Prevalence of Clinically Significant Differences in Sodium Measurements Due to Abnormal Protein Concentrations Using an Indirect Ion-Selective Electrode Method.

Waddah Katrangi1, Nikola A Baumann1, Ryan C Nett1, Brad S Karon1, Darci R Block2.   

Abstract

BACKGROUND: Indirect ion-selective electrode (ISE) is the primary method used to measure sodium in automated clinical laboratories and is susceptible to the electrolyte exclusion effect. Pseudohyponatremia due to hyperproteinemia can affect patient management. The aims of this study were to (a) establish the relationship between serum total protein (TP) concentration and the magnitude of the electrolyte exclusion effect on indirect ISE-measured sodium values (b) estimate the frequency at which TP concentrations outside the reference interval may impact indirect-ISE measured sodium values, and (c) determine whether clinical decision support (middleware) rules in the laboratory would be effective for detecting cases of pseudohyponatremia.
METHODS: Residual waste serum specimens from physician-ordered TP testing were collected (n = 112). Sodium concentration was measured using indirect ISE (Cobas 8000, Roche Diagnostics) and direct ISE (ABL 825, Radiometer) methods. The difference in sodium concentration (Δ[Na+]) was calculated as follows: ([Na+]indirect-ISE - [Na+]direct-ISE). Retrospective TP results reported from July 31, 2013, to September 24, 2014, were stratified by ordering location and sodium and TP co-ordering rates were quantified.
RESULTS: Δ[Na+] was inversely proportional to TP concentration (y = -1.29x + 8.6, R = -0.883). The average difference (SD, range) was -6.1(3.4, -16-0) mmol/L when TP >7.9 g/dL (>79g/L), with 69% of samples demonstrating differences ≥4.0 mmol/L. A majority of intensive care unit patients (81%) were hypoproteinemic (<6.3 g/dL, <63g/L). Only 10.9% of sodium test orders include an order for TP on the same collection.
CONCLUSIONS: Indirect sodium measurement is impacted when TP concentrations are increased. TP concentration outside the reference interval is prevalent and sodium is usually not ordered with TP. Health systems need to be aware of the limitations of their indirect-ISE method for sodium measurement.
© 2019 American Association for Clinical Chemistry.

Entities:  

Year:  2019        PMID: 31659081     DOI: 10.1373/jalm.2018.028720

Source DB:  PubMed          Journal:  J Appl Lab Med        ISSN: 2475-7241


  1 in total

1.  Clinical factors within a week of birth influencing sodium level difference between an arterial blood gas analyzer and an autoanalyzer in VLBWIs: A retrospective study.

Authors:  Hyun Ho Kim; Jin Kyu Kim
Journal:  Medicine (Baltimore)       Date:  2021-12-10       Impact factor: 1.817

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.