Literature DB >> 31623924

Effects of lysosomal deposition on the friction coefficient of hydrogel contact lenses.

Chen-Ying Su1, Lung-Kun Yeh2, Chi-Chun Lai2, Kuan-Yi Li1, Ching-Li Tseng3, Hsu-Wei Fang4.   

Abstract

When a contact lens is in contact with the eyelid and the cornea, tear proteins start to be accumulated on the lens and may subsequently undergo conformational changes. Protein adsorption or conformational changes on the lens may lead to discomfort, red eye, or even inflammatory reactions. Although measuring the friction coefficient of contact lens has been linked to comfort degree in vivo, there is not much research about the effects of tear protein deposition on the friction coefficient of the contact lens. Therefore, we investigated the friction coefficient of three distinct materials of contact lenses in two different lysosomal concentrations. We also studied lysozyme deposition on the lens without the influence of friction. The results demonstrated that although the amount of lysozyme deposition was high on all the materials we tested, it was not corresponded to higher friction coefficient. In addition, we investigated the effect of the contact lens care solution we developed on friction coefficient and lysozyme deposition of three different materials of contact lenses. The results showed that the care solution could reduce the increased friction coefficient caused by high lysosomal concentration. Therefore, we proposed a potential mechanism of why lysosomal deposition may result in high friction coefficient for certain types of hydrogel contact lenses.
Copyright © 2019 British Contact Lens Association. Published by Elsevier Ltd. All rights reserved.

Keywords:  Friction coefficient; Hydrogel contact lens; Lysozyme deposition

Mesh:

Substances:

Year:  2019        PMID: 31623924     DOI: 10.1016/j.clae.2019.09.007

Source DB:  PubMed          Journal:  Cont Lens Anterior Eye        ISSN: 1367-0484            Impact factor:   3.077


  3 in total

1.  Correlation between Tribological Properties and the Quantified Structural Changes of Lysozyme on Poly (2-hydroxyethyl methacrylate) Contact Lens.

Authors:  You-Cheng Chang; Chen-Ying Su; Chia-Hua Chang; Hsu-Wei Fang; Yang Wei
Journal:  Polymers (Basel)       Date:  2020-07-23       Impact factor: 4.329

2.  The Bio-Tribological Effect of Poly-Gamma-Glutamic Acid in the Lysozyme-Ionic Contact Lens System.

Authors:  Chen-Ying Su; Lung-Kun Yeh; Chi-Chun Lai; Mihaela Dubuisson; Yi-Fei Tsao; Ching-Li Tseng; And Hsu-Wei Fang
Journal:  Polymers (Basel)       Date:  2020-01-07       Impact factor: 4.329

3.  Evaluation of the Lubricating Effect of Hyaluronic Acid on Contact Lenses Using a Pendulum-Type Friction Tester Under Mimicking Physiological Conditions.

Authors:  Hiroko Iwashita; Kiyoshi Mabuchi; Takashi Itokawa; Yukinobu Okajima; Takashi Suzuki; Yuichi Hori
Journal:  Eye Contact Lens       Date:  2022-02-01       Impact factor: 2.018

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.