Literature DB >> 31778796

Performance of ICDAS II and fluorescence methods on detection of occlusal caries-An ex vivo study.

Grażyna Marczuk-Kolada1, Elzbieta Luczaj-Cepowicz2, Marta Obidzinska2, Janusz Rozycki3.   

Abstract

BACKGROUND: The aim of this ex vivo study was to compare the performance of the visual ICDAS II scale and two fluorescence methods (DIAGNOdent Pen and VistaCam iX) in detecting occlusal caries lesions in permanent teeth using cone beam computed tomography (CBCT) as a reference test.
METHODS: One hundred-sixty molars and premolars were qualified for the study. The visual, the DIAGNOdent Pen and VistaCam iX examinations were carried out by two examiners. The actual extent of the lesions was determined using CBCT. For the three methods, inter- and intra-examiner reproducibility was calculated using intraclass correlation coefficient. The sensitivity, specificity, and accuracy of these methods were compared using the mid-P McNemar test. The level of significance was set at p < 0.05.
RESULTS: For all data, intraclass-correlation coefficients (ICC) for intra- and inter-examiner reproducibility were high. With regard to the enamel threshold, the sensitivity and accuracy were significantly higher for VistaCam iX and ICDAS II compared with the DIAGNOdent Pen (p < 0.05). The results obtained with the DIAGNOdentu Pen indicated significantly higher specificity values for the enamel threshold (p < 0.05). The results obtained with the VistaCam iX showed the lowest sensitivity but the significantly highest specificity and accuracy for the dentin threshold (p < 0.05).
CONCLUSIONS: The ICDAS II is an effective method for detecting early carious lesions, characterized by a higher sensitivity compared with devices using fluorescence. To improve the diagnostic efficiency of the evaluated devices, modifying the cut-off values recommended by the manufacturers should be considered.
Copyright © 2019. Published by Elsevier B.V.

Entities:  

Keywords:  Cone beam computed tomography; Fluorescence methods; ICDAS II; Occlusal caries detection

Mesh:

Substances:

Year:  2019        PMID: 31778796     DOI: 10.1016/j.pdpdt.2019.101609

Source DB:  PubMed          Journal:  Photodiagnosis Photodyn Ther        ISSN: 1572-1000            Impact factor:   3.631


  2 in total

1.  Clinical effectiveness of high definition fluorescence camera in detection of initial occlusal caries.

Authors:  Mohamed Salama; Olfat Hassanein; Omar Shaalan; Asmaa Yassen
Journal:  J Clin Exp Dent       Date:  2022-02-01

Review 2.  Emerging Technologies for Dentin Caries Detection-A Systematic Review and Meta-Analysis.

Authors:  Christa Serban; Diana Lungeanu; Sergiu-David Bota; Claudia C Cotca; Meda Lavinia Negrutiu; Virgil-Florin Duma; Cosmin Sinescu; Emanuela Lidia Craciunescu
Journal:  J Clin Med       Date:  2022-01-28       Impact factor: 4.241

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.