A Maleb1, O Bouayadi2, J El Malki2, S Rifai2, S Lamrabat2, E Benaissa3, Y Ben Lahlou3, M Frikh3, M Elouennass3.
Abstract
We evaluated the body fluid module on Sysmex UF-1000i (UF-1000i-BF) for analysis of white blood cell (WBC) and red blood cell (RBC) in cerebrospinal fluid. We collected 93 cerebrospinal fluid samples and compared the results of the UF-1000i-BF mode with the Fast-Read 102 disposable counting cell. Results shows a good correlation between the UF-1000i and the microscopic examination. The concordance percentage is 99.06% for white blood cells and 85.18% for red blood cells. The UF-1000i-BF mode offers rapid and reliable total WBC and RBC counts for initial screening of cerebrospinal fluid, and can improve the workflow in a routine laboratory.Keywords: Body fluid; Cytology; Red blood cell; White blood cell
Year: 2019 PMID: 31704087 DOI: 10.1016/j.ab.2019.113501