A self-checking treatment couch coordinate calculation system in radiotherapy.

Pingfang Tsai1, Chihray Liu1, Darren L Kahler1, Jonathan G Li1, Bo Lu1, Guanghua Yan1.   

Abstract

PURPOSE: Traditionally, the treatment couch coordinates (TCCs) for patients undergoing radiotherapy can only be determined at the time of treatment, placing pressure on the treating therapists and leaving several pathways for errors such as wrong-site treatment or wrong treatment table shift from a reference point. The purpose of this work is to propose an accurate, robust, and streamlined system that calculates TCC in advance.
METHODS: The proposed system combines the advantages of two different calculation methods that use an indexed immobilization device. The first method uses an array of reference ball bearings (BBs) embedded in the CT scanner's couch-top. To obtain the patient-specific TCC, the spatial offset of the treatment planning isocenter from the reference BB is used. The second method performs a calculation using the one-to-one mapping relationship between the CT scanner's DICOM (Digital Imaging and Communications in Medicine) coordinate system and the TCC system. Both methods use a reference point in the CT coordinate system to correlate a point in the TCC system to perform the coordinate transfer between the two systems. Both methods were used to calculate the TCC and the results were checked against each other, creating an integrated workflow via automated self-checking. The accuracy of the calculation system was retrospectively evaluated with 275 patients, where the actual treatment position determined with cone-beam CT was used as a reference.
RESULTS: An efficient workflow transparent to the therapists at both CT simulation and treatment was created. It works with any indexed immobilization device and can be universally applied to all treatment sites. The two methods had comparable accuracy, with 95% of the calculations within 3 mm. The inter-fraction variation was within ± 1.0 cm for 95% of the coordinates across all the treatment sites.
CONCLUSIONS: A robust, accurate, and streamlined system was implemented to calculate TCCs in advance. It eases the pressure on the treating therapists, reduces patient setup time, and enhances the patient safety by preventing setup errors.
© 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  couch coordinate; error prevention; patient setup; workflow automation

Year:  2019        PMID: 31737999     DOI: 10.1002/acm2.12771

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


  3 in total

1.  Principles for Designing and Developing a Workflow Monitoring Tool to Enable and Enhance Clinical Workflow Automation.

Authors:  Danny T Y Wu; Lindsey Barrick; Mustafa Ozkaynak; Katherine Blondon; Kai Zheng
Journal:  Appl Clin Inform       Date:  2022-01-19       Impact factor: 2.342

2.  Automatic couch position calculation using eclipse scripting for external beam radiotherapy.

Authors:  Hesheng Wang; Anthony Rea; Benedikt Rudek; Ting Chen; Allison McCarthy; David Barbee
Journal:  J Appl Clin Med Phys       Date:  2021-01-13       Impact factor: 2.102

3.  Machine learning-based treatment couch parameter prediction in support of surface guided radiation therapy.

Authors:  Geert De Kerf; Michaël Claessens; Isabelle Mollaert; Wim Vingerhoed; Dirk Verellen
Journal:  Tech Innov Patient Support Radiat Oncol       Date:  2022-08-12
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.