What is the safety and efficacy of the use of automated fastener in heart valve surgery?

M Yousuf Salmasi1, Lueh Chien2, Philip Hartley2, Amer Al-Balah2, Kulvinder Lall3, Aung Oo3, Roberto Casula4, Thanos Athanasiou1.   

Abstract

INTRODUCTION: Cor-Knot automated fastener has been used as an adjunct in heart valve surgery to eliminate the need for manual tying during valve implantation. Although reduced operative time and facilitation for minimally invasive surgery are clear benefits, whether their use translates to improved patient outcome remains debatable. This study aims to review the safety and efficacy of automated fasteners in heart valve surgeries.
METHOD: Specific searches were conducted via online medical databases (Pubmed, Embase, Ovid) between 1950 and June 2019. Longitudinal studies were included that provided operative parameters.
RESULTS: The initial literature search identified 3773 articles, but only eight met the inclusion criteria and were used for analysis: four studies related to aortic valve replacement (AVR), four related to mitral valve (MV) intervention (total n = 810). The meta-analysis revealed the significantly shorter aortic cross-clamp time in the Cor-knot group compared to manual tying, both in AVR and MV surgeries (P < .05). Cardiopulmonary bypass time was significantly shorter in the Cor-knot group when analyzing studies in MV surgery (weighted mean difference [WMD]: 110.0; 95% confidence interval: 12.3-207.7; P = .027) The use of Cor-Knot did not increase the risk of permanent pacemaker implantation, paravalvular leak, and 30-day mortality. The majority of studies reported no change in the length of intensive unit care and total hospital stay.
CONCLUSION: We confirmed that the majority of existing literatures indicated the safety and intraoperative efficacy with automated fastener application. Nevertheless, there is currently no evidence to support automated fastened sutures can translate its intraoperative advantages to improved patient outcome.
© 2019 Wiley Periodicals, Inc.

Entities:  

Keywords:  aortic; automated fastener; clinical review; cor-knot; heart valves; mitral

Mesh:

Year:  2019        PMID: 31725943     DOI: 10.1111/jocs.14265

Source DB:  PubMed          Journal:  J Card Surg        ISSN: 0886-0440            Impact factor:   1.620


  2 in total

1.  Safety and Effectiveness Outcomes of a Novel Automated Titanium Suture Fastener Device Applied for Heart Valve Surgery in an Ovine Model.

Authors:  Bin Li; Shanshan Bai; Guangxin Yue; Jinyan Zhu; Min Zhang; Baiqing Yang; Jiafei Luo; Yang Sun; Leisheng Zhang; Xin Wang
Journal:  Front Cardiovasc Med       Date:  2022-02-11

2.  Cor-Knot automated fastener in distal anastomosis of total aortic arch replacement: A case report.

Authors:  Omar M Sharaf; Tomas D Martin; Eric I Jeng
Journal:  Clin Case Rep       Date:  2021-07-21
  2 in total