Measuring diaphragm movement and respiratory frequency using a novel ultrasound device in healthy volunteers.

Håvard Andreassen Sæverud1, Ragnhild Sørum Falk2, Adam Dowrick3, Morten Eriksen3, Sigurd Aarrestad4,5, Ole Henning Skjønsberg4,6.   

Abstract

PURPOSE: To evaluate the ability of a novel ultrasound (US) device, DiaMon, to monitor diaphragm movement via its proxy liver movement, and compare it with the respired flow measured with a flowmeter, in awake and healthy volunteers. We wanted to (1) establish the optimal anatomical position for attaching the DiaMon device to the abdominal wall, and (2) evaluate the accuracy of continuous monitoring of respiratory frequency.
METHODS: Thirty healthy subjects were recruited. The DiaMon probe was applied subcostally in four different positions with the subjects in five different postures. The subjects breathed tidal volumes into a spirometer for 30-60 s with the DiaMon recording simultaneously.
RESULTS: The device detected a readable signal in 83-100% of the position/posture-combinations. The technical correlation between the two signals was highest in the anterior axillary-supine position (mean ± SD: 0.95 ± 0.03), followed by paramidline-supine (0.90 ± 0.09) and midclavicular-supine (0.89 ± 0.12). The frequency measurements yielded a mean difference of 0.03 (95% limits of agreement - 0.11, 0.16) breaths per minute in the anterior axillary-supine position.
CONCLUSION: The DiaMon device is able to detect liver movement in most subjects, and it measures breathing frequency accurately.

Entities:  

Keywords:  Diaphragm; Non-invasive; Respiration; Ultrasound

Mesh:

Year:  2019        PMID: 31691921      PMCID: PMC7925735          DOI: 10.1007/s40477-019-00412-2

Source DB:  PubMed          Journal:  J Ultrasound        ISSN: 1876-7931


  35 in total

1.  Respiratory kinematics of the upper abdominal organs: a quantitative study.

Authors:  H W Korin; R L Ehman; S J Riederer; J P Felmlee; R C Grimm
Journal:  Magn Reson Med       Date:  1992-01       Impact factor: 4.668

2.  Importance of vital signs to the early diagnosis and severity of sepsis: association between vital signs and sequential organ failure assessment score in patients with sepsis.

Authors:  Tsuneaki Kenzaka; Masanobu Okayama; Shigehiro Kuroki; Miho Fukui; Shinsuke Yahata; Hiroki Hayashi; Akihito Kitao; Daisuke Sugiyama; Eiji Kajii; Masayoshi Hashimoto
Journal:  Intern Med       Date:  2012-04-15       Impact factor: 1.271

3.  Quantification of diaphragm function using ultrasound: evaluation of a novel technique.

Authors:  Thomas L Gethin-Jones; Vicki E Noble; Christopher R Morse
Journal:  Ultrasound Med Biol       Date:  2010-09-26       Impact factor: 2.998

4.  Diaphragm thickening during inspiration.

Authors:  D Cohn; J O Benditt; S Eveloff; F D McCool
Journal:  J Appl Physiol (1985)       Date:  1997-07

5.  Ultrasound evaluation of diaphragm function in mechanically ventilated patients: comparison to phrenic stimulation and prognostic implications.

Authors:  Bruno-Pierre Dubé; Martin Dres; Julien Mayaux; Suela Demiri; Thomas Similowski; Alexandre Demoule
Journal:  Thorax       Date:  2017-03-30       Impact factor: 9.139

6.  Respiration can be monitored by photoplethysmography with high sensitivity and specificity regardless of anaesthesia and ventilatory mode.

Authors:  L Nilsson; A Johansson; S Kalman
Journal:  Acta Anaesthesiol Scand       Date:  2005-09       Impact factor: 2.105

7.  The risk factors for hospital-acquired pneumonia in the Intensive Care Unit.

Authors:  Marta Wałaszek; Alicja Kosiarska; Agnieszka Gniadek; Małgorzata Kołpa; Zdzisław Wolak; Wiesław Dobroś; Jolanta Siadek
Journal:  Przegl Epidemiol       Date:  2016

8.  Is diaphragm motion a good surrogate for liver tumor motion?

Authors:  Juan Yang; Jing Cai; Hongjun Wang; Zheng Chang; Brian G Czito; Mustafa R Bashir; Manisha Palta; Fang-Fang Yin
Journal:  Int J Radiat Oncol Biol Phys       Date:  2014-09-12       Impact factor: 7.038

9.  Can diaphragmatic ultrasonography performed during the T-tube trial predict weaning failure? The role of diaphragmatic rapid shallow breathing index.

Authors:  Savino Spadaro; Salvatore Grasso; Tommaso Mauri; Francesca Dalla Corte; Valentina Alvisi; Riccardo Ragazzi; Valentina Cricca; Giulia Biondi; Rossella Di Mussi; Elisabetta Marangoni; Carlo Alberto Volta
Journal:  Crit Care       Date:  2016-09-28       Impact factor: 9.097

Review 10.  Improving detection of patient deterioration in the general hospital ward environment.

Authors:  Jean-Louis Vincent; Sharon Einav; Rupert Pearse; Samir Jaber; Peter Kranke; Frank J Overdyk; David K Whitaker; Federico Gordo; Albert Dahan; Andreas Hoeft
Journal:  Eur J Anaesthesiol       Date:  2018-05       Impact factor: 4.330

View more
  1 in total

1.  Ultrasound variations of diaphragm activity between prone position versus supine position in ventilated patients: a cross-sectional comparative study.

Authors:  Ahlem Trifi; Sami Abdellatif; Cyrine Abdennebi; Foued Daly; Yosr Touil; Salah Ben Lakhal
Journal:  J Ultrasound       Date:  2020-09-01
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.