purpose of this work is to report the workflow and the accuracy of the new markerless dynamic tumor tracking (MLDTT) method of the Vero 4DRT system introduced with ExacTrac 3.6.1.
MATERIAL AND METHODS: Phantom measurements were performed to assess the accuracy of the MLDTT algorithm by using the QA-tool which is provided by the vendor. A patient breathing curve was used as the motion trajectory of the phantom and the target positions detected by the MLDTT algorithm were compared to the defined positions. Furthermore, eight patients have been treated with MLDTT between May 2018 and July 2019. A log-file analysis is used to evaluate MLDTT treatment data.
RESULTS: The accuracy of the MLDTT detection is 0.12mm ± 0.12mm, 0.12mm ± 0.11mm, 0.20mm ± 0.21mm for the x-, y-, z-direction, respectively. These values are comparable to the accuracy of marker based DTT at the Vero system. The median treatment time was 21min 34seconds and 175kV images were acquired during treatment for monitoring the target motion.
CONCLUSION: The accuracy of the MLDTT algorithm is comparable to the marker based approach and the accuracy reported for the XSight Lung of the CyberKnife. Eight patients were treated successfully using MLDTT and the treatment times are comparable to a standard DTT treatment.
Copyright © 2019. Published by Elsevier GmbH.

Entities:  

Keywords:  NSCLC; Real-time monitoring; Stereotactic body radiation therapy; Tumor Tracking

Mesh:

Year:  2019        PMID: 31780095     DOI: 10.1016/j.zemedi.2019.10.003

Source DB:  PubMed          Journal:  Z Med Phys        ISSN: 0939-3889            Impact factor:   4.820


  2 in total

1.  The markerless lung target tracking AAPM Grand Challenge (MATCH) results.

Authors:  Marco Mueller; Per Poulsen; Rune Hansen; Wilko Verbakel; Ross Berbeco; Dianne Ferguson; Shinichiro Mori; Lei Ren; John C Roeske; Lei Wang; Pengpeng Zhang; Paul Keall
Journal:  Med Phys       Date:  2021-12-29       Impact factor: 4.071

2.  Markerless Real-Time 3-Dimensional kV Tracking of Lung Tumors During Free Breathing Stereotactic Radiation Therapy.

Authors:  Kimmie de Bruin; Max Dahele; Hassan Mostafavi; Berend J Slotman; Wilko F A R Verbakel
Journal:  Adv Radiat Oncol       Date:  2021-04-20
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.