Literature DB >> 31679157

Beam modeling and beam model commissioning for Monte Carlo dose calculation-based radiation therapy treatment planning: Report of AAPM Task Group 157.

Chang Ming Charlie Ma1, Indrin J Chetty2, Jun Deng3, Bruce Faddegon4, Steve B Jiang5, Jinsheng Li6, Jan Seuntjens7, Jeffrey V Siebers8, Erik Traneus9.   

Abstract

Dose calculation plays an important role in the accuracy of radiotherapy treatment planning and beam delivery. The Monte Carlo (MC) method is capable of achieving the highest accuracy in radiotherapy dose calculation and has been implemented in many commercial systems for radiotherapy treatment planning. The objective of this task group was to assist clinical physicists with the potentially complex task of acceptance testing and commissioning MC-based treatment planning systems (TPS) for photon and electron beam dose calculations. This report provides an overview on the general approach of clinical implementation and testing of MC-based TPS with a specific focus on models of clinical photon and electron beams. Different types of beam models are described including those that utilize MC simulation of the treatment head and those that rely on analytical methods and measurements. The trade-off between accuracy and efficiency in the various source-modeling approaches is discussed together with guidelines for acceptance testing of MC-based TPS from the clinical standpoint. Specific recommendations are given on methods and practical procedures to commission clinical beam models for MC-based TPS.
© 2019 American Association of Physicists in Medicine.

Keywords:  Monte Carlo dose calculation; acceptance testing; beam model commissioning; beam modeling; radiation therapy; treatment planning

Mesh:

Year:  2019        PMID: 31679157     DOI: 10.1002/mp.13898

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Determination and validation of the initial beam parameters of Elekta Agility collimator head by Monte Carlo simulations.

Authors:  Bhagat Chand; Ranjit Singh; Mukesh Kumar
Journal:  Phys Eng Sci Med       Date:  2022-07-18

2.  Commissioning of and preliminary experience with a new fully integrated computed tomography linac.

Authors:  Lei Yu; Jun Zhao; Zhen Zhang; Jiazhou Wang; Weigang Hu
Journal:  J Appl Clin Med Phys       Date:  2021-06-20       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.