Literature DB >> 31658505

Implementation and study of dopant-assisted photoionization with a miniature capillary inlet ion trap mass spectrometer.

Qian Zhang1,2, Yuan Tian1, Mushage Aliang1, Quan Yu1, Xiaohao Wang1,2.   

Abstract

RATIONALE: Dopant-assisted photoionization (PI) has been widely used in the mass spectrometric analysis of volatile compounds. Exploring simple doping methods will benefit parameter optimization and promote the application of this technique.
METHODS: A previously built miniature ion trap mass spectrometer was used to study dopant-assisted vacuum PI. The sampling system of this device was modified to provide three inlets for the simultaneous introduction of analytes, dopants, and auxiliary air. Then, dopant solution was directly injected into the ion trap chamber through a self-aspirating capillary inlet and rapidly evaporated without heating. Various dopant solutions were prepared and switched during the experiments.
RESULTS: When analyzing some aniline compounds, the signals of all analytes were improved by more than 10 times after the injection of 2% anisole solution as a dopant. In addition, anisole can provide analyte signals more than three times stronger than those provided by the other dopants. On the basis of the ionization energy selectivity of dopant-assisted PI, some isomers in the mass spectrometric analysis were distinguished using different additives.
CONCLUSIONS: In general, liquid doping is as feasible and as effective as other traditional methods, and using appropriate dopants with high PI efficiency or feeding more dopants contributes to the ionization of analytes. The proposed method also offers several unique merits, such as simple operation, low consumption, and smooth switching with minimal residue.
© 2019 John Wiley & Sons, Ltd.

Entities:  

Year:  2020        PMID: 31658505     DOI: 10.1002/rcm.8621

Source DB:  PubMed          Journal:  Rapid Commun Mass Spectrom        ISSN: 0951-4198            Impact factor:   2.419


  1 in total

1.  Portable Digital Linear Ion Trap Mass Spectrometer Based on Separate-Region Corona Discharge Ionization Source for On-Site Rapid Detection of Illegal Drugs.

Authors:  Lingfeng Li; Tianyi Zhang; Deting Wang; Yunjing Zhang; Xingli He; Xiaozhi Wang; Peng Li
Journal:  Molecules       Date:  2022-05-30       Impact factor: 4.927

  1 in total