Literature DB >> 31743293

Electrode Array Type and Its Impact on Impedance Fluctuations and Loss of Residual Hearing in Cochlear Implantation.

Nicholas J Thompson1, Margaret T Dillon, Emily Buss, Lisa R Park, Harold C Pillsbury, Brendan P O'Connell, Kevin D Brown.   

Abstract

HYPOTHESIS/
OBJECTIVE: Determine variables associated with electrode impedance fluctuations and loss of residual hearing in cochlear implant (CI) recipients.
BACKGROUND: CI recipients with postoperative hearing preservation demonstrate superior speech perception with an electric-acoustic stimulation (EAS) device as compared with a CI-alone device. Maintaining superior speech perception over time relies on long-term hearing preservation; therefore, understanding variables that may contribute to loss of residual hearing is needed. Recent reports suggest a relationship between changes in electrode impedance and loss of residual hearing. The variables influencing this relationship have yet to be determined.
METHODS: Review of pediatric and adult CI cases from 2013 to 2016 who presented with preoperative residual hearing. Regression analysis was performed to evaluate effects of array type (lateral wall vs. perimodiolar), manufacturer, age at implantation, and preoperative hearing on impedance. The correlation between peak impedance change and change in low-frequency hearing was determined.
RESULTS: One hundred forty-six CI recipients presented with preoperative residual hearing. A multivariate regression analysis demonstrated a statistically significant association between preoperative hearing thresholds (p = 0.017), device manufacturer (p = 0.011), and array type (p = 0.038) on postoperative impedance changes. Hearing preservation rates and change in impedance differed by electrode array type. The association between peak impedance changes and loss of residual hearing differed between manufacturers (R = 0.208, p = 0.029 vs. R = 0.016, p = 0.609).
CONCLUSION: Impedance fluctuation appears to be a marker for loss of residual hearing for specific electrode array types and manufacturers. Specific arrays may affect the cochlear microenvironment differently, with different effects on postoperative hearing preservation.

Mesh:

Year:  2020        PMID: 31743293     DOI: 10.1097/MAO.0000000000002457

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  4 in total

1.  Timing of Acoustic Hearing Changes After Cochlear Implantation.

Authors:  Megan J Jensen; Heba Isaac; Helin Hernandez; Jacob Oleson; Camille Dunn; Bruce J Gantz; Marlan R Hansen
Journal:  Laryngoscope       Date:  2021-12-09       Impact factor: 2.970

2.  Access and Polarization Electrode Impedance Changes in Electric-Acoustic Stimulation Cochlear Implant Users with Delayed Loss of Acoustic Hearing.

Authors:  Viral D Tejani; Hyejin Yang; Jeong-Seo Kim; Helin Hernandez; Jacob J Oleson; Marlan R Hansen; Bruce J Gantz; Paul J Abbas; Carolyn J Brown
Journal:  J Assoc Res Otolaryngol       Date:  2021-10-22

3.  Increasing the reliability of real-time electrocochleography during cochlear implantation: a standardized guideline.

Authors:  K Schuerch; M Waser; G Mantokoudis; L Anschuetz; M Caversaccio; W Wimmer; S Weder
Journal:  Eur Arch Otorhinolaryngol       Date:  2022-01-20       Impact factor: 3.236

4.  Cochlear Implant Electrode Impedance as Potential Biomarker for Residual Hearing.

Authors:  Wilhelm Wimmer; Luca Sclabas; Marco Caversaccio; Stefan Weder
Journal:  Front Neurol       Date:  2022-06-27       Impact factor: 4.086

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.