Assessing the Use of Wrist-Worn Devices in Patients With Heart Failure: Feasibility Study.

Yasbanoo Moayedi1, Raghad Abdulmajeed2, Juan Duero Posada1, Farid Foroutan1, Ana Carolina Alba1, Joseph Cafazzo2, Heather Joan Ross1.   

Abstract

BACKGROUND: Exercise capacity and raised heart rate (HR) are important prognostic markers in patients with heart failure (HF). There has been significant interest in wrist-worn devices that track activity and HR.
OBJECTIVE: We aimed to assess the feasibility and accuracy of HR and activity tracking of the Fitbit and Apple Watch.
METHODS: We conducted a two-phase study assessing the accuracy of HR by Apple Watch and Fitbit in healthy participants. In Phase 1, 10 healthy individuals wore a Fitbit, an Apple Watch, and a GE SEER Light 5-electrode Holter monitor while exercising on a cycle ergometer with a 10-watt step ramp protocol from 0-100 watts. In Phase 2, 10 patients with HF and New York Heart Association (NYHA) Class II-III symptoms wore wrist devices for 14 days to capture overall step count/exercise levels.
RESULTS: Recorded HR by both wrist-worn devices had the best agreement with Holter readings at a workload of 60-100 watts when the rate of change of HR is less dynamic. Fitbit recorded a mean 8866 steps/day for NYHA II patients versus 4845 steps/day for NYHA III patients (P=.04). In contrast, Apple Watch recorded a mean 7027 steps/day for NYHA II patients and 4187 steps/day for NYHA III patients (P=.08).
CONCLUSIONS: Both wrist-based devices are best suited for static HR rate measurements. In an outpatient setting, these devices may be adequate for average HR in patients with HF. When assessing exercise capacity, the Fitbit better differentiated patients with NYHA II versus NYHA III by the total number of steps recorded. This exploratory study indicates that these wrist-worn devices show promise in prognostication of HF in the continuous monitoring of outpatients. ©Yasbanoo Moayedi, Raghad Abdulmajeed, Juan Duero Posada, Farid Foroutan, Ana Carolina Alba, Joseph Cafazzo, Heather Joan Ross. Originally published in JMIR Cardio (http://cardio.jmir.org), 19.12.2017.

Entities:  

Keywords:  Apple watch; Fitbit; MeSH: exercise physiology; heart failure; heart rate tracker; steps; wrist worn devices

Year:  2017        PMID: 31758789      PMCID: PMC6857992          DOI: 10.2196/cardio.8301

Source DB:  PubMed          Journal:  JMIR Cardio        ISSN: 2561-1011


Introduction

Exercise capacity and raised heart rate (HR) are important prognostic markers in patients with heart failure (HF) [1,2]. In clinic, we rely on patients’ self-reported exercise capacity and classify their symptoms based on the New York Heart Association (NYHA) scale. Although widely used, this classification is subjective and poorly reproducible [2]. Furthermore, clinicians are exposed to only a snapshot of the patients’ HR in the ambulatory setting. There has been significant growth in wrist-worn fitness devices that track activity and HR [3]. These wearable devices use infrared and green light emitting diodes to track HR using the photoplethysmography (PPG) method [4]. The aim of this study was to validate the accuracy of HR monitoring using Fitbit and Apple Watch at rest and during structured cardiopulmonary exercise testing in healthy individuals and to then examine the relationship of physical activity in patients with HF.

Methods

We conducted a two-phase study assessing the accuracy of HR using Apple Watch and Fitbit wrist-based devices in healthy participants and then as continuous HR monitoring in patients with HF. In Phase 1, 10 healthy individuals wore a Fitbit, an Apple Watch, and a GE SEER Light 5-electrode Holter monitor while exercising on a cycle ergometer with a 10-watt step ramp protocol. During the first 60 seconds of the test, the workload was set to 0 watts and followed by increments of 10 watts with a maximum workload of 100 watts. In the recovery period, the workload was decreased to 10 watts. In Phase 1, two participants were excluded as data from one device could not be recorded. In Phase 2, 10 patients with HF with NYHA Class II-III symptoms wore both wrist devices for 14 days to capture overall step count/exercise levels. Two patients were excluded due to incomplete data recorded. For Phase 1, we calculated a single measures intraclass correlation (ICC) and a 95% confidence interval, specific to each exercise workload, between HR measured by Holter (gold standard) and HR measured by the Fitbit and Apple Watch devices. The mean ICC and 95% confidence interval for each device against the Holter as the gold standard was plotted against exercise workload. For Phase 2, we used a Kruskal-Wallis rank test to compare the mean number of steps, calculated by either device, between patients in NYHA Class II or III. We used STATA 13.1 and SPSS 22 statistical packages for our statistical analysis. The study protocol was approved by the University Health Network Research Ethics Board.

Results

Recorded HR by both devices was not significantly related with the Holter HR at rest (Fitbit ICC=.263, 95% CI 0.257-0.447, Apple Watch ICC=.218, 95% CI 0.010-0.408). However, with cycle ergometer workloads of 60-100 watts, both devices had stronger agreement with the Holter HR (Figure 1).
Figure 1

Intraclass correlation curves (solid) with 95% confidence intervals (dotted) for workload comparing Fitbit to Holter and Apple Watch to Holter in healthy individuals.

Table 1 shows the baseline demographics of the patients included in Phase 2. Patients were predominantly male (5/8, 63%), with an average age of 58 years and ischemic cardiomyopathy (5/8, 63%). All patients were on guideline-directed medical therapy including a betablocker and either an angiotensin-converting enzyme (ACE) inhibitor or angiotensin-receptor blocker (ARB) when indicated. As shown in Figure 2, Fitbit recorded a mean 8866 steps/day for NYHA II patients versus 4845 steps/day for NYHA III patients (P=.04). In contrast, Apple Watch recorded a mean 7027 steps/day for NYHA II patients and 4187 steps/day for NYHA III patients (P=.08).
Table 1

Demographics and baseline data.

NumberAge (years)GenderLVEFa, %Etiology of HFbNYHAc classMedicationsd
BetablockerAmiodaroneOther
167Male40Ischemic3Bisoprolol 2.5 mgNoneCandesartan 8 mg
268Male18Ischemic2Bisoprolol 10 mg200Irbesartan 300 mg
363Male25Ischemic3Bisoprolol 10 mgNonePerindopril 8 mg
461Female27Non-ischemic2Bisoprolol 10 mgNonePerindopril 4 mg
552Male25Ischemic2Bisoprolol 10 mgNonePerindopril 8 mg
657Female27Non-ischemic3Carvedilol 25 mgNoneRamipril 2.5 mg
758Female60Familial2NoneNoneNone
835Male33Hypertrophic3Carvedilol 50 mgNoneRamipril 10 mg

aLVEF: left ventricular ejection fraction.

bHF: heart failure.

cNYHA: New York Heart Association.

dDrug doses are total daily dose.

Figure 2

Total daily steps recorded (with 95% confidence interval) by Fitbit and Apple watch over 14 days for patients with New York Heart Association (NYHA) Class II and III symptoms .

Intraclass correlation curves (solid) with 95% confidence intervals (dotted) for workload comparing Fitbit to Holter and Apple Watch to Holter in healthy individuals. Demographics and baseline data. aLVEF: left ventricular ejection fraction. bHF: heart failure. cNYHA: New York Heart Association. dDrug doses are total daily dose. Total daily steps recorded (with 95% confidence interval) by Fitbit and Apple watch over 14 days for patients with New York Heart Association (NYHA) Class II and III symptoms .

Discussion

Principal Findings

In this study, both PPG-based monitors had difficulty predicting HR when compared to a 5-lead electrode electrocardiogram (ECG) Holter as the gold standard in a cardiopulmonary study where HR is expected to change quickly over a 10-minute period. A recent study by Wang et al reported similar variability among four popular wrist-worn devices in relation to standard ECG limb leads and a Polar H7 chest strap monitor [5]. The HR underestimation was due to the inherent limitation of PPG that requires a longer settling time processing and averaging to eliminate optical and motion artifact. As a result, PPG monitors underestimated HR and showed poor correlation until the latter stages of the ramp study where HR would tend to saturate and the settling time was sufficient for the PPG HR estimation to compare favorably to the ECG gold standard. Although this indicates that PPG is less suited for dynamic HR measurements, in an outpatient monitoring context, PPG may be suitable for long-term static measurement of HR over long periods of time where settling time would not be an issue [6].

Strengths and Limitations

When assessing exercise capacity, the Fitbit better differentiated patients with NYHA II versus III by the total number of steps recorded. The limitations in this study include the small sample size; larger studies will be needed to confirm these findings. This is the first study in the literature that suggests the possibility of better classifying patients, quantitatively, and potentially remotely, rather than the current practice of determining class through self-reported symptoms with its inherent limitations. Reductions in daily step counts may herald the onset of progressive NYHA symptoms alerting physicians to assess patients in a timely manner.

Conclusion

This exploratory study indicates that wrist-worn devices show promise in prognostication of HF in the continuous monitoring of outpatients, but they require further validation.
  6 in total

Review 1.  Photoplethysmography and its application in clinical physiological measurement.

Authors:  John Allen
Journal:  Physiol Meas       Date:  2007-02-20       Impact factor: 2.833

2.  Accuracy of smartphone applications and wearable devices for tracking physical activity data.

Authors:  Meredith A Case; Holland A Burwick; Kevin G Volpp; Mitesh S Patel
Journal:  JAMA       Date:  2015-02-10       Impact factor: 56.272

3.  Accuracy of Wrist-Worn Heart Rate Monitors.

Authors:  Robert Wang; Gordon Blackburn; Milind Desai; Dermot Phelan; Lauren Gillinov; Penny Houghtaling; Marc Gillinov
Journal:  JAMA Cardiol       Date:  2017-01-01       Impact factor: 14.676

4.  Limitations of the New York Heart Association functional classification system and self-reported walking distances in chronic heart failure.

Authors:  Claire Raphael; Cathy Briscoe; Justin Davies; Zachary Ian Whinnett; Charlotte Manisty; Richard Sutton; Jamil Mayet; Darrel P Francis
Journal:  Heart       Date:  2006-09-27       Impact factor: 5.994

5.  Association of heart rate and outcomes in a broad spectrum of patients with chronic heart failure: results from the CHARM (Candesartan in Heart Failure: Assessment of Reduction in Mortality and morbidity) program.

Authors:  Davide Castagno; Hicham Skali; Madoka Takeuchi; Karl Swedberg; Salim Yusuf; Christopher B Granger; Eric L Michelson; Marc A Pfeffer; John J V McMurray; Scott D Solomon
Journal:  J Am Coll Cardiol       Date:  2012-05-15       Impact factor: 24.094

6.  Accuracy of a Wrist-Worn Wearable Device for Monitoring Heart Rates in Hospital Inpatients: A Prospective Observational Study.

Authors:  Ryan R Kroll; J Gordon Boyd; David M Maslove
Journal:  J Med Internet Res       Date:  2016-09-20       Impact factor: 5.428

  6 in total
  6 in total

Review 1.  The Role of Wearables in Heart Failure.

Authors:  Arvind Singhal; Martin R Cowie
Journal:  Curr Heart Fail Rep       Date:  2020-08

2.  Wearable Monitoring and Interpretable Machine Learning Can Objectively Track Progression in Patients during Cardiac Rehabilitation.

Authors:  Hélène De Cannière; Federico Corradi; Christophe J P Smeets; Melanie Schoutteten; Carolina Varon; Chris Van Hoof; Sabine Van Huffel; Willemijn Groenendaal; Pieter Vandervoort
Journal:  Sensors (Basel)       Date:  2020-06-26       Impact factor: 3.576

Review 3.  Smart Wearables for Cardiac Monitoring-Real-World Use beyond Atrial Fibrillation.

Authors:  David Duncker; Wern Yew Ding; Susan Etheridge; Peter A Noseworthy; Christian Veltmann; Xiaoxi Yao; T Jared Bunch; Dhiraj Gupta
Journal:  Sensors (Basel)       Date:  2021-04-05       Impact factor: 3.576

Review 4.  Digital Technology Application for Improved Responses to Health Care Challenges: Lessons Learned From COVID-19.

Authors:  Darshan H Brahmbhatt; Heather J Ross; Yasbanoo Moayedi
Journal:  Can J Cardiol       Date:  2021-12-01       Impact factor: 5.223

Review 5.  Digital Technologies to Support Better Outcome and Experience of Care in Patients with Heart Failure.

Authors:  K C C McBeath; C E Angermann; M R Cowie
Journal:  Curr Heart Fail Rep       Date:  2022-04-29

Review 6.  Digital Health: Implications for Heart Failure Management.

Authors:  Arvind Singhal; Martin R Cowie
Journal:  Card Fail Rev       Date:  2021-05-11
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.