The early failure of the gamma nail and the dynamic hip screw in femurs with a wide medullary canal. A biomechanical study of intertrochanteric fractures.

Marcin Ceynowa1, Krzysztof Zerdzicki2, Pawel Klosowski3, Rafał Pankowski4, Marek Rocławski4, Tomasz Mazurek4.   

Abstract

BACKGROUND: Intertrochanteric fractures may occur in a bone with a wide medullary canal that may lead to significant mobility of a intramedullary nail, contrary to an extramedullary device. This study evaluates the Dynamic Hip Screw and the gamma nail in AO 31.A2.1 fractures in these circumstances.
METHODS: Synthetic femora with canals drilled to 18 mm were used. Five fixation types were examined: a 2 - hole and a 4 - hole Dynamic Hip Screw with a 2 - hole plate, a standard gamma nail with dynamic and static distal locking and a long gamma nail. The specimens were tested with cyclic axial loading, from 500 N increasing of 50 N increments in each cycle. Force at failure, overall stiffness, stiffness at the fracture site, location and mode of failure were recorded.
FINDINGS: The short gamma nails dislocated into varus under preload because the nail migrated laterally. The Dynamic Hip Screw was initially stable, but some specimens rotated around the lag screw. The gamma nail was rotationally stable. Both implants failed through femur fracture. The long gamma nailed failed by screw cut - out at forces lower than the ultimate force of the short gamma nail.
INTERPRETATION: This study shows that the gamma nail is unstable in a large medullary canal but offers better rotational stability of the proximal fragment. A modification of the nail design or the operative technique may be considered.
Copyright © 2019 The Authors. Published by Elsevier Ltd.. All rights reserved.

Entities:  

Keywords:  Biomechanical study; Dynamic hip screw; Gamma nail; Intertrochanteric fracture

Year:  2019        PMID: 31775090     DOI: 10.1016/j.clinbiomech.2019.11.006

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  8 in total

1.  Effect of Fracture Reduction with Different Medial Cortical Support on Stability After Cephalomedullary Nail Fixation of Unstable Pertrochanteric Fractures: A Biomechanical Analysis.

Authors:  Ling Ling; Zhongyong Qu; Kaihua Zhou
Journal:  Indian J Orthop       Date:  2021-06-30       Impact factor: 1.033

2.  The cement-bone bond is weaker than cement-cement bond in cement-in-cement revision arthroplasty. A comparative biomechanical study.

Authors:  Marcin Ceynowa; Krzysztof Zerdzicki; Pawel Klosowski; Maciej Zrodowski; Rafal Pankowski; Marek Roclawski; Tomasz Mazurek
Journal:  PLoS One       Date:  2021-02-11       Impact factor: 3.240

3.  Short cephalomedullary nail toggle: a closer examination.

Authors:  Albert V George; Kamil Bober; Erik B Eller; William M Hakeos; Joseph Hoegler; Ali H Jawad; S Trent Guthrie
Journal:  OTA Int       Date:  2022-01-26

4.  Low filling ratio of the distal nail segment to the medullary canal is a risk factor for loss of anteromedial cortical support: a case control study.

Authors:  Hui Song; Shi-Min Chang; Sun-Jun Hu; Shou-Chao Du
Journal:  J Orthop Surg Res       Date:  2022-01-15       Impact factor: 2.359

5.  Biomechanical study of extramedullary and intramedullary fixation in the treatment of unstable intertrochanteric reversed-tilt fractures of the femur.

Authors:  Guo-Liang Lu; Song-Jun Li; Wen-Xue Li
Journal:  Ann Transl Med       Date:  2022-02

6.  Biomechanical Comparisons of Trochanteric Hip Fracture Fixation Using Short-, Mid-, and Long-Length Proximal Femoral Nails.

Authors:  Tomohiro Matsumura; Tsuneari Takahashi; Ryusuke Ae; Katsushi Takeshita
Journal:  Geriatr Orthop Surg Rehabil       Date:  2022-09-02

7.  Drill holes decrease cancellous bone strength: A comparative study of 33 paired osteoporotic human and 9 paired artificial bone samples.

Authors:  Marcin Ceynowa; Krzysztof Zerdzicki; Pawel Klosowski; Rafal Pankowski; Marek Roclawski; Tomasz Mazurek
Journal:  PLoS One       Date:  2020-10-29       Impact factor: 3.240

8.  Static vs dinamic short nail in pertrochanteric fractures: experience of two center in Northern Italy.

Authors:  Fabrizio Quattrini; Alessandro Ortolani; Corrado Ciatti; Calogero Puma Pagliarello; Antonio Martucci; Luca Gurrieri; Stefano Stallone; Giuseppe Melucci; Tigani Domenico; Pietro Maniscalco
Journal:  Acta Biomed       Date:  2021-07-26
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.