Designing parallel transmit head coil arrays based on radiofrequency pulse performance.

Zhipeng Cao1,2, Xinqiang Yan1,2, John C Gore1,2,3, William A Grissom2,3.   

Abstract

PURPOSE: A new approach to design parallel transmit (pTx) head arrays is proposed that integrates transmit radiofrequency pulse designs with electromagnetic modeling of array coil elements. THEORY AND METHODS: An approach to design pTx head arrays is proposed that finds optimal groupings of a large number of coils into a small number of channels. An algorithm is proposed to extend array-compressed parallel transmit pulse design by adding the ability to optimally select and prune coil elements, in addition to optimizing compression weights. The performance of the method is demonstrated in simulations of dynamic multislice shimming of the human brain in axial, coronal, and sagittal directions, and of reduced field-of-view excitation targeting the human occipital lobe, with simulated electromagnetic field maps from a group of 5 human head models at 7T.
RESULTS: For both dynamic multislice shimming and reduced field-of-view excitation, the method successfully designed pTx arrays that simultaneously achieved in general 15% lower mean excitation errors with 20% lower SDs, along with 20% lower mean global averaged specific absorption rate and 50% lower SD than previously reported pTx head array designs.
CONCLUSION: With the proposed optimal coil element selection algorithm, the array-compressed parallel transmit pulse design can be extended to design pTx transmit head arrays with joint consideration of the fields within the sample and the radiofrequency pulse. The pTx arrays from such an approach achieved higher transmit excitation accuracy, lower radiofrequency heating in subjects, and more robust performance across subjects compared with previously reported pTx head arrays with the same number of channels.
© 2019 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  coil design; dynamic multislice shimming; excitation accuracy; global SAR; parallel transmit; rFOV excitation

Mesh:

Year:  2019        PMID: 31722120      PMCID: PMC7047538          DOI: 10.1002/mrm.28068

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  32 in total

1.  Bloch-based MRI system simulator considering realistic electromagnetic fields for calculation of signal, noise, and specific absorption rate.

Authors:  Zhipeng Cao; Sukhoon Oh; Christopher T Sica; John M McGarrity; Timothy Horan; Wei Luo; Christopher M Collins
Journal:  Magn Reson Med       Date:  2013-09-04       Impact factor: 4.668

2.  Local specific absorption rate control for parallel transmission by virtual observation points.

Authors:  Gabriele Eichfelder; Matthias Gebhardt
Journal:  Magn Reson Med       Date:  2011-05-20       Impact factor: 4.668

3.  kT -points: short three-dimensional tailored RF pulses for flip-angle homogenization over an extended volume.

Authors:  M A Cloos; N Boulant; M Luong; G Ferrand; E Giacomini; D Le Bihan; A Amadon
Journal:  Magn Reson Med       Date:  2011-05-16       Impact factor: 4.668

4.  A 16-channel dual-row transmit array in combination with a 31-element receive array for human brain imaging at 9.4 T.

Authors:  G Shajan; Mikhail Kozlov; Jens Hoffmann; Robert Turner; Klaus Scheffler; Rolf Pohmann
Journal:  Magn Reson Med       Date:  2014-02       Impact factor: 4.668

5.  A highly decoupled transmit-receive array design with triangular elements at 7T.

Authors:  Gang Chen; Bei Zhang; Martijn A Cloos; Daniel K Sodickson; Graham C Wiggins
Journal:  Magn Reson Med       Date:  2018-03-23       Impact factor: 4.668

6.  Small-tip-angle spokes pulse design using interleaved greedy and local optimization methods.

Authors:  William A Grissom; Mohammad-Mehdi Khalighi; Laura I Sacolick; Brian K Rutt; Mika W Vogel
Journal:  Magn Reson Med       Date:  2012-03-05       Impact factor: 4.668

7.  Parallel-transmission-enabled magnetization-prepared rapid gradient-echo T1-weighted imaging of the human brain at 7 T.

Authors:  M A Cloos; N Boulant; M Luong; G Ferrand; E Giacomini; M-F Hang; C J Wiggins; D Le Bihan; A Amadon
Journal:  Neuroimage       Date:  2012-06-01       Impact factor: 6.556

8.  Joint design of large-tip-angle parallel RF pulses and blipped gradient trajectories.

Authors:  Zhipeng Cao; Manus J Donahue; Jun Ma; William A Grissom
Journal:  Magn Reson Med       Date:  2015-04-27       Impact factor: 4.668

9.  Simultaneous multislice excitation by parallel transmission.

Authors:  Benedikt A Poser; Robert James Anderson; Bastien Guérin; Kawin Setsompop; Weiran Deng; Azma Mareyam; Peter Serano; Lawrence L Wald; V Andrew Stenger
Journal:  Magn Reson Med       Date:  2013-05-28       Impact factor: 4.668

10.  Self-decoupled radiofrequency coils for magnetic resonance imaging.

Authors:  Xinqiang Yan; John C Gore; William A Grissom
Journal:  Nat Commun       Date:  2018-08-28       Impact factor: 14.919

View more
  1 in total

Review 1.  Clinical 7-T MRI for neuroradiology: strengths, weaknesses, and ongoing challenges.

Authors:  Brian J Burkett; Andrew J Fagan; Joel P Felmlee; David F Black; John I Lane; John D Port; Charlotte H Rydberg; Kirk M Welker
Journal:  Neuroradiology       Date:  2021-01-02       Impact factor: 2.804

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.