Red blood cell tolerance to shear stress above and below the subhemolytic threshold.

Jarod T Horobin1,2, Surendran Sabapathy3, Michael J Simmonds3,4.   

Abstract

Mechanical circulatory support device (MCS) design has improved over the years and yet blood damage (e.g., hemolysis) remains a problem. Accumulating evidence indicates a subhemolytic threshold for red blood cells (RBC)-a threshold at which RBC deformability is impaired prior to hemolysis. The current study aimed to assess the deformability of RBC exposed to supra-physiological shear stresses that are typical of MCS devices and assess whether a method used to estimate an individualized subhemolytic threshold, accurately demarcates whether future application of shear stress was damaging. Suspensions of RBC were "conditioned" with discrete magnitudes of shear stress (5-100 Pa) for specific durations (1-16 s). Cellular deformability was subsequently measured via ektacytometry and a mechanical sensitivity (MS) index was calculated to identify the subhemolytic threshold. Thereafter, fresh RBC suspensions were exposed to a magnitude of shear stress 10 Pa above, 10 Pa below, or matched to a donor's previously estimated subhemolytic threshold for a given duration (1, 4, 16 s) to ascertain the sensitivity of the subhemolytic threshold. The MS index of RBC was significantly impaired following exposure to 10 Pa above the subhemolytic threshold (p < 0.0001), and significantly enhanced following exposure to 10 Pa below the subhemolytic threshold (p < 0.01). For all shear conditions, there was no significant increase in free hemoglobin. Functional assessments of RBC may be useful when conducting biocompatibility testing of MCS devices, to detect trauma to blood prior to overt cell rupture being induced.

Entities:  

Keywords:  Erythrocyte; Haemorheology; Hemocompatibility; Mechanical damage; Membrane elasticity

Mesh:

Substances:

Year:  2019        PMID: 31720887     DOI: 10.1007/s10237-019-01252-z

Source DB:  PubMed          Journal:  Biomech Model Mechanobiol        ISSN: 1617-7940


  4 in total

1.  Sub-Fractions of Red Blood Cells Respond Differently to Shear Exposure Following Superoxide Treatment.

Authors:  Marijke Grau; Lennart Kuck; Thomas Dietz; Wilhelm Bloch; Michael J Simmonds
Journal:  Biology (Basel)       Date:  2021-01-11

2.  Near-infrared photobiomodulation of blood reversibly inhibits platelet reactivity and reduces hemolysis.

Authors:  Tomasz Walski; Karolina Grzeszczuk-Kuć; Katarzyna Gałecka; Natalia Trochanowska-Pauk; Raghvendra Bohara; Albert Czerski; Konstanty Szułdrzyński; Wiesław Królikowski; Jerzy Detyna; Małgorzata Komorowska
Journal:  Sci Rep       Date:  2022-03-08       Impact factor: 4.379

3.  Computational fluid dynamics analysis and experimental hemolytic performance of three clinical centrifugal blood pumps: Revolution, Rotaflow and CentriMag.

Authors:  Dong Han; Joshua L Leibowitz; Lu Han; Shigang Wang; Ge He; Bartley P Griffith; Zhongjun J Wu
Journal:  Med Nov Technol Devices       Date:  2022-06-19

4.  Erythrocyte morphological symmetry analysis to detect sublethal trauma in shear flow.

Authors:  Antony P McNamee; Michael J Simmonds; Masataka Inoue; Jarod T Horobin; Masaya Hakozaki; John F Fraser; Nobuo Watanabe
Journal:  Sci Rep       Date:  2021-12-07       Impact factor: 4.379

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.