Abstract

OBJECTIVES: This study investigated the agreement at the categorical level between the Copan WASPLab incorporating the BioRad expert system against the SIRscan 2000 automatic for antimicrobial disc diffusion susceptibility testing.
METHODS: The 338 clinical strains (67 Pseudomonas aeruginosa, 19 methicillin-resistant Staphylococcus aureus, 75 methicillin-sensitive S. aureus and 177 Enterobacterales isolates) analysed in this study were non-duplicate isolates obtained from consecutive clinical samples referred to the clinical bacteriology laboratory at Geneva University Hospitals between June and August 2019. For the WASPLab the inoculum suspension was prepared in strict accordance with the manufacturer's instruction (Copan WASP srl, Brescia, Italy) by adding 2 mL of the 0.5 McFarland primary suspension used for the SIRscan analysis into a sterile tube filled with 4 mL of sterile saline (1:3 dilution). The inoculum (2 × 30 μL loop/spreader) was spread over the entire surface of Mueller-Hinton agar plates according to the AST streaking pattern defined by Copan. The antibiotic discs were dispensed by the WASP and inoculated media were loaded on conveyors for transfer to the automatic incubators. The plates were incubated for 16 h, and several digital images were acquired. Inhibition zone diameters were automatically read by the WASPLab and were adjusted manually whenever necessary. For the SIRscan 2000 automatic, the antimicrobial disc diffusion susceptibility testing was performed according to the EUCAST guidelines. The gradient strip method was used to resolve discrepancies.
RESULTS: The overall categorical agreement between the compared methods reached 99.1% (797/804; 95% CI 98.2%-99.6%), 99.5% (1029/1034; 95% CI 98.9%-99.8%), and 98.8% (2798/2832; 95% CI 98.3%-99.1%) for P. aeruginosa, S. aureus and the Enterobacterales, respectively.
CONCLUSIONS: WASPLab incorporating the BioRad expert system provides a fully automated solution for antimicrobial disc diffusion susceptibility testing with equal or better accuracy than other available phenotypic methods.
Copyright © 2019 European Society of Clinical Microbiology and Infectious Diseases. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Accuracy; Antimicrobial susceptibility testing; Automation; BioRad expert system; Copan WASPLab; Disc diffusion; Medical value; Quality control; SIRscan 2000 automatic; Turn-around time

Mesh:

Substances:

Year:  2019        PMID: 31733376     DOI: 10.1016/j.cmi.2019.11.008

Source DB:  PubMed          Journal:  Clin Microbiol Infect        ISSN: 1198-743X            Impact factor:   8.067


  4 in total

1.  Computer Vision and Artificial Intelligence Are Emerging Diagnostic Tools for the Clinical Microbiologist.

Authors:  Daniel D Rhoads
Journal:  J Clin Microbiol       Date:  2020-05-26       Impact factor: 5.948

Review 2.  Mini Review: Clinical Routine Microbiology in the Era of Automation and Digital Health.

Authors:  Stefano Leo; Abdessalam Cherkaoui; Gesuele Renzi; Jacques Schrenzel
Journal:  Front Cell Infect Microbiol       Date:  2020-11-30       Impact factor: 5.293

Review 3.  Total Laboratory Automation for Rapid Detection and Identification of Microorganisms and Their Antimicrobial Resistance Profiles.

Authors:  Abdessalam Cherkaoui; Jacques Schrenzel
Journal:  Front Cell Infect Microbiol       Date:  2022-02-03       Impact factor: 5.293

Review 4.  Recent Developments in Phenotypic and Molecular Diagnostic Methods for Antimicrobial Resistance Detection in Staphylococcus aureus: A Narrative Review.

Authors:  Andrea Sanchini
Journal:  Diagnostics (Basel)       Date:  2022-01-15
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.