BACKGROUND: Many nuclear medicine departments have equipped themselves with automated dispensing systems (ADS) for PET radiopharmaceuticals, in both the operators' and the patients' interests. Whether partially or fully automated, to date there is no marketed ADS representing a true "closed-system". Despite the sterile, injectable nature of ready-to-use radiopharmaceutical drug solutions manipulated by these systems, neither manufacturer's recommendation nor literature report was found about specific qualification of the process' sterility, or about operators' qualification on these dispensing systems. We set up a master plan validation in a radiopharmacy equipped with Trasis Unidose®, including: 1) monthly process-simulating media-fill tests and microbiological contamination assessments of the ADS surfaces; 2) initial and periodic qualification of the operators. The microbiological qualification consisted in surface biocontamination assessment on critical zones with Tryptic-Soy agar. The operator qualification consisted in the evaluation of the operators' knowledge and skills for using the ADS.Entities:
Keywords: Dispensing system; PET; Qualification; Radiopharmaceuticals
Year: 2019 PMID: 31659501 PMCID: PMC6565780 DOI: 10.1186/s41181-019-0061-8
Source DB: PubMed Journal: EJNMMI Radiopharm Chem ISSN: 2365-421X
Fig. 1Datasheet for microbiological qualification of Unidose®
Fig. 2Initial and periodic qualification scale for operators of Unidose®
Fig. 3The importance of an adequate biodecontamination on microbiological qualification. Microbiological qualification results on 3 areas of interest: the carousel with ISO4.8 specifications (a), the reservoir surface (b) and the gloves (c) both with ISO7 specifications, before and after biodecontamination (BD). *P < 0.05, n = 7; two-tailed Wilcoxon matched-pairs signed rank test.
Initial Trasis Unidose® IQ/OQ/PQ
| IQ | OQ | PQ |
|---|---|---|
• Setup check-list • Network configuration • Trainings | • Labels data consistency • Air bubble size • Cartridge closure | • Requested activity vs filled activity precision • Internal dose calibrators • precision vs external calibrated dose calibrator • Particular qualification • Microbiological qualification (externalized) • External dose rate • PC/Unidose bridge |
Legend: This table provides summarized data about installation qualification (IQ), operational qualification (OQ) and performance qualification (PQ) initially performed by the supplier.