Image Quality and Activity Optimization in Oncologic 18F-FDG PET Using the Digital Biograph Vision PET/CT System.

Joyce van Sluis1, Ronald Boellaard2, Rudi A J O Dierckx2, Gilles N Stormezand2, Andor W J M Glaudemans2, Walter Noordzij2.   

Abstract

The first Biograph Vision PET/CT system (Siemens Healthineers) was installed at the University Medical Center Groningen. Improved performance of this system could allow for a reduction in activity administration or scan duration. This study evaluated the effects of reduced scan duration in oncologic 18F-FDG PET imaging on quantitative and subjective imaging parameters and its influence on clinical image interpretation.
Methods: Patients referred for a clinical PET/CT scan were enrolled in this study, received a weight-based 18F-FDG injected activity, and underwent list-mode PET acquisition at 180 s per bed position (s/bp). Acquired PET data were reconstructed using the vendor-recommended clinical reconstruction protocol (hereafter referred to as "clinical"), using the clinical protocol with additional 2-mm gaussian filtering (hereafter referred to as "clinical+G2"), and-in conformance with European Association of Nuclear Medicine Research Ltd. (EARL) specifications-using different scan durations per bed position (180, 120, 60, 30, and 10 s). Reconstructed images were quantitatively assessed for comparison of SUVs and noise. In addition, clinically reconstructed images were qualitatively evaluated by 3 nuclear medicine physicians.
Results: In total, 30 oncologic patients (22 men, 8 women; age: 48-88 y [range], 67 ± 9.6 y [mean ± SD]) received a single weight-based (3 MBq/kg) 18F-FDG injected activity (weight: 45-123 kg [range], 81 ± 15 kg [mean ± SD]; activity: 135-380 MBq [range], 241 ± 47.3 MBq [mean ± SD]). Significant differences in lesion SUVmax were found between the 180-s/bp images and the 30- and 10-s/bp images reconstructed using the clinical protocols, whereas no differences were found in lesion SUVpeak EARL-compliant images did not show differences in lesion SUVmax or SUVpeak between scan durations. Quantitative parameters showed minimal deviation (∼5%) in the 60-s/bp images. Therefore, further subjective image quality assessment was conducted using the 60-s/bp images. Qualitative assessment revealed the influence of personal preference on physicians' willingness to adopt the 60-s/bp images in clinical practice. Although quantitative PET parameters differed minimally, an increase in noise was observed.
Conclusion: With the Biograph Vision PET/CT system for oncologic 18F-FDG imaging, scan duration or activity administration could be reduced by a factor of 3 or more with the use of the clinical+G2 or the EARL-compliant reconstruction protocol.
© 2020 by the Society of Nuclear Medicine and Molecular Imaging.

Entities:  

Keywords:  PET/CT; activity optimization; image quality; scan duration; silicon photomultiplier–based detector

Year:  2019        PMID: 31628214     DOI: 10.2967/jnumed.119.234351

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  13 in total

1.  Digital versus analogue PET in [68Ga]Ga-PSMA-11 PET/CT for recurrent prostate cancer: a matched-pair comparison.

Authors:  Ian Alberts; George Prenosil; Christos Sachpekidis; Thilo Weitzel; Kuangyu Shi; Axel Rominger; Ali Afshar-Oromieh
Journal:  Eur J Nucl Med Mol Imaging       Date:  2019-12-02       Impact factor: 9.236

Review 2.  Kinetic modeling and parametric imaging with dynamic PET for oncological applications: general considerations, current clinical applications, and future perspectives.

Authors:  Antonia Dimitrakopoulou-Strauss; Leyun Pan; Christos Sachpekidis
Journal:  Eur J Nucl Med Mol Imaging       Date:  2020-05-19       Impact factor: 9.236

3.  Digital PET/CT allows for shorter acquisition protocols or reduced radiopharmaceutical dose in [18F]-FDG PET/CT.

Authors:  Ian Alberts; Christos Sachpekidis; George Prenosil; Marco Viscione; Karl Peter Bohn; Clemens Mingels; Kuangyu Shi; Ali Ashar-Oromieh; Axel Rominger
Journal:  Ann Nucl Med       Date:  2021-02-07       Impact factor: 2.668

4.  Evaluation of [68Ga]Ga-PSMA PET/CT images acquired with a reduced scan time duration in prostate cancer patients using the digital biograph vision.

Authors:  Manuel Weber; Walter Jentzen; Regina Hofferber; Ken Herrmann; Wolfgang Peter Fendler; Maurizio Conti; Axel Wetter; David Kersting; Christoph Rischpler; Pedro Fragoso Costa
Journal:  EJNMMI Res       Date:  2021-02-28       Impact factor: 3.138

5.  The role of activity, scan duration and patient's body mass index in the optimization of FDG imaging protocols on a TOF-PET/CT scanner.

Authors:  Roberta Matheoud; Naema Al-Maymani; Alessia Oldani; Gian Mauro Sacchetti; Marco Brambilla; Alessandro Carriero
Journal:  EJNMMI Phys       Date:  2021-04-06

6.  EARL compliance measurements on the biograph vision Quadra PET/CT system with a long axial field of view.

Authors:  George A Prenosil; Michael Hentschel; Thilo Weitzel; Hasan Sari; Kuangyu Shi; Ali Afshar-Oromieh; Axel Rominger
Journal:  EJNMMI Phys       Date:  2022-04-08

7.  Phantom-based image quality assessment of clinical 18F-FDG protocols in digital PET/CT and comparison to conventional PMT-based PET/CT.

Authors:  Silvano Gnesin; Christine Kieffer; Konstantinos Zeimpekis; Jean-Pierre Papazyan; Renaud Guignard; John O Prior; Francis R Verdun; Thiago V M Lima
Journal:  EJNMMI Phys       Date:  2020-01-06

8.  Clinical performance of long axial field of view PET/CT: a head-to-head intra-individual comparison of the Biograph Vision Quadra with the Biograph Vision PET/CT.

Authors:  Ian Alberts; Jan-Niklas Hünermund; George Prenosil; Clemens Mingels; Karl Peter Bohn; Marco Viscione; Hasan Sari; Bernd Vollnberg; Kuangyu Shi; Ali Afshar-Oromieh; Axel Rominger
Journal:  Eur J Nucl Med Mol Imaging       Date:  2021-04-02       Impact factor: 9.236

9.  The influence of digital PET/CT on diagnostic certainty and interrater reliability in [68Ga]Ga-PSMA-11 PET/CT for recurrent prostate cancer.

Authors:  Ian Alberts; Jan-Niklas Hünermund; Christos Sachpekidis; Clemens Mingels; Viktor Fech; Karl Peter Bohn; Axel Rominger; Ali Afshar-Oromieh
Journal:  Eur Radiol       Date:  2021-04-15       Impact factor: 5.315

10.  Reducing Radiation Exposure to Paediatric Patients Undergoing [18F]FDG-PET/CT Imaging.

Authors:  Hunor Kertész; Thomas Beyer; Kevin London; Hamda Saleh; David Chung; Ivo Rausch; Jacobo Cal-Gonzalez; Theo Kitsos; Peter L Kench
Journal:  Mol Imaging Biol       Date:  2021-04-12       Impact factor: 3.488

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.