Faezeh Bakhshi1,2,3, Ommoleila Molavi2,4, Mohammad Reza Rashidi1,2, Ali Shayanfar5, Hassan Amini6,7.
Abstract
OBJECTIVE: Silibinin is an antioxidant agent and is shown to have anticancer effects in different cancers including lung, breast, colorectal, liver, prostate, and kidney. There are challenges in the clinical use of silibinin. The main limitation is low solubility, poor oral absorption, and extensive hepatic metabolism. We aim to develop a High-Performance Liquid Chromatography (HPLC) sensitive method for quantification of silibinin in aqueous samples to quantify its concentration in new formulations. A reverse-phase high-performance liquid chromatography (RP-HPLC) composed of C18 column as stationary phase and the mixture of methanol (90%) and water (10%) as mobile phase. The developed method was validated based on the established guidelines.Entities:
Keywords: Analysis method; HPLC; Method validation; Precision and accuracy; Silibinin
Mesh:
Substances:
Year: 2019 PMID: 31727143 PMCID: PMC6854794 DOI: 10.1186/s13104-019-4774-2
Source DB: PubMed Journal: BMC Res Notes ISSN: 1756-0500
Fig. 1a Silibinin chromatogram in aqueous samples with a concentration of 50 µg/ml. b Silibinin calibration curve in aqueous samples: the horizontal axis shows the injected silibinin concentrations (µg/ml), and the vertical axis shows peak height of chromatograms)
Linear parameters for in vitro samples
| Parameters | Value |
|---|---|
| Slope | 3110.6 |
| Intercept | − 17,877 |
| Standard deviation of calibration curve | 7515.2 |
| Linear range (µg/ml) | 10–100 |
| Coefficient of determination (R2) | 0.996 |
| LLOQ (µg/ml) | 10 |
| ULOQ (µg/ml) | 100 |
| LOD (µg/ml) | 7.25 |
| LOQ (µg/ml) | 24.2 |
Calibration curve parameters, linear range, LLOQ, ULOQ, LOD, LOQ for Silibinin in in vitro samples
Accuracy and precision
| Added concentration | Calculated concentration | RSD (%) | Accuracy (%) |
|---|---|---|---|
| Inter-day | |||
| 15.0 | 14.3 | 10.6 | 95.1 |
| 50.0 | 53.0 | 3.1 | 105.9 |
| 75.0 | 75.8 | 2.7 | 101.1 |
| Intra-day | |||
| 15.0 | 13.2 | 10.9 | 88.0 |
| 50.0 | 50.6 | 4.0 | 101.2 |
| 75.0 | 75.4 | 0.8 | 100.5 |
Inter-day and intra-day accuracy and precision of the developed method for in vitro method