| Literature DB >> 31691368 |
Jingxian Wang1,2,3, Zhifeng Huang2, Mingshan Xue2, Huimin Huang2, Xiaomao Zheng2, Nanshan Zhong1,2, Baoqing Sun2.
Abstract
BACKGROUND: The sputum saccharide chain antigen (Krebs von den Lungen-6 [KL-6]) is a serum biomarker of lung injury. We aimed to evaluate the clinical performance of the automated immunoassay analyzer HISCL-5000 in detecting KL-6 by comparing it with LUMIPULSE G1200 and determine the diagnostic value of KL-6 in interstitial lung disease (ILD).Entities:
Keywords: HISCL-5000; KL-6; LUMIPULSE G1200; agreement; comparison; interstitial lung disease
Year: 2019 PMID: 31691368 PMCID: PMC7031560 DOI: 10.1002/jcla.23070
Source DB: PubMed Journal: J Clin Lab Anal ISSN: 0887-8013 Impact factor: 2.352
Patients’ Demographic Characteristics
| Characteristic | No. (%) |
|---|---|
| Sample size | 145 |
| Sex | |
| Male | 83 (57.24%) |
| Female | 62 (42.76%) |
| Age | |
| Median (interquartile range) | 56 (46‐67) |
| Range | 7‐88 |
| Diagnosis | |
| Lung cancer | 25 (17.24%) |
| Interstitial lung disease | 56 (38.62%) |
| Other lung disease | 35 (24.14%) |
| Control group | 29 (20.00%) |
Since the patients' age distribution is non‐normal, expressed in interquartile range.
Evaluation of the consistency of KL‐6 detected by LUMIPULSE G1200 analyzer and HISCL‐5000 analyzer
| SYSTEM A (ng/mL) | CO | SE | SP | PPV | NPV | Kappa (95% CI) |
| ICC | |||
|---|---|---|---|---|---|---|---|---|---|---|---|
| ≤500 | >500 | ||||||||||
| SYSTEM B (ng/mL) | ≤500 | 72 | 0 | 99.20% | 100.00% | 98.63% | 98.11% | 100.00% |
0.984 (0.952‐1.000) | .991 | 0.984 |
| >500 | 1 | 52 | |||||||||
The consistency of the SYSTEM B was evaluated using the SYSTEM A as a reference method. And 500 U/mL is used as the cutoff value of the KL‐6.
Abbreviations: CO, consistency; ICC, intraclass correlation coefficient; NPV, negative predictive value; PPV, positive predictive value; rS, Spearman's rho; SE, sensitivity; SP, specificity; SYSTEM A, LUMIPULSE G1200 analyzer; SYSTEM B, HISCL‐5000 analyzer.
Correlation is significant at the 0.01 level (2‐tailed).
Figure 1A, A scatter plot and (B) Bland‐Altman plot based on the KL‐6 titers detected by two systems. SYSTEM A, LUMIPULSE G1200 analyzer; SYSTEM B, HISCL‐5000 analyzer. “R 2” represents the linear coefficient of the fitting curve of the two indexes, and “r S” represents the correlation coefficient of the Spearman correlation analysis. The lines in the scatter plot are the cutoff values of the two systems. In the (B) figure, the black line represents the zero line, red dashed line represents the average difference value, and the upper and lower two green dashed lines represent 95% limits of agreement
The clinical diagnostic performance of KL‐6 detected by LUMIPULSE G1200 analyzer and HISCL‐5000 analyzer
| ILD | CO | SE | SP | PPV | NPV | Kappa (95% CI) | AUC (95% CI) | |||
|---|---|---|---|---|---|---|---|---|---|---|
| Positive | Negative | |||||||||
| SYSTEM A (ng/ml) | ≤500 | 62 | 11 | 85.60% | 80.36% | 89.86% | 86.54% | 84.93% |
0.707 (0.582‐0.832) |
0.901 (0.835‐0.947) |
| >500 | 7 | 45 | ||||||||
| SYSTEM B (ng/ml) | ≤500 | 80 | 11 | 86.21% | 80.36% | 89.89% | 83.33% | 87.91% |
0.707 (0.588‐0.826) |
0.888 (0.820‐0.938) |
| >500 | 9 | 45 | ||||||||
Analyze the clinical diagnostic performance of the two analyzers with the exact diagnosis of the disease as a gold standard and compare the differences between the two analyzers.
Abbreviations: AUC, area under the curve; CO, consistency; ILD, interstitial lung disease; NPV, negative predictive value; PPV, positive predictive value; SE, sensitivity; SP, specificity; SYSTEM A, LUMIPULSE G1200 analyzer; SYSTEM B, HISCL‐5000 analyzer.
Figure 2Relative operating characteristic curve. SYSTEM A, LUMIPULSE G1200 analyzer; SYSTEM B, HISCL‐5000 analyzer; AUC, area under the curve
Figure 3Serum KL‐6 level in the ILD and non‐ILD groups. ILD, interstitial lung disease; SYSTEM A, LUMIPULSE G1200 analyzer; SYSTEM B, HISCL‐5000 analyzer. Nonparametric quantitative data were presented as median (interquartile range). P‐values were calculated using Mann‐Whitney U test between two groups
Figure 4Serum KL‐6 levels in the lung cancer, ILD, other respiratory disease, and healthy control groups. ILD, interstitial lung disease; A, SYSTEM A, LUMIPULSE G1200 analyzer; B, SYSTEM B, HISCL‐5000 analyzer. Nonparametric quantitative data were presented as median (interquartile range). P‐values were calculated using Kruskal‐Wallis H test among the four groups