Titanium Implant Characteristics After Implantoplasty: An In Vitro Study on Two Different Kinds of Instrumentation.

Philipp Sahrmann, Sandra Luso, Constanze Mueller, Andreas Ender, Thomas Attin, Bogna Stawarczyk, Patrick R Schmidlin.   

Abstract

PURPOSE: To assess surface characteristics and implant stability after implantoplasty performed by two different instrument sequences regarding material loss, surface roughness, and fracture load resistance. Additionally, operators' subjective experience during instrumentation and the damage to neighboring teeth were evaluated.
MATERIALS AND METHODS: Titanium implants were placed in the position of both first maxillary molars in models exposing 6 mm of their surface. Implantoplasty was performed in phantom heads: Exposed surfaces were instrumented with diamonds and Arkansas stones or abrasive stones and silicone polishers. Operators reported on abrasion, gloss, effectiveness, and tactility using a visual analog scale (VAS). Residual wall thickness of implants was measured on radiographs, material abrasion using three-dimensional (3D) scans, and surface roughness by contact profilometry. Maximum bending moments were measured.
RESULTS: Residual thickness and weight loss were comparable after both treatments (0.3 ± 0.1 and 0.25 ± 0.07 mm and 0.22 ± 0.01 g, and 0.03 ± 0.01 mm and 0.02 ± 0.01 g, respectively, P > .05). Mean surface roughness was lower (P = .0001) for the group with the silicone polishers (0.4 ± 0.2 μm) compared with the group employing diamonds (0.8 ± 0.1 μm). Maximum bending moments showed neither intergroup differences nor stability loss compared with untreated implants. The stone-and-silicone polisher group showed less abrasion (4.6 ± 2.2) and higher gloss values (8.1 ± 1.4) than the diamond-and-Arkansas group (3.1 ± 1.3 and 4.1 ± 2.1, respectively). Superficial tooth injuries at proximal neighbor teeth were common (73% and 80%).
CONCLUSION: Implantoplasty did not weaken implant stability. The use of silicone polishers revealed lower surface roughness. Regarding surface smoothness, the instrumentation sequence employing silicon carbide and Arkansas stones followed by silicone polishers seems to be superior to the combination of diamond and Arkansas stones.

Entities:  

Mesh:

Substances:

Year:  2019        PMID: 31711072     DOI: 10.11607/jomi.7410

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  3 in total

1.  Electrospun Bioresorbable Membrane Eluting Chlorhexidine for Dental Implants.

Authors:  Pierre Pouponneau; Ophélie Perrey; Céline Brunon; Carol Grossiord; Nicolas Courtois; Vincent Salles; Antoine Alves
Journal:  Polymers (Basel)       Date:  2020-01-02       Impact factor: 4.329

2.  Does implantoplasty affect the failure strength of narrow and regular diameter implants? A laboratory study.

Authors:  Kristina Bertl; Flemming Isidor; Per Vult von Steyern; Andreas Stavropoulos
Journal:  Clin Oral Investig       Date:  2020-09-07       Impact factor: 3.573

3.  Implantoplasty Improves Clinical Parameters over a 2-Year Follow-Up: A Case Series.

Authors:  Orlando Martins; Philipp Sahrmann; João Ramos; Francisco Caramelo; Sérgio Matos; Isabel Poiares Baptista
Journal:  Medicina (Kaunas)       Date:  2022-01-12       Impact factor: 2.430

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.