| Literature DB >> 31766452 |
Rodolfo Rocha Vieira Leocádio1,2, Alan Kardek Rêgo Segundo1, Cibelle Ferreira Louzada2.
Abstract
This work proposes adapting an existing sensor and embedding it on mannequins used in cardiopulmonary resuscitation (CPR) training to accurately measure the amount of air supplied to the lungs during ventilation. Mathematical modeling, calibration, and validation of the sensor along with metrology, statistical inference, and spirometry techniques were used as a base for aquiring scientific knowledge of the system. The system directly measures the variable of interest (air volume) and refers to spirometric techniques in the elaboration of its model. This improves the realism of the dummies during the CPR training, because it estimates, in real-time, not only the volume of air entering in the lungs but also the Forced Vital Capacity (FVC), Forced Expiratory Volume (FEVt) and Medium Forced Expiratory Flow (FEF20-75%). The validation of the sensor achieved results that address the requirements for this application, that is, the error below 3.4% of full scale. During the spirometric tests, the system presented the measurement results of (305 ± 22, 450 ± 23, 603 ± 24, 751 ± 26, 922 ± 27, 1021 ± 30, 1182 ± 33, 1326 ± 36, 1476 ± 37, 1618 ± 45 and 1786 ± 56) × 10-6 m3 for reference values of (300, 450, 600, 750, 900, 1050, 1200, 1350, 1500, 1650 and 1800) × 10-6 m3, respectively. Therefore, considering the spirometry and pressure boundary conditions of the manikin lungs, the system achieves the objective of simulating valid spirometric data for debriefings, that is, there is an agreement between the measurement results when compared to the signal generated by a commercial spirometer (Koko brand). The main advantages that this work presents in relation to the sensors commonly used for this purpose are: (i) the reduced cost, which makes it possible, for the first time, to use a respiratory volume sensor in medical simulators or training dummies; (ii) the direct measurement of air entering the lung using a noninvasive method, which makes it possible to use spirometry parameters to characterize simulated human respiration during the CPR training; and (iii) the measurement of spirometric parameters (FVC, FEVt, and FEF20-75%), in real-time, during the CPR training, to achieve optimal ventilation performance. Therefore, the system developed in this work addresses the minimum requirements for the practice of ventilation in the CPR maneuvers and has great potential in several future applications.Entities:
Keywords: cardiopulmonary resuscitation; medical mannequins; sensor; spirometric techniques
Mesh:
Year: 2019 PMID: 31766452 PMCID: PMC6929026 DOI: 10.3390/s19235095
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1YF-S201 flow sensor: (a) Input profile; (b) Control volume; (c) Output profile; (d) Propeller compartment; (e) Airflow profile; (f) Detail of the propeller.
The relationship between the length of stem and volume provided by the syringe.
|
| 0 | 300 | 450 | 600 | 750 | 900 | 1050 | 1200 | 1350 | 1500 | 1650 | 1800 |
|
| 0 | 42 | 64 | 85 | 106 | 127 | 148 | 169 | 191 | 212 | 233 | 254 |
Figure 2Volume versus Time chart generated by Koko spirometer.
Figure 3Components for calibrating the spirometer: (a) Calibration syringe; (b) Koko flow spirometer model 313105.
Figure 4Air route inside the manikin. Adapted from CanStock and SimulaCare.
Figure 5Error curve of the YF-S201 sensor.
Measurements after calibration.
|
| 300 | 450 | 600 | 750 | 900 | 1050 | 1200 | 1350 | 1500 | 1650 | 1800 |
|
| 301 | 450 | 601 | 751 | 901 | 1050 | 1201 | 1356 | 1500 | 1650 | 1800 |
|
| 22 | 23 | 24 | 26 | 27 | 30 | 33 | 36 | 37 | 45 | 56 |
Figure 6Result of validation of YF-S201 sensor with air: (a) 300 × 10−6 m3; (b) 450 × 10−6 m3; (c) 600 × 10−6 m3; (d) 750 × 10−6 m3; (e) 900 × 10−6 m3; (f) 1050 × 10−6 m3; (g) 1200 × 10−6 m3; (h) 1350 × 10−6 m3; (i) 1500 × 10−6 m3; (j) 1650 × 10−6 m3 and (k) 1800 × 10−6 m3.
R-Square of the non-linear adjustments.
| Volume | R-Square | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| BTZ | LG | ML | Doseresp | Gompertz | Slogistic | LA | ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| 300 | I | N | I | N | I | 0.99999997354852 | I | 0.99999996735480 | I | N | I | 0.99999968697100 | I | N |
| 450 | 0.99999998774221 | 0.99999998313179 | 0.99999956213635 | |||||||||||
| 600 | 0.99999999387587 | 0.99999998738478 | 0.99999994155838 | |||||||||||
| 750 | 0.99999997787916 | 0.99999998623968 | 0.99999996494737 | |||||||||||
| 900 | 0.99999999363261 | 0.99999999373928 | 0.99999998003168 | |||||||||||
| 1050 | 0.99999998596560 | 0.99999998676343 | 0.99999977894465 | |||||||||||
| 1200 | 0.99999999843628 | 0.99999999853441 | 0.99999994501474 | |||||||||||
| 1350 | 0.99999999976504 | 0.99999999797433 | 0.99999994523958 | |||||||||||
| 1500 | 0.99999999614389 | 0.99999999677671 | 0.99999993443600 | |||||||||||
| 1650 | 0.99999999808238 | 0.99999999767503 | 0.99999680496200 | |||||||||||
| 1800 | 0.99999998519947 | 0.99999998367641 | 0.99999792109656 | |||||||||||
IA—Inadequate accuracy. NC—Not converged.
Parameters of convergence of non-linear adjustment applied to the results.
| Volume (×10−6 m3) | Y0 | xc | C | s |
|---|---|---|---|---|
| 300 | 98 ± 21 | 0.30 ± 0.07 | 258 ± 29 | 0.22 ± 0.02 |
| 450 | 182 ± 10 | 0.47 ± 0.02 | 320 ± 14 | 0.19 ± 0.01 |
| 600 | 284 ± 3 | 0.72 ± 0.01 | 372 ± 5 | 0.18 ± 0.01 |
| 750 | 364 ± 6 | 0.89 ± 0.02 | 467 ± 12 | 0.21 ± 0.02 |
| 900 | 515 ± 12 | 1.20 ± 0.03 | 735 ± 39 | 0.38 ± 0.03 |
| 1050 | 535 ± 9 | 1.15 ± 0.03 | 745 ± 33 | 0.33 ± 0.03 |
| 1200 | 608 ± 4 | 1.43 ± 0.01 | 891 ± 15 | 0.45 ± 0.01 |
| 1350 | 672 ± 5 | 1.62 ± 0.01 | 999 ± 19 | 0.51 ± 0.02 |
| 1500 | 709 ± 6 | 1.56 ± 0.02 | 997 ± 14 | 0.47 ± 0.01 |
| 1650 | 881 ± 6 | 2.40 ± 0.02 | 1218 ± 17 | 0.66 ± 0.02 |
| 1800 | 1006 ± 24 | 2.98 ± 0.06 | 1354 ± 54 | 0.77 ± 0.05 |
Comparison between spirometric results of Koko and the sensor developed in this work.
| Reference | Measured Volume | FVC | tFVC | FEVt=1 s | FEF25–75% | |||||
|---|---|---|---|---|---|---|---|---|---|---|
| YF-S201 | Koko | YF-S201 | Koko | YF-S201 | Koko | YF-S201 | Koko | YF-S201 | Koko | |
| 300 | 305 ± 22 | 320 ± 23 | 305 ± 22 | 320 ± 23 | 1.4 ± 0.1 | 1.3 ± 0.1 | 274 ± 22 | 260 ± 23 | 355 ± 26 | 230 ± 40 |
| 450 | 450 ± 23 | 450 ± 45 | 450 ± 23 | 450 ± 45 | 1.7 ± 0.1 | 1.6 ± 0.2 | 384 ± 23 | 310 ± 45 | 500 ± 77 | 270 ± 81 |
| 600 | 603 ± 24 | 610 ± 80 | 603 ± 24 | 610 ± 80 | 2.0 ± 0.1 | 2.0 ± 0.3 | 463 ± 24 | 360 ± 80 | 617 ± 87 | 330 ± 143 |
| 750 | 751 ± 26 | 760 ± 100 | 751 ± 26 | 760 ± 100 | 2.1 ± 0.1 | 2.2 ± 0.3 | 414 ± 26 | 390 ± 100 | 565 ± 78 | 330 ± 143 |
| 900 | 922 ± 27 | 890 ± 100 | 922 ± 27 | 890 ± 100 | 2.1 ± 0.1 | 2.1 ± 0.2 | 395 ± 27 | 490 ± 100 | 585 ± 67 | 480 ± 150 |
| 1050 | 1051 ± 30 | 1050 ± 100 | 1051 ± 30 | 1050 ± 100 | 2.4 ± 0.1 | 2.4 ± 0.2 | 419 ± 30 | 490 ± 100 | 635 ± 78 | 420 ± 124 |
| 1200 | 1182 ± 33 | 1200 ± 100 | 1182 ± 33 | 1200 ± 100 | 2.7 ± 0.1 | 2.7 ± 0.2 | 334 ± 33 | 530 ± 100 | 565 ± 75 | 490 ± 139 |
| 1350 | 1326 ± 36 | 1350 ± 100 | 1326 ± 36 | 1350 ± 100 | 3.2 ± 0.1 | 3.2 ± 0.2 | 316 ± 36 | 560 ± 100 | 597 ± 89 | 470 ± 129 |
| 1500 | 1476 ± 37 | 1500 ± 100 | 1476 ± 37 | 1500 ± 100 | 3.6 ± 0.1 | 3.7 ± 0.2 | 372 ± 37 | 570 ± 100 | 610 ± 96 | 440 ± 118 |
| 1650 | 1618 ± 45 | 1650 ± 100 | 1618 ± 45 | 1650 ± 100 | 4.1 ± 0.1 | 4.2 ± 0.3 | 197 ± 45 | 530 ± 100 | 565 ± 107 | 430 ± 156 |
| 1800 | 1786 ± 56 | 1800 ± 100 | 1786 ± 56 | 1800 ± 100 | 4.9 ± 0.1 | 5.0 ± 0.3 | 179 ± 56 | 530 ± 100 | 525 ± 127 | 380 ± 136 |
Simultaneous measurements of the Laerdal® and YF-S201 sensors.
| Laerdal® (×10−6 m3) | Indicators | This Work (×10−6 m3) |
|---|---|---|
| 0 | Off | 196 ± 2 |
| Orange | 215 ± 2 | |
| Orange | 282 ± 2 | |
| Orange | 328 ± 3 | |
| Orange | 373 ± 3 | |
| ≤400 ± 60 | Orange | 419 ± 3 |
| >400 ± 60 | Green | 557 ± 4 |
| Green | 663 ± 5 | |
| ≤600 ± 90 | Green | 851 ± 6 |
| >600 ± 90 | Red | 1096 ± 2 |