| Literature DB >> 31769483 |
Irena Koniarová1,2,3, Ondřej Konček2,4.
Abstract
The Exradin W1 plastic scintillator (Standard Imaging) was examined for measurement suitability in high-energy photon beams, high-energy electron beams and tomotherapy in terms of dose to water for reference fields. For photon beams, pulse repetition rate dependence, calibration stability, noise from the photodiode enclosure, detector motion during treatment, output factors, off-axis doses and percentage depth doses were tested. For electron and tomotherapy beams, energy non-dependence was verified. All features make the detector suitable in small and non-standards fields, and for electron beams.Entities:
Mesh:
Substances:
Year: 2019 PMID: 31769483 DOI: 10.1093/rpd/ncz230
Source DB: PubMed Journal: Radiat Prot Dosimetry ISSN: 0144-8420 Impact factor: 0.972