INTRODUCTION: Woven EndoBridge (WEB) is an innovative device for the treatment of intracranial aneurysms especially wide-necked bifurcation aneurysms. Here we present our experience with the WEB device.
MATERIAL AND METHODS: Patients treated using only the WEB device between September 2014 and November 2018 were included in the study. Follow up imaging studies and medical records of the patients were retrospectively reviewed.
RESULTS: Forty-two aneurysm of 42 patients (27 female, 15 male; median age: 56, range: 32-76) were treated using the WEB device. The mean diameter of the aneurysms was 6.6 mm (range: 3-12 mm). The neck diameter was ≥4 mm in 57% of the aneurysms. The locations of the aneurysms were the middle cerebral artery bifurcation in 29 (69%), basilar tip in 5 (12%), anterior communicating artery in 5 (12%), internal carotid artery tip in 2 (5%), and M1 segment of the middle cerebral artery in 1 (2%) of the patients. Five patients had subarachnoid hemorrhage due to aneurysm rupture. The device could be successfully deployed in all of the cases. There were control imaging studies available for 36 patients who were followed up for a median of 7 months (range: 1-33 months). Adequate occlusion was observed in 35 of these 36 patients (97%). There was no treatment related morbidity or mortality.
CONCLUSION: Although long term follow-up data are not available, WEB intrasaccular flow disruptor seems to be effective and safe for intracranial bifurcation aneurysm treatment in the mid-term follow up.
Copyright © 2019. Published by Elsevier Inc.

Entities:  

Keywords:  Endovascular treatment; Intracranial aneurysm; Woven EndoBridge

Mesh:

Year:  2019        PMID: 31715514     DOI: 10.1016/j.clinimag.2019.09.001

Source DB:  PubMed          Journal:  Clin Imaging        ISSN: 0899-7071            Impact factor:   1.605


  4 in total

1.  Residual Flow Inside the Woven EndoBridge Device at Follow-Up: Potential Predictors of the Bicêtre Occlusion Scale Score 1 Phenomenon.

Authors:  M T Nawka; A Lohse; M Bester; J Fiehler; J-H Buhk
Journal:  AJNR Am J Neuroradiol       Date:  2020-06-25       Impact factor: 3.825

2.  Trends of Expanding Indications of Woven EndoBridge Devices for the Treatment of Intracranial Aneurysms: A Systematic Review and Meta-analysis.

Authors:  Ki Baek Lee; Chong Hyun Suh; Yunsun Song; Boseong Kwon; Mi Hyeon Kim; Jong-Tae Yoon; Deok Hee Lee
Journal:  Clin Neuroradiol       Date:  2022-08-29       Impact factor: 3.156

3.  Safety and Efficacy of the Woven EndoBridge Device for Treatment of Ruptured Intracranial Aneurysms: A Systematic Review and Meta-analysis.

Authors:  M A Essibayi; G Lanzino; W Brinjikji
Journal:  AJNR Am J Neuroradiol       Date:  2021-06-11       Impact factor: 4.966

Review 4.  Advances in endovascular aneurysm management: flow modulation techniques with braided mesh devices.

Authors:  Jessica K Campos; Barry Cheaney Ii; Brian V Lien; David A Zarrin; Chau D Vo; Geoffrey P Colby; Li-Mei Lin; Alexander L Coon
Journal:  Stroke Vasc Neurol       Date:  2020-03-25
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.