| Literature DB >> 31773299 |
Toshiyuki Komaki1, Takao Hiraki2, Tetsushi Kamegawa3, Takayuki Matsuno4, Jun Sakurai5, Ryutaro Matsuura6, Takuya Yamaguchi7, Takanori Sasaki8, Toshiharu Mitsuhashi5, Soichiro Okamoto1, Mayu Uka1, Yusuke Matsui1, Toshihiro Iguchi1, Hideo Gobara9, Susumu Kanazawa1.
Abstract
OBJECTIVES: To evaluate the accuracy of robotic CT-guided out-of-plane needle insertion in phantom and animal experiments.Entities:
Keywords: Animal experiments; Interventional radiology; Robotics
Year: 2019 PMID: 31773299 PMCID: PMC7033049 DOI: 10.1007/s00330-019-06477-1
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315
Fig. 1The robotic system. A robot (left) with 6 degrees of freedom and an interface (right) comprising a touch panel (long arrow) and a controller (arrowhead). The robot may be manipulated by either button operation of the controller or numerical inputs on software displayed on the touch panel. With the former technique, the robot moves while the buttons of the controller are manually pressed, whereas with the latter technique, the robot moves to a certain place semi-automatically after numerical inputs. In the present study, typically, alteration of needle angles for needle targeting and adjustment of needle orientation was performed by the latter, while needle insertion was done by the former. The needle (short arrow) is attached to a plastic needle holder at the end of a robot arm
Needle Angles for Insertion in the Phantom Experiment
| Needle angles (°) | No. of insertions | |||||
|---|---|---|---|---|---|---|
| Robotic | Manual | |||||
| Smartphone-guided | Freehand | |||||
| XY plane | YZ plane | Operator 1 | Operator 2 | Operator 1 | Operator 2 | |
| 40 | -20 | 2 | 1 | 1 | 1 | 1 |
| 30 | -30 | 2 | 1 | 1 | 1 | 1 |
| 20 | -40 | 2 | 1 | 1 | 1 | 1 |
| -20 | 20 | 2 | 1 | 1 | 1 | 1 |
| -30 | 30 | 2 | 1 | 1 | 1 | 1 |
| -40 | 40 | 2 | 1 | 1 | 1 | 1 |
The needle angle is defined as the angle between the perpendicular line and the needle on CT images. When the needle is oriented to the right and left side of the perpendicular line, the needle angle is expressed as a positive and negative value, respectively
Fig. 2Robotic out-of-plane needle insertion in the phantom experiment. The needle (arrow) at the predetermined angles in the XY and YZ planes is inserted into a sponge phantom (arrowhead) fixed on the CT table
Needle Angles for Insertion in the Animal Experiment
| Needle angles (°) | No. of insertions | ||||
|---|---|---|---|---|---|
| With adjustments | Without adjustments | ||||
| XY plane | YZ plane | Hip muscle | Kidney | Hip muscle | Kidney |
| +30 ± 15 | -30 ± 10 | 2 | 2 | 2 | 2 |
| 0 ± 15 | -30 ± 10 | 2 | 2 | 2 | 2 |
| -30 ± 15 | -30 ± 10 | 2 | 2 | 2 | 2 |
The needle angle is defined as the angle between the perpendicular line and the needle on CT images. When the needle is oriented to the right and left side of the perpendicular line, the needle angle is expressed as a positive and negative value, respectively
Fig. 3Techniques of needle insertion with adjustment in the animal experiment. Needle orientation is checked by CT scanning at two points: middle of the tract and 1 cm behind the target. Needle orientation is then evaluated in maximum intensity projections reconstructed from CT data in the XY and YZ planes. If the needle orientation is not satisfactory, it is adjusted until it becomes satisfactory
Fig. 4Schemata of the needle angle being changed by the robot in the air (a) and in vivo (b). The robot provides the remote-center-of-motion function, by which the needle angle is changed around its tip in the air. In vivo, however, the needle angle is changed as if a pivot point is at the approximately half length of needle in the tissue; this is attributable mainly to resistance of the tissue. Considering this characteristic needle behavior in vivo, the corrected needle angle in the planes may be calculated to compensate for the deviation between the ideal and actual needle angles
Results of the phantom experiment
| Robotic | Manual | |||
|---|---|---|---|---|
| Smartphone-guided | Freehand | |||
| Needle insertion time (s) | 5.0 ± 0.1 (4.8–5.1) | 24.7 ± 8.1 (13.8–43.3) | 4.4 ± 1.8 (3.0–9.4) | < 0.001 |
| Needle insertion accuracy (°) | ||||
| | 0.4 ± 0.4 (0.0–1.6) | 3.7 ± 2.3 (0.0–7.4) | 7.0 ± 5.7 (1.0–17.1) | < 0.001 |
| | 0.6 ± 0.4 (0.0–1.2) | 0.6 ± 0.4 (0.1–1.5) | 6.3 ± 3.5 (1.8–12.1) | 0.65 |
| Predicted needle tip deviation at a depth of 8 cm (mm) | 1.0 ± 0.7 (0.1–2.5) | 4.9 ± 2.9 (0.9–9.7) | 13.0 ± 7.0 (4.5–29.4) | < 0.001 |
Data are means ± standard deviations, with ranges in parentheses
aComparison between robotic and manual smartphone-guided insertions with Student’s t test
Results of the animal experiment
| Robotic insertion | |||
|---|---|---|---|
| With adjustment | Without adjustment | ||
| Needle tract length (mm)a | 82.2 ± 5.3 (71.1–87.5) | 82.5 ± 8.0 (69.9–92.0) | 0.903 |
| No. of needle adjustments during insertionb | 2 (0–5) | 0 (0–0) | < 0.001 |
| No. of CT scans during insertionb | 4 (2–8) | 0 (0–0) | < 0.001 |
| Needle insertion time (s)a | 716.9 ± 396.0 (316–1851) | 14.4 ± 2.6 (11–19) | < 0.001 |
| Distance of target movement (mm)a | 4.6 ± 2.5 (1.2–11.2) | 4.8 ± 0.8 (3.8–6.6) | 0.776 |
| Radiation exposure to swine during insertion | |||
| Tube current-time product (mAs)a | 7210.1 ± 2746.4 (3999–14,533) | 1310.8 ± 77.1 (1066–1333) | < 0.001 |
| Dose-length product (mGy·cm)a | 998.3 ± 380.1 (553.8–2011.7) | 181.5 ± 10.7 (147.7–184.6) | < 0.001 |
| Needle insertion accuracy (mm) | |||
| Hip muscle ( | 2.5 ± 1.0 (0.9–3.8) | 5.1 ± 1.2 (3.7–6.6) | 0.003 |
| Kidney ( | 2.4 ± 0.5 (1.9–3.3) | 5.0 ± 2.2 (2.7–8.3) | 0.019 |
| Total ( | 2.5 ± 0.8 (0.9–3.8) | 5.0 ± 1.7 (2.7–8.3) | < 0.001 |
aData are means ± standard deviations, with ranges in parentheses
bData are medians, with ranges in parentheses