Short- and medium-term follow-up of transcatheter closure of perimembranous ventricular septal defects.

Hao Li1, Youyang Shi1, Songyue Zhang1, Yue Ren1, Xing Rong1, Zhenquan Wang1, Tianhe Xia1, Tingting Wu1, Haitang Xu1, Yaoyao Yan1, Yi Zhan1, Maoping Chu2, Rongzhou Wu3.   

Abstract

BACKGROUND: Transcatheter closure of perimembranous ventricular septal defects is one of the greatest challenges in interventional cardiology. Short- and midium-term follow-up data for large samples are limited. This report presents our experience with transcatheter closure of perimembranous ventricular septal defects using an occluder.
METHODS: Two hundred fifty-three patients included in the database of the Second Affiliated Hospital and Yuying Children's Hospital from January 2011- December 2015 with transcatheter closure of perimembranous ventricular septal defects and discharged from follow-up. All patients were invited for clinical and transthoracic echocardiography, electrocardiogram, and thoracic radiography check-up.
RESULTS: Device implantation was successful in 252 of 253 patients (99.6%). The median age was 42 months (range 27-216 months). The median follow-up duration was 36 months (range 6-60 months). The mean defect diameter was 3.5 ± 1.4 mm and the mean size of the ventricular septal defect rim below the aortic valve was 3.7 ± 1.8 mm. The mean diameter of the devices used was 4 mm. Thirty-seven patients developed arrhythmia after the procedure and recovered within 24 months; four patients had hemolysis and four had moderate tricuspid valve regurgitation. No other serious adverse event occurred during the follow-up period.
CONCLUSION: Transcatheter closure of perimembranous ventricular septal defects using an occluder is safe and effective in most patients.

Entities:  

Keywords:  Complications; Short- and midium-term follow-up; Transcatheter closure; Ventricular septal defect

Mesh:

Year:  2019        PMID: 31619172      PMCID: PMC6794751          DOI: 10.1186/s12872-019-1188-y

Source DB:  PubMed          Journal:  BMC Cardiovasc Disord        ISSN: 1471-2261            Impact factor:   2.298


Background

The ventricular septal defect (VSD) is one of most common congenital heart defects worldwide. Perimembranous ventricular septal defect, involve a more extensive area than normal membranous septum, also known as aortic subvalvular type. The defect is surrounded by membrane, or the upper edge is fibrous and the lower edge is muscular. With advances in imaging and screening of infants, the detection rate of confirmed cases of VSD has increased notably [1]. Almost 45% of VSDs occurring in isolation close spontaneously [2]. However, for patients with medium and larger (perimembranous or muscular) defects, surgical treatment is often recommended. Admittedly, traditional surgical repair has yielded great results. Nevertheless, residual VSDs are frequently associated with complications, such as reoperation, infection, sternotomy scarring, and even death [3, 4]. In 1987, Lock [5] was the first to use the Rashkind double-umbrella occluder to close a VSD. Since then, transcatheter VSD closure has been used increasingly in selected patients worldwide. Nevertheless, this technique is associated with complications, such as embolization of the device, hemolysis, aortic insufficiency, and complete heart block. Although various centers in China have reported the use of interventional therapy for VSDs, few data on long-term results are available. We assessed 253 patients through periodic follow-up. In this study, we report the short- and midium-term follow-up results of transcatheter closure of perimembranous VSDs using an occluder in 253 selected patients at our medical center.

Methods

Study design

This retrospective, descriptive, non-randomized study utilized data from January 2011–December 2015 from a single medical center (The Second Affiliated Hospital and Yuying Children’s Hospital, Institute of Cardiovascular Development and Translational Medicine, Wenzhou Medical University, Wenzhou, Zhejiang). Two hundred fifty-three patients (126 males, 127 females) were included (Table 1, Additional file 2: Table S2). Informed consent was obtained from all patients or their parents. All patients had clinical and echocardiographic evidence of ventricular septal defect (VSD). The position and distance between each valve and the defect should be carefully assessed preoperatively using echocardiography.
Table 1

General Characteristics of Patients

253 patientsFemale 127 (50.2%)/ Male 126 (49.8%)
Age (months)Median = 42 (27–216)
 0–3672 (28%)
 36–72132 (52%)
 72–10824 (10%)
 108–14415 (6%)
 144–21610 (4%)
Weight (kg)Median = 14.75 (9.5–86)
VSD size (narrowest diameter) (mm)Mean = 3.5 ± 1.4
VSD rim to the aortic valveMean = 3.7 ± 1.8
LVEDd (mm)Mean = 35.6 ± 5.1
TR-PG (mmHg)Mean = 31.6 ± 11.6
CTRMean = 0.533 ± 0.044
VSD typePerimembranous
Multiple VSDs43 (17%)

VSD Ventricular septal defect, LVEDd left ventricular end-diastolic dimension, TR-PG tricuspid valve regurgitation pressure gradient, CTR cardiothoracic ratio

General Characteristics of Patients VSD Ventricular septal defect, LVEDd left ventricular end-diastolic dimension, TR-PG tricuspid valve regurgitation pressure gradient, CTR cardiothoracic ratio

Closure protocol

According to the guidelines for interventional therapy via catheter for congenital heart disease, all patients underwent combined intravenous anesthesia and were monitored via electrocardiogram and blood pressure and oxygen saturation measurement. The peritoneal inguinal puncture site was anesthetized locally with 2% lidocaine, and the femoral vein and femoral artery were punctured. Left and right cardiac catheterization and left ventricular and ascending aortic angiography were performed to determine the size, location, shape, and distance from the aortic valve of the defect. The femoral artery, aorta, left ventricle, VSD, right ventricle, and femoral vein were identified. A long sheath was inserted into the left ventricle along the track from the side of the femoral vein, and then fed into the appropriate sealing device in an attempt to plug the defect. Generally, the sealing device must be 1–2 mm larger than the diameter of the defect. Following closure, the patient must be monitored closely for heart murmur via auscultation, left ventricular angiography should be repeated after 15 min to check for residual shunt, and root imaging should be used to assess the presence of aortic valve regurgitation and prolapse. Following successful closure, thoracic echocardiography (TTE) should be used to monitor the occluder location and sharp, and check for residual shunt and tricuspid valve, and to assess the movement of the aortic valve.

Device details

The following occluder devices were used (Tables 2, 3, 4 and 5): the Amplatzer Membranous VSD Occluder (AGA Medical Corporation) and Amplatzer-like occluders including Membranous Symmetric VSD Occluder (SHSMA Medical Corporation, China) and Cera™ Membranous VSD (Symmetric) Occluder (ShenZhen Lifetech Scientific Inc., China).
Table 2

Embolic materials used in transcatheter embolization of VSD

Device typeCompanyNationNumber
Amplatzer Membranous VSD OccluderAGA Medical CorporationAmerican7
Membranous Symmetric VSD OccluderSHSMA Medical CorporationChina167
Cera™ Membranous VSD (Symmetric) OccluderLifetech Scientific IncChina79
Table 3

Device type, sizes and recommended sheath size of CeraTM Membranous VSD (Symmetric) Occluder

TypeD (mm)D1(mm)D2(mm)L (mm)Minimum Recommended Sheath Size
LT-VSD-Sym-0448835F
LT-VSD-Sym-0559935F
LT-VSD-Sym-066101036F
LT-VSD-Sym-077111136F
LT-VSD-Sym-088121237F
LT-VSD-Sym-1010141437F
LT-VSD-Sym-1212161639F
LT-VSD-Sym-1414191939F
LT-VSD-Sym-1616212139F
LT-VSD-Sym-18182323310F
LT-VSD-Sym-20202525310F
LT-VSD-Sym-22222727312F
LT-VSD-Sym-24242929312F

D, Waist Diameter; D1, Right Disc Diameter; D2, Left Disc Diameter; L, Waist Length

Table 4

Device type, sizes and recommended sheath size of Membranous Symmetric VSD Occluder

TypeA (mm)B (mm)C (mm)H (mm)Minimum Recommended Sheath Size
SQFDQ-II 044883.56F
SQFDQ-II 0559946F
SQFDQ-II 066101046F
SQFDQ-II 077111147F
SQFDQ-II 088121247F
SQFDQ-II 09913134.58F
SQFDQ-II 101014144.58F
SQFDQ-II 121216154.59F
SQFDQ-II 141418174.59F
SQFDQ-II 16162220510F
SQFDQ-II 18182422510F
SQFDQ-II 20202624512F

A, Waist Diameter; B, Left Disc Diameter; C, Right Disc Diameter; H, Waist Length

Table 5

Device type, sizes and recommended sheath size of Amplatzer Membranous VSD Occluder

TypeA (mm)B (mm)C (mm)Minimum Recommended Sheath Size
9-VSDMEMB-00410847F
9-VSDMEMB-00511957F
9-VSDMEMB-006121067F
9-VSDMEMB-007131177F
9-VSDMEMB-008141287F
9-VSDMEMB-009151397F
9-VSDMEMB-0101614107F
9-VSDMEMB-0111715117F
9-VSDMEMB-0121816127F
9-VSDMEMB-0131917138F
9-VSDMEMB-0142018148F
9-VSDMEMB-0152119159F
9-VSDMEMB-0162220169F
9-VSDMEMB-0172321179F
9-VSDMEMB-0182422189F

A, Left Disc Diameter; B, Right Disc Diameter; C, Waist Diameter

Embolic materials used in transcatheter embolization of VSD Device type, sizes and recommended sheath size of CeraTM Membranous VSD (Symmetric) Occluder D, Waist Diameter; D1, Right Disc Diameter; D2, Left Disc Diameter; L, Waist Length Device type, sizes and recommended sheath size of Membranous Symmetric VSD Occluder A, Waist Diameter; B, Left Disc Diameter; C, Right Disc Diameter; H, Waist Length Device type, sizes and recommended sheath size of Amplatzer Membranous VSD Occluder A, Left Disc Diameter; B, Right Disc Diameter; C, Waist Diameter

Follow-up

Immediately following the procedure, and in months 1, 3, and 6 and years 1, 2, 3, 4, and 5 of follow-up, the shape and orientation of the device and residual shunt were reevaluated through transthoracic echocardiography, electrocardiogram, and thoracic radiography.

Statistical analysis

Continuous data were expressed as means ± standard deviations with ranges and were compared using analysis of variance (ANOVA) and the student’s t-test, while categorical data are expressed as number and percentage, and were compared using the X2-test or Fisher method. All statistical analyses were performed using commercially available software (SPSS version 11.0 for Windows; SPSS Inc., Chicago, IL, USA). P values < 0.05 were considered to be significant.

Results

Patient groups

The median age at implantation was 42 months (range 27–216 months) and the median weight was 14.75 kg (range 9.5–86 kg). All patients had continuous systolic murmurs and left-to-right shunts with enlarged left ventricular diastolic dimensions. No patient received diuretics or angiotensin-converting enzyme inhibitors before the procedure for there was not any grade II or higher cardiac dysfunction after preoperative evaluation. Transcatheter femoral venous and arterial approaches were used in all patients. The mean VSD size, measured echocardiographically, was 3.5 ± 1.4 mm. Pulmonary-to-systemic flow ratio (Qp/Qs) was 2.3 ± 0.4 (range 1.5–3.0). One patient with Down syndrome underwent transcatheter closure, and a residual shunt, hemolysis, and arrhythmia was found 24 h postoperative. A surgical operation was subsequently performed. (Tables 1 and 6, Additional file 3: Table S3).
Table 6

Procedure data

VariablesValues
Procedure time (hours)Mean = 1.9 ± 0.6
Qp/QsMean = 2.3 ± 0.4
Systolic PA pressure (mmHg)Mean = 30.6 ± 7.8
 Mild pulmonary hypertension; 30–50N = 120
 Moderate pulmonary hypertension; 50–70N = 4
 Severe pulmonary hypertension; ≥70N = 1
VSD size (mm)Mean = 3.5 ± 1.4

Qp/Qs pulmonary-to-systemic flow ratio, VSD Ventricular septal defect, PA Pulmonary artery

Procedure data Qp/Qs pulmonary-to-systemic flow ratio, VSD Ventricular septal defect, PA Pulmonary artery

Safety and efficacy

In 252 of 253 (99.6%) procedures, defect closure was successful (Fig. 1). Stable device positioning was demonstrated by angiography and transthoracic echocardiography. Significant shunt reduction was detected in all 252 patients after successful VSD closure, and only minor residual shunts were documented. Procedural data are provided in Table 6. No increase in aortic or tricuspid valve regurgitation was observed postoperatively in any patient.
Fig. 1

Left ventricular angiography in preoperative and 15-min postoperative. a Left ventricular angiography showed the defect with the rim of 3.5 mm (the shunts was designated by arrow); b The 5 mm occluder closed the defect without residual shunt

Left ventricular angiography in preoperative and 15-min postoperative. a Left ventricular angiography showed the defect with the rim of 3.5 mm (the shunts was designated by arrow); b The 5 mm occluder closed the defect without residual shunt

Valve regurgitation and residual shunt

Pre-operatively, four patients had moderate tricuspid valve regurgitation and nine had mild aortic valve regurgitation. No significant increase or new regurgitation of the aortic or tricuspid valve occurred during or after the procedure. All cases of valve regurgitation had resolved without treatment at the 2-year follow-up evaluation (Fig. 2).
Fig. 2

Echocardiography: four-chamher view with VSD, from preoperative to 2 years follow-up (26-month-old patient). VSD:ventricular septal defect; occluder: the device closed the defect without residual shunt. a: preoperative(d = 3.5 mm); b: at 1 month after the procedure; c: at 3 months after the procedure; d: at 6 months after the procedure; e: at 1 year after the procedure; f: at 2 years after the procedure

Echocardiography: four-chamher view with VSD, from preoperative to 2 years follow-up (26-month-old patient). VSD:ventricular septal defect; occluder: the device closed the defect without residual shunt. a: preoperative(d = 3.5 mm); b: at 1 month after the procedure; c: at 3 months after the procedure; d: at 6 months after the procedure; e: at 1 year after the procedure; f: at 2 years after the procedure Closure was achieved with the device in all but one patient. In this case, echocardiography revealed a significant shunt at the center and edge of the occluder, and thoracic surgery was performed. Immediate complete occlusion, determine by angiography, was achieved in 38.6% of patients. Echocardiography performed 24 h postoperatively documented an occlusion rate of 86.6% (219 patients). Echocardiography confirmed the achievement of complete closure in the remaining patients, excluding the patient with a significant shunt described above, at 1 month (n = 11), 3 months (n = 3), 6 months (n = 7), 1 year (n = 1), and 2 years (n = 1) postoperatively. The occlusion rate at 2 years postoperatively was 96% (Fig. 3).
Fig. 3

Complete closure percentage immediately on angiography and 24 h, 1 month, 3 months, 6 months, 1 year and 2 years follow up by echocardiography

Complete closure percentage immediately on angiography and 24 h, 1 month, 3 months, 6 months, 1 year and 2 years follow up by echocardiography

Cardiothoracic ratio and tricuspid valve regurgitation pressure gradient

The cardiothoracic ratio decreased gradually from 1 to 24 months postoperatively, and had become stable by the 36-month follow-up evaluation (Fig. 4; Table 7). TTE was used to measure the tricuspid valve regurgitation pressure gradient (TR-PG) for the estimation of pulmonary artery pressure. A significant decrease was detected during follow-up (Table 7).
Fig. 4

Frontal CTR: from preoperative to 6 months follow-up (26-month-old patient). a preoperative; b at 1 month after the procedure; c at 3 months after the procedure; d at 6 months after the procedure

Table 7

Residual shunt, tricuspid valve pressure gradient and cardiothoracic ratio

TimeTR-PGCTR
Mean (mmHg)P-valueMeanP-value
Before closure31.0 ± 11.60.533 ± 0.044
3-day25.0 ± 6.8<  0.05
1-month20.9 ± 6.6<  0.050.531 ± 0.052<  0.05
3-month21.9 ± 4.0<  0.050.531 ± 0.047<  0.05
6-month20.4 ± 5.1<  0.050.527 ± 0.047<  0.05
12-month21.7 ± 4.7<  0.050.518 ± 0.042<  0.05
24-month22.2 ± 6.6<  0.050.514 ± 0.037<  0.05
36-month20.0 ± 2.5<  0.050.505 ± 0.048NS
48-month20.3 ± 3.6<  0.050.503 ± 0.028NS
60-month21.0 ± 4.7<  0.050.504 ± 0.035NS

TR-PG tricuspid valve regurgitation pressure gradient, CTR cardiothoracic ratio

Frontal CTR: from preoperative to 6 months follow-up (26-month-old patient). a preoperative; b at 1 month after the procedure; c at 3 months after the procedure; d at 6 months after the procedure Residual shunt, tricuspid valve pressure gradient and cardiothoracic ratio TR-PG tricuspid valve regurgitation pressure gradient, CTR cardiothoracic ratio

Complications

No death occurred. Arrhythmia was detected in 37 patients, hemolysis in 4, moderate tricuspid valve regurgitation in 4, and mild aortic valve regurgitation in 9 patients directly after implantation or during follow-up. No significant complication, such as procedure-related obstruction, cardiac tamponade, thrombosis or air embolism, aortic rupture, or increased valve failure, was observed. One significant complication occurred in a 3-year-old boy, who developed hematuria; significant shunt reduction and complete right bundle branch block were detected.

Hemolysis

No significant hemolysis necessitating device retrieval occurred. Although four (1.6%) patients showed erythrocyte positivity (+++) on routine urine tests during the first 24 h postoperatively, the hemolysis was not sufficiently severe in any case to warrant retrieval of the device. Blood analysis showed high values for hemolysis parameters with no relevant decrease in the hemoglobin concentration. Hemolysis regressed by 5 days postoperatively with hydration and urine alkalization. Blood transfusion was not required in any case.

Arrhythmia

Minor arrhythmic complications occurred in 37 (17.4%) patients: 14 patients had incomplete right bundle branch block and were treated with a conventional dose of methylprednisolone (20–30 ml/kg/d), which was gradually tapered. 8 had complete right bundle branch block, 3 had left bundle branch block, and 12 had left anterior fascicular block. All patients had recovered, as demonstrated by ECG, at the 24-month follow-up evaluation. Patients with right bundle branch blocks were treated with a conventional dose of methylprednisolone (1–2 mg/kg, q8h), which was gradually tapered. As left bundle and left anterior branch blocks have been reported to lead to death due to heart failure in China, the methylprednisolone dose can be increased to 20–30 mg/kg/d to relieve ventricular septal edema. VSD closure with an occluder was performed successfully in 252 patient. The patients were followed for a median follow-up period of 36 months (range 6–60 months). And the 6 months of follow-up with clinical examination, chest radiography, TTE and ECG was completed in 250 patients (250/253), 1 year in 236 (236/243), 2 years in 190 (190/208), 3 years in 151 (151/169), 4 years in 101 (101/116), 5 years in 32 (32/45) (Table 8). There were 23 patients had lost at follow-up for the following reasons: surgical follow-up for operation, patients from migrant workers’ families going back to their household registration, contact information been replaced. In addition to 9 patients who changed contact information, none of the rest of the patients lost had significant complications such as death or syncope, and they had so far insisted on regular annual follow-up at local children’s heart centers or comprehensive large hospitals. Pre- and postoperatively, TTE was used to detect the tricuspid valve pressure gradient for the estimation of pulmonary artery pressure. TTE showed that the left atrium was reduced compared with preoperative, and gradually increased during follow-up. The TR-PG decreased significantly during the postoperative follow-up period. TTE showed that the LVEDd, RV, and LA had decreased at 3 days postoperatively, decreased further at 1–6 months postoperatively, and had regressed at the 12-month follow-up examination (Table 9).
Table 8

Data of follow-up duration

Follow-up durationNumber of followersLoss of follow-up rate (%)
6 months250/2534.3%
1 year236/2432.8%
2 years190/2088.7%
3 years151/16910.7%
4 years101/11612.9%
5 years32/4528.9%
Table 9

Echocardiography Follow-up Datas: preoperative and postoperative

Age (months)Parameter (mm)Preoperative3d1 m3 m6 m12 m24 m36 m48 m60 m
< 36LVEDd33.4 ± 3.6*31.8 ± 3.231.6 ± 3.431.6 ± 2.131.8 ± 2.432.3 ± 2.332.4 ± 2.535.1 ± 2.036.7 ± 1.536.9 ± 2.9
RV15.0 ± 4.2*15.4 ± 3.915.2 ± 3.615.4 ± 3.516.0 ± 3.616.7 ± 4.716.5 ± 3.416.0 ± 3.418.7 ± 3.519.2 ± 1.8
LA23.0 ± 3.2*22.1 ± 3.120.9 ± 3.421.1 ± 3.221.3 ± 2.821.5 ± 3.121.3 ± 2.923.6 ± 1.923.2 ± 4.723.6 ± 2.5
EF70.2 ± 3.670.8 ± 4.570.1 ± 3.770.9 ± 3.169.2 ± 3.769.3 ± 4.769.2 ± 3.671.2 ± 3.270.0 ± 3.671.2 ± 1.2
36–72LVEDd34.7 ± 3.6*32.8 ± 3.333.0 ± 3.333.0 ± 3.233.4 ± 3.133.5 ± 3.834.9 ± 3.335.5 ± 3.836.8 ± 2.136.5 ± 1.4
RV14.4 ± 3.6*14.7 ± 3.614.7 ± 3.7.9 ± 3.315.6 ± 3.916.1 ± 4.016.3 ± 3.616.2 ± 4.516.4 ± 2.816.5 ± 1.7
LA22.3 ± 4.3*21.8 ± 3.321.3 ± 3.221.4 ± 3.121.1 ± 2.922.2 ± 3.422.2 ± 3.823.2 ± 3.824.7 ± 1.423.8 ± 1.3
EF70.1 ± 4.370.7 ± 4.469.3 ± 2.569.8 ± 3.569.8 ± 3.869.8 ± 3.769.5 ± 3.670.1 ± 2.770.8 ± 3.269.8 ± 2.4
72-108LVEDd39.2 ± 4.3*35.7 ± 4.536.8 ± 3.138.1 ± 3.938.3 ± 3.538.0 ± 2.338.6 ± 2.139.5 ± 1.840.1 ± 2.341.5 ± 3.6
RV15.7 ± 3.9*15.3 ± 3.416.3 ± 4.418.0 ± 5.719.0 ± 5.819.1 ± 2.218.4 ± 4.19.3 ± 2.519.5 ± 3.120.1 ± 3.5
LA24.6 ± 3.4*22.7 ± 3.123.0 ± 2.624.4 ± 3.623.8 ± 3.723.9 ± 1.924.5 ± 3.025.1 ± 3.626.1 ± 4.526.3 ± 5.1
EF71.6 ± 4.470.0 ± 3.869.8 ± 3.468.2 ± 3.767.1 ± 3.968.8 ± 2.269.1 ± 3.168.9 ± 3.569.0 ± 2.070.1 ± 2.3
108-144LVEDd39.3 ± 2.5*37.6 ± 2.137.9 ± 3.739.5 ± 3.438.2 ± 2.838.9 ± 2.039.1 ± 2.439.5 ± 3.640.2 ± 1.842.3 ± 3.7
RV15.8 ± 2.6*15.8 ± 2.716.8 ± 1.918.6 ± 5.718.3 ± 5.318.5 ± 2.717.4 ± 3.218.9 ± 3.618.2 ± 3.420.1 ± 3.6
LA24.6 ± 3.5*24.7 ± 2.324.7 ± 3.625.3 ± 3.224.1 ± 3.824.5 ± 2.825.1 ± 3.125.7 ± 3.226.1 ± 6.726.8 ± 3.4
EF70.9 ± 3.470.3 ± 3.367.9 ± 4.271.3 ± 3.670.2 ± 3.670.1 ± 4.571.2 ± 3.670.9 ± 1.868.6 ± 3.670.0 ± 6.3
144-216LVEDd50.0 ± 2.5*48.5 ± 2.347.6 ± 1.547.5 ± 2.148.1 ± 2.348.5 ± 3.449.1 ± 3.549.5 ± 3.150.1 ± 3.550.8 ± 3.6
RV21.5 ± 5.2*23.7 ± 4.322.3 ± 4.423.2 ± 3.522.2 ± 3.522.6 ± 3.623.1 ± 3.123.5 ± 3.524.5 ± 3.125.1 ± 4.1
LA32.7 ± 2.8*30.5 ± 5.232.0 ± 3.431.9 ± 3.331.7 ± 2.532.8 ± 3.132.9 ± 3.533.1 ± 3.633.5 ± 3.134.5 ± 2.9
EF68.0 ± 1.471.1 ± 3.766.3 ± 3.168.9 ± 3.569.9 ± 1.969.1 ± 2.369.3 ± 2.670.1 ± 2.368.1 ± 3.167.1 ± 2.5

P△ < 0.05 is statistically different from the previous group, P* < 0.05 is statistically different from normal datas

Data of follow-up duration Echocardiography Follow-up Datas: preoperative and postoperative P△ < 0.05 is statistically different from the previous group, P* < 0.05 is statistically different from normal datas

Discussion

Ventricular septum defect (VSD) is the most common congenital heart disease, with 40 to 60% of patients with spontaneous closure, mostly in preschool children [6-8]. Surgical closure of VSD has always been a traditional treatment, but there are many complications [6, 9, 10]. With the continuous improvement of occluder instruments in recent years and the selection of appropriate occluder devices according to the shape and size of VSD, the interventional therapy is becoming more and more reliable, safe and effective. Muscular and perimembranous VSDs have been closed using transcatheter devices at our center since 2002. Use of the transcatheter approach for the treatment of congenital heart diseases is appreciated by patients and their parents because of its decreased psychological impact (avoidance of a large scar), shorter hospital time, reduced pain and discomfort, and avoidance of admission to the intensive care unit [11]. In our study, the results of transcatheter VSD closure with an occluder device were satisfactory: the procedure was successful in 252 of 253 (99.6%) patients, confirming the results reported for interventional VSD closure in other published studies. The complete VSD closure rate immediately after the procedure was 86.6%, as documented by left ventricular angiography and TTE. This rate increased to 95.3% at the 6-month follow-up evaluation and 96% at the 2-year follow-up evaluation. The follow-up was remained for those children with residual shunts, and most residual shunts resolve postoperatively. These good closure rates are similar to those obtained in other studies of transcatheter VSD closure. Residual shunt is commonly seen in cases of multiple VSDs, as the occluder size must be appropriate for complete defect coverage. Thus the location of the occlusion and the type of occluder should be flexibly selected according to the specific conditions. The general principle to achieve the goal of completely closure of the defect without affecting the function of the aortic valve and the right atrioventricular valve, as well as to reduce the occurrence of complications. Hemolysis occurred in 4 (1.6%) patients in this study, in whom the mean VSD diameter was 8.1 ± 1.3 mm. All four of these patients had residual shunts. Postoperative hemolysis occurred in association with large defects, high shunt pressure, and residual shunt after closure. Squeezing and collision associated with rapid blood flow through the occluder can cause the mechanical rupture of red blood cells, leading to hemolysis. In these cases, aspirin was stopped and 5% sodium bicarbonate solution (5 ml/kg/d) was used for hydration and alkalization, after which three patients returned to normal after 5 days. The remaining patient developed severe hemolysis accompanied by arrhythmia and residual shunt, necessitating removal of the occluder via surgical thoracotomy. Thirty-seven (17.4%) patients developed arrhythmia postoperatively. Patients with right bundle branch blocks were treated with a conventional dose of methylprednisolone (1–2 ml/kg, q8h), which was gradually tapered. As left bundle and left anterior branch blocks have been reported to lead to death due to heart failure in China, the methylprednisolone dose can be increased to 20–30 ml/kg/d to relieve ventricular septal edema. Conduction block is related to oppression of the left and right bundle branches due to repeated stimulation of the right ventricular surface at the defect edge by the intraoperative catheter, leading to inflammation of the marginal tissue, edema, and/or exudation. Therefore, excessive long-term intracardiac procedures should be avoided, and the occluder should be slightly larger than the defect diameter of 1–2 mm to avoid compression of the irritated edge tissue (which would lead to tissue edema), thereby preventing arrhythmia. However, evaluating patients with postoperative arrhythmia, ECG monitoring and not 24 h Holter monitoring are limited. For valve regurgitation, valve insufficiency is associated with the use of an oversized occluder or an insufficient distance from the occluder edge to the valve. A left ventricular disc diameter of the occluder exceeding 50% of the circumference of the subaortic outflow tract has been reported to lead to outflow tract deformation, and thus to valve insufficiency [12]. Therefore, the position and distance between each valve and the defect should be carefully assessed preoperatively using echocardiography. Care should be taken during operation to avoid serious reflux or prolapse, which can damage the valve. Postoperative follow-up chest radiography showed gradual reduction of the cardiothoracic ratio, with eventual stabilization at 50%, which is close to the normal physiological index. TTE performed immediately after occlusion and 6 months postoperatively showed gradual reduction of the left ventricular end diastolic diameter, right ventricular diameter, and left atrial diameter, indicating smaller inner diameter of the heart cavity and the restored heart structure. At the 12-month follow-up evaluation, the inner diameter indexes of patients’ hearts had increased, but were consistent with the normal developmental ranges in all age groups, suggesting the improvement of hemodynamics. The postoperative TR-PG was significantly lower than preoperatively, suggesting that the pulmonary artery pressure had decreased. The pre- and postoperative left ventricular ejection fraction values did not differ significantly, likely due to the relatively small sizes of defects in this sample. TTE showed that the left atrium of patients was reduced compared with preoperative, and gradually increased during follow-up. Preoperative left ventricular volume overload and heart enlargement was detected with TTE; postoperatively, due to disappear shunt, the load of the heart was cut down and the heart became smaller than preoperatively. During the follow-up period, the heart became larger due to children growth. For children with growth and development, it is more intuitive to use the z-values to evaluate the heart. However, the z-values commonly used in the world are mainly applicable to Western countries such as Europe and the United States, while these are not suitable for East Asian Chinese. Recently, many centers are now gradually conducting research on the application of Z-values for children’s TTE indicators in East Asian countries such as China (Additional file 1: Table S1).

Conclusion

Transcatheter closure of VSDs is safe and effective, with a high success rate and few complications. However, indications must be strictly controlled, and the appropriate size of occluder should be selected based on angiographic findings. Patients should be monitored carefully intraoperatively via echocardiography. Further studies are needed to more fully investigate the longer term effects of this procedure.

Limitation

Holter 24-h monitoring for patients with postoperative arrhythmias is superior to ECG follow-up. Very few patients lost follow-up. Lack of normal echocardiography control. Table S1. Echocardiography Follow-up Datas: preoperative and postoperative. (DOCX 16 kb) Table S2. Size of Perimembranous ventricular septal defect. (DOCX 15 kb) Table S3. Data for pulmonary artery pressure. (DOCX 15 kb)
  9 in total

1.  Outcome following tricuspid valve detachment for ventricular septal defects closure.

Authors:  J W Gaynor; J E O'Brien; J Rychik; G R Sanchez; W M DeCampli; T L Spray
Journal:  Eur J Cardiothorac Surg       Date:  2001-03       Impact factor: 4.191

Review 2.  Percutaneous closure of ventricular septal defects.

Authors:  Gianfranco Butera; Massimo Chessa; Luciane Piazza; Diana Negura; Angelo Micheletti; Mario Carminati
Journal:  Expert Rev Cardiovasc Ther       Date:  2006-09

3.  Primary surgical closure of ventricular septal defect in the first year of life: results in 128 infants.

Authors:  S B Yeager; M D Freed; J F Keane; W I Norwood; A R Castaneda
Journal:  J Am Coll Cardiol       Date:  1984-05       Impact factor: 24.094

4.  Conduction disturbances after surgical correction of ventricular septal defect by the atrial approach.

Authors:  S M Hobbins; T Izukawa; D J Radford; W G Williams; G A Trusler
Journal:  Br Heart J       Date:  1979-03

Review 5.  The incidence of congenital heart disease.

Authors:  Julien I E Hoffman; Samuel Kaplan
Journal:  J Am Coll Cardiol       Date:  2002-06-19       Impact factor: 24.094

6.  Second natural history study of congenital heart defects. Results of treatment of patients with ventricular septal defects.

Authors:  L Kidd; D J Driscoll; W M Gersony; C J Hayes; J F Keane; W M O'Fallon; D R Pieroni; R R Wolfe; W H Weidman
Journal:  Circulation       Date:  1993-02       Impact factor: 29.690

7.  Transcatheter closure of perimembranous ventricular septal defects: early and long-term results.

Authors:  Gianfranco Butera; Mario Carminati; Massimo Chessa; Luciane Piazza; Angelo Micheletti; Diana Gabriella Negura; Raul Abella; Alessandro Giamberti; Alessandro Frigiola
Journal:  J Am Coll Cardiol       Date:  2007-09-04       Impact factor: 24.094

8.  Transcatheter closure of ventricular septal defects.

Authors:  J E Lock; P C Block; R G McKay; D S Baim; J F Keane
Journal:  Circulation       Date:  1988-08       Impact factor: 29.690

9.  Ventricular septal defect: early against late surgical repair.

Authors:  Waqas J Siddiqui; Praneet Iyer; Samridhi Amba; Salman Muddassir; Oleg Cheboterav
Journal:  J Community Hosp Intern Med Perspect       Date:  2016-02-17
  9 in total
  6 in total

1.  Clinical Experience of Transcatheter Closure for Ventricular Septal Defects in Children Weighing under 15 kg.

Authors:  Tsung-Yen Chen; Ying-Tzu Ju; Yu-Jen Wei; Min-Ling Hsieh; Jing-Ming Wu; Jieh-Neng Wang
Journal:  Acta Cardiol Sin       Date:  2021-11       Impact factor: 2.672

2.  Unplanned Surgery After Transcatheter Closure of Ventricular Septal Defect in Children: Causes and Risk Factors.

Authors:  Penghui Yang; Zhijun Wu; Zhiyuan Liu; Jing Zhang; Hao Zhou; Xiaojuan Ji; Qijian Yi; Mi Li
Journal:  Front Pediatr       Date:  2021-11-30       Impact factor: 3.418

3.  Influence of percutaneous catheter intervention for congenital perimembranous ventricular septal defects in children on the cardiac conduction system and associated risk factors: a meta-analysis.

Authors:  Yu-Qing Lei; Wen-Hao Lin; Shi-Hao Lin; Wen-Peng Xie; Jian-Feng Liu; Qiang Chen; Hua Cao
Journal:  J Cardiothorac Surg       Date:  2022-02-16       Impact factor: 1.637

4.  Study on Inflammatory Factors in Aneurysmal Perimembranous Ventricular Septal Defect in Congenital Heart Disease.

Authors:  Jin Zhou; Ying Liu; Jing Wang; Wei Yan; Yongjian Liu; Litao Chen; Zhixing Du; Qilian Xie
Journal:  Biomed Res Int       Date:  2022-07-19       Impact factor: 3.246

5.  Left Anterior Fascicular Block After Transcatheter Closure of Ventricular Septal Defect in Children.

Authors:  Zhijun Wu; Penghui Yang; Ping Xiang; Xiaojuan Ji; Jie Tian; Mi Li
Journal:  Front Cardiovasc Med       Date:  2021-06-11

6.  Echocardiographic Classification of Perimembranous Ventricular Septal Defect Guides Selection of the Occluder Design for Their Transcatheter Device Closure.

Authors:  Anil Kumar Singhi; Kothandam Sivakumar
Journal:  J Cardiovasc Imaging       Date:  2021-03-30
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.