| Literature DB >> 31727047 |
Francesca Gazzani1, Roberta Lione2,3, Chiara Pavoni2,3, Gianluca Mampieri2, Paola Cozza2,4.
Abstract
BACKGROUND: The aim of the present investigation was to evaluate enamel reduction efficiency, abrasive property decay, and enamel effects between oscillating mechanical and manual systems for interproximal enamel reduction (IPR).Entities:
Keywords: Interproximal enamel reduction; Manual strips; Mechanical strips; SEM analysis
Mesh:
Substances:
Year: 2019 PMID: 31727047 PMCID: PMC6854626 DOI: 10.1186/s12903-019-0934-y
Source DB: PubMed Journal: BMC Oral Health ISSN: 1472-6831 Impact factor: 2.757
Fig. 1Oscillating diamond strip and contra-angle adapted on the Instron Universal Testing Machine for the experimental analysis
Descriptive statistics and statistical comparisons (independent-samples t tests) of the enamel reduction efficiency in mm of removed enamel. Mean values obtained from 24 cycles performed for 3 oscillating strips and 24 cycles performed for 3 manual strips
| Variable | Oscillating mechanical system (3 strips) | Manual system (3 strips) | Diff. | 95% CI of the difference | ||||
|---|---|---|---|---|---|---|---|---|
| Mean | SD | Mean | SD | Lower | Upper | |||
| Enamel reduction (mm) | 0.16 | 0.04 | 0.09 | 0.04 | 0.07 | 0.004* | 0.025 | 0.110 |
SD Standard Deviations, Diff. Differences, CI Confidence interval
* p < 0.005
Fig. 2Enamel reduction efficiency (mm) comparing two IPR systems during one cycle test
Fig. 3SEM analysis (100X) of non-used mechanical and manual IPR systems
Fig. 4SEM analysis of manual IPR system surface after 8 IPR test cycles. a 50X. b 100X. C 200X
Fig. 5SEM analysis of mechanical IPR system surface after 8 IPR test cycles. a 50X. b 100X. C 200X
Fig. 6SEM analysis of enamel surface after 8 IPR cycles by means of manual IPR system. a 30X. b 140X. C 300X
Fig. 7SEM analysis of enamel surface after 8 IPR cycles by means of mechanical IPR system. a 30X. b 140X. C 300X