Antegrade versus retrograde screw fixation of anterior column acetabular fractures: a biomechanical in vitro study.

Georg Osterhoff1, Dag Wulsten2, Satish Babu3, Mark Heyland2, Carlotta Pari4.   

Abstract

BACKGROUND: To compare the mechanical strength of antegrade versus retrograde lag screw fixation of anterior column acetabular fractures.
METHODS: Standardised anterior column fractures were created in synthetic pelvis models and stabilised by either antegrade (ANTE, n = 4) or retrograde (RETRO, n = 4) anterior column screw fixation. In a validated setup, a cyclic loading protocol was applied with increasing axial force (750 cycles, 250-750 N) followed by load to failure. Construct survival, energy absorbed, construct stiffness, and load to failure were assessed. Descriptive and opto-metric methods were used to describe the mode of failure.
RESULTS: All constructs failed with loads below 1500 N. With regard to energy absorbed until failure, the ANTE group resisted to 3.763 × 105 N*cycles (range 3.760 × 105-3.763 × 105) and the RETRO group to 3.762 × 105 N*cycles (range 3.761 × 105-3.765 × 105; p = 1.0). The load to failure was 1254 N (range 977-1299) in the ANTE group and 1234 N (range 1087-1456) in the RETRO group (p = 1.0). Construct stiffness with 250 N was not different between the two groups (ANTE 192 N/mm vs. RETRO 215 N/mm, p = 0.486). In all samples, the mode of failure was a transiliac fracture with screw breakout due to rotation of the pubic fragment around the axis of the screw with a range of rotational motion [ROM] during cyclic testing of 0.96° in one ANTE sample and 0.82° in one RETRO sample for 750 N, and ROM at failure of 2.53° in one ANTE sample and 2.23° in one RETRO sample. There was some plastic deformation of the screws in all cases but no breakage.
CONCLUSIONS: In this in vitro mechanical study, antegrade screw fixation of an anterior column acetabular fracture was not different in construct survival, load to failure, stiffness, and mode of failure when compared to retrograde screw fixation.

Entities:  

Keywords:  Acetabular fracture; Acetabulum; Biomechanical; Osteoporosis; Pelvis; Screw fixation

Year:  2019        PMID: 31664465     DOI: 10.1007/s00068-019-01255-6

Source DB:  PubMed          Journal:  Eur J Trauma Emerg Surg        ISSN: 1863-9933            Impact factor:   3.693


  2 in total

Review 1.  Biomechanical analysis of fixation methods in acetabular fractures: a systematic review of test setups.

Authors:  Nico Hinz; Julius Dehoust; Matthias Münch; Klaus Seide; Tobias Barth; Arndt-Peter Schulz; Karl-Heinz Frosch; Maximilian J Hartel
Journal:  Eur J Trauma Emerg Surg       Date:  2022-03-19       Impact factor: 2.374

2.  Focus on acetabular fractures.

Authors:  Pol M Rommens; Johannes D Bastian
Journal:  Eur J Trauma Emerg Surg       Date:  2021-10       Impact factor: 3.693

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.