Literature DB >> 31683473

V Flow technology in measurement of wall shear stress of common carotid arteries in healthy adults: Feasibility and normal values.

Yijie Qiu1, Daohui Yang1, Qi Zhang1, Kailing Chen1, Yi Dong1, Wen-Ping Wang1.   

Abstract

OBJECTIVE: To evaluate the feasibility of vector flow imaging technique (V Flow) in measurement of wall shear stress (WSS) of common carotid arteries (CCA) in healthy adults and to provide the normal WSS values assessed by V Flow. METHODS & MATERIALS: This prospective study was approved by the Ethics Committee of our University. Eighty healthy adult volunteers were included (mean age 43.3 y, 47 females, 33 males). The volunteers were classified into three groups according to their age: group I (age 20 - 39 y), group II (age 40 - 59 y) and group III (age 60 - 80 y). Mindray Resona 8 ultrasound machine and a linear array transducer (3-9 MHz) was used, equipped with the updated V Flow function. Common carotid arteries of both sides were evaluated in three segments (initial segment, middle segment and near bifurcation segment). The WSS values of CCA were measured by two independent radiologists. The intraclass correlation coefficient (ICC) of observer reliability in WSS measurement was calculated. Inter-observer reproducibility was also evaluated with the 95% Bland-Altman limits of agreement (LOA).
RESULTS: V Flow measurements were performed successfully in 79 volunteers (98.8 %, 79/80). The mean value of WSS in right CCA was (0.66±0.24) Pa, in left CCA was (0.66±0.18) Pa (P > 0.05). Mean WSS value had a moderately negative correlation with age group (P < 0.05). The mean WSS value of group I(mean±SD, 0.75±0.25 Pa) is larger than group II (mean±SD, 0.62±0.13 Pa) and group III (mean±SD, 0.49±0.11 Pa) (P < 0.05). The ICC of observer reliability of group I, II and III was 0.96 (95% confidence interval (95% CI) 0.92-0.98), 0.94 (95% CI 0.88-0.97), 0.93 (95% CI 0.76-0.98) respectively. The Bland-Altman plots showed that the 95% LOA were -0.17-0.12 (Pa) for group I, -0.09-0.13 (Pa) for group II and -0.08-0.10 (Pa) for group III.
CONCLUSION: V Flow measurement is a simple, rapid and feasible imaging method for the WSS assessment of CCA in healthy volunteers, which will probably be an important tool for assessing CCA function.

Entities:  

Keywords:  Vector flow imaging (V Flow); common carotid artery zzm321990(CCA); feasibility; normal value; wall shear stress (WSS)

Mesh:

Year:  2020        PMID: 31683473     DOI: 10.3233/CH-190719

Source DB:  PubMed          Journal:  Clin Hemorheol Microcirc        ISSN: 1386-0291            Impact factor:   2.375


  3 in total

1.  Wall Shear Stress Measurements Based on Ultrasound Vector Flow Imaging: Theoretical Studies and Clinical Examples.

Authors:  Yigang Du; Alfredo Goddi; Chandra Bortolotto; Yingying Shen; Alex Dell'Era; Fabrizio Calliada; Lei Zhu
Journal:  J Ultrasound Med       Date:  2020-03-03       Impact factor: 2.153

2.  Quantitative Evaluation of Post-stenotic Blood Flow Disturbance in Canine Femoral Artery Stenosis Model: An Early Experience With Vector Flow Imaging.

Authors:  Rui Zhao; Haining Zheng; Wei Wang; Yigang Du; Yisha Tong; Chaoyang Wen
Journal:  Front Cardiovasc Med       Date:  2022-02-24

Review 3.  Advances in vascular anatomy and pathophysiology using high resolution and multiparametric sonography.

Authors:  Petrucci Ilaria; Meola Mario; Fiorina Ilaria
Journal:  J Vasc Access       Date:  2021-08-02       Impact factor: 2.283

  3 in total