An inexpensive servo-respirator based upon regulation of a shunt resistance.

J A Daubenspeck1, D Pichon, K V Knuth, D Bartlett, W M St John.   

Abstract

We have designed and built an inexpensive servo-respirator for use in investigations of respiratory control in small animals. The device uses a butterfly valve to alter the resistance of an outflow shunt from a manifold that connects the animal's tracheal cannula to a pressure source. Tracheal pressure is regulated in response to a command provided by a suitably processed neural signal, often the integrated phrenic neurogram. As the valve opens, tracheal pressure approaches atmospheric; as it closes, tracheal pressure approaches the source pressure. An electronic controller circuit was developed to permit experimental procedures that include withholding volume delivery while maintaining a desired level of positive end-expiratory pressure. The device is able to track the neural command signal satisfactorily, and its performance appears to be limited primarily by the constraints applied by the respiratory system mechanics.

Mesh:

Year:  1988        PMID: 3175357     DOI: 10.1016/0034-5687(88)90129-6

Source DB:  PubMed          Journal:  Respir Physiol        ISSN: 0034-5687


  2 in total

1.  Discharge of vagal pulmonary receptors differentially alters neural activities during various stages of expiration in the cat.

Authors:  W M St John; D Zhou
Journal:  J Physiol       Date:  1990-05       Impact factor: 5.182

2.  Alteration of ventilatory activity by intralaryngeal CO2 in the cat.

Authors:  D Bartlett; S L Knuth; J C Leiter
Journal:  J Physiol       Date:  1992-11       Impact factor: 5.182

  2 in total