Reproducibility of hepatic MR elastography across field strengths, pulse sequences, scan intervals, and readers.

Hye Jin Kim1, Bohyun Kim2, Hyun Jeong Yu1, Jimi Huh1, Jei Hee Lee1, Seung Soo Lee3, Kyung Won Kim3, Jai Keun Kim1.   

Abstract

PURPOSE: To evaluate the reproducibility of hepatic MRE under various combinations of settings of field strength, pulse sequence, scan interval, and reader in non-alcoholic fatty liver disease (NAFLD) patients.
METHODS: Adult NAFLD patients were prospectively enrolled for serial hepatic MRE with 1.5 T using 2D GRE sequence and 3.0 T using 2D SE-EPI sequence on the same day and after 2 weeks, resulting a total of four MRE examinations per patient. Three readers with various levels of background knowledge in MRE technique and liver anatomy measured liver stiffness after a training session. Linear regression, Bland-Altman analysis, within-subject coefficient of variation, and reproducibility coefficient (RDC) were used to determine reproducibility of hepatic MRE measurement.
RESULTS: Twenty patients completed the MRE sessions. Liver stiffness through MRE showed pooled RDC of 26% (upper 95% CI 30.6%) and corresponding limits of agreement (LOA) within 0.55 kPa across field strengths, MRE sequences, and 2-week interscan interval in three readers. Small mean biases and narrow LOA were observed among readers (0.05-0.19 kPa ± 0.53).
CONCLUSION: The magnitude of change across combinations of scan parameters is within acceptable clinical range, rendering liver stiffness through MRE a reproducible quantitative imaging biomarker. A lower reproducibility was observed for measurements under different field strengths/MRE sequences at a longer (2 weeks) interscan interval. Operators should be trained to acquire region of interest consistently in repeat examinations.

Entities:  

Keywords:  Magnetic resonance elastography; Non-alcoholic fatty liver disease; Reproducibility of results

Mesh:

Year:  2020        PMID: 31720766     DOI: 10.1007/s00261-019-02312-9

Source DB:  PubMed          Journal:  Abdom Radiol (NY)


  2 in total

Review 1.  Magnetic resonance elastography of the liver: everything you need to know to get started.

Authors:  Kay M Pepin; Christopher L Welle; Flavius F Guglielmo; Jonathan R Dillman; Sudhakar K Venkatesh
Journal:  Abdom Radiol (NY)       Date:  2021-11-01

2.  Reliability of Gradient-Echo Magnetic Resonance Elastography of Lumbar Muscles: Phantom and Clinical Studies.

Authors:  Tsyh-Jyi Hsieh; Ming-Chung Chou; Yi-Chu Chen; Yi-Chen Chou; Chien-Hung Lin; Clement Kuen-Huang Chen
Journal:  Diagnostics (Basel)       Date:  2022-06-03
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.