Madeline Gaynor1, Abbey Sawyer1,2,3, Sue Jenkins1,2,3, Jamie Wood1,2,3.
Abstract
In people with cystic fibrosis (CF), greater cardiorespiratory fitness is associated with improved survival and quality of life. Wearable activity monitors are a popular method of monitoring exercise, with measures of heart rate used to indicate exercise intensity. We assessed the agreement of heart rate recordings obtained using the Fitbit Charge HR™, Polar® H7 heart rate sensor and Masimo SET® Rad-5v pulse oximeter with the three-lead ECG during continuous and interval exercise. Adults with CF completed two exercise sessions, of 15-min duration per session, on a cycle ergometer while wearing the previously mentioned monitors. Firstly, participants cycled at 30% of estimated peak workload (W peak). Secondly, participants cycled at 1-min intervals at 60% of W peak interspersed with 2 min of unloaded cycling. Heart rate readings on all devices were recorded at minute intervals and their agreement was analysed using the Bland-Altman method. The Polar® H7 heart rate sensor had the best agreement with three-lead ECG, with a bias of 0±1 bpm during both continuous and interval exercise. The Masimo SET® Rad-5v pulse oximeter had good agreement, with a bias of 1±7 bpm and 1±5 bpm during continuous and interval exercise, respectively. The Fitbit Charge HR™ demonstrated less agreement, with a bias of 9±17 bpm and 5±13 bpm during continuous and interval exercise, respectively. The Fitbit Charge HR™ is not recommended for assessing heart rate during exercise in adults with CF. Findings support the use of the Polar® H7 for accurate heart rate monitoring.Entities:
Year: 2019 PMID: 31687372 PMCID: PMC6819990 DOI: 10.1183/23120541.00006-2019
Source DB: PubMed Journal: ERJ Open Res ISSN: 2312-0541
Participant characteristics
| 11:4 | |
| 34±11.20 (19–57) | |
| 23.6±3.2 (19.4–31.1) | |
| 2.29±0.87 (0.87–3.83) | |
| 59±22 (32–103) | |
| 3.57±0.81 (2.18–5.30) | |
| 77±14 (57–106) | |
| 80 |
Data are presented as mean±sd (range), unless otherwise stated. n=15. BMI: body mass index; FEV1: forced expiratory volume in 1 s; FVC: forced vital capacity.
FIGURE 1Bland–Altman plots depicting heart rate measurements of all participants from each monitor compared to three-lead ECG during continuous exercise. a) Polar® H7 heart rate sensor; b) Fitbit Charge HR™; c) Masimo SET® Rad-5v pulse oximeter. HR: heart rate; bpm: beats per minute; LoA: 95% limits of agreement.
FIGURE 2Bland–Altman plots depicting heart rate measurements of all participants from each monitor compared to three-lead ECG during interval exercise. a) Polar® H7 heart rate sensor; b) Fitbit Charge HR™; c) Masimo SET® Rad-5v pulse oximeter. HR: heart rate; bpm: beats per minute; LoA: 95% limits of agreement.
FIGURE 3Mean percentage of maximum heart rate of all participants from each monitor during a) continuous and b) interval exercise. HRmax: maximum heart rate.