Keiichiro Mie1, Akiyoshi Hayashi2, Hidetaka Nishida1, Mari Okamoto1, Kazuo Yasuda3, Mio Nakata1,3, Kazuyuki Fukatsu4, Norie Matsunami4, Shogo Yamashita5, Fumihito Ohashi1, Hideo Akiyoshi1.
Abstract
The accuracy of urine analyzers used for dogs and cats has remained uncertain. This study examines the agreement between results of urine analysis obtained using two devices marketed for animals and for humans and the results of quantitative biochemical analysis. The degrees of concordance for bilirubin and ketones in the same category were ~80%, but for pH these were only ~60% in dogs and cats. Degrees of concordance for protein and the UP/C ratio clearly differed between the devices for animals and humans. We found that values for bilirubin and ketones obtained using urine analyzers may be reliable, but pH is unlikely to be accurate enough to be clinically useful for dogs and cats.Entities:
Keywords: cat; dog; pH; urine analyzer
Year: 2019 PMID: 31611480 PMCID: PMC6895637 DOI: 10.1292/jvms.18-0468
Source DB: PubMed Journal: J Vet Med Sci ISSN: 0916-7250 Impact factor: 1.267
The ranges of quantitative values matching categories defined by the urine analyzer results
The degrees of concordance between the urine analyzer results and the quantitative biochemical analysis for results in the same category
| Number | Animal (%) | Human (%) | ||
|---|---|---|---|---|
| Dogs | ||||
| Glucose | 365 | 77 | 73 | |
| Protein | 329 | 72 a) | 46 | |
| pH | 296 | 63 | 63 | |
| Bilirubin | 152 | 88 | 82 | |
| Ketone | 150 | 82 | 81 | |
| Albumin | 152 | 58 | 55 | |
| UP/C | 320 | 77 a) | 64 | |
| Cats | ||||
| Glucose | 307 | 83 b) | 75 | |
| Protein | 276 | 49 b) | 39 | |
| pH | 245 | 64 | 64 | |
| Bilirubin | 102 | 86 | 86 | |
| Ketone | 102 | 85 | 81 | |
| Albumin | 99 | 64 a) | 29 | |
| UP/C | 119 | 81 b) | 66 | |
a) P<0.01 and b) P<0.05 between animal and human analyzers.
The degrees of concordance between the urine analyzer results and the quantitative biochemical analysis results in the ± 1 category
| Number | Animal (%) | Human (%) | ||
|---|---|---|---|---|
| Dogs | ||||
| Glucose | 365 | 100 | 100 | |
| Protein | 329 | 97 a) | 89 | |
| pH | 296 | 96 | 96 | |
| Bilirubin | 152 | 100 | 99 | |
| Ketone | 150 | 100 | 100 | |
| Albumin | 152 | 93 a) | 78 | |
| UP/C | 320 | 97 | 96 | |
| Cats | ||||
| Glucose | 307 | 100 | 99 | |
| Protein | 276 | 95 a) | 80 | |
| pH | 245 | 100 | 100 | |
| Bilirubin | 102 | 100 | 100 | |
| Ketone | 102 | 100 | 100 | |
| Albumin | 99 | 90 a) | 59 | |
| UP/C | 119 | 97 | 93 | |
a) P<0.01 between animal and human analyzers.
The degrees of concordance of glucose between the results of the urine analyzer for animals and the quantitative biochemical analysis results in dogs
The sensitivities and specificities of the urine analyzers
| Number | Animals | Humans | Expected category | |||||
|---|---|---|---|---|---|---|---|---|
| Sensitivity (%) | Specificity (%) | Sensitivity (%) | Specificity (%) | Animals | Humans | |||
| Dogs | ||||||||
| Glucose | 365 | 67.2 | 88.6 | 50.0 | 100.0 | - | - | |
| Protein | 329 | 87.1 | 81.7 | 98.6 | 60.0 | - | - | |
| pH | 296 | 63.4 | 97.0 | 63.4 | 97.0 | 6.0–7.5 | 6.0–7.5 | |
| Bilirubin | 152 | 75.7 | 93.9 | 75.7 | 93.9 | - | - | |
| Ketone | 150 | 65.1 | 95.3 | 65.1 | 95.3 | - | - | |
| Albumin | 152 | 94.1 | 58.2 | 93.2 | 57.8 | 1.0 | 10.0 | |
| UP/C | 320 | 90.2 | 91.9 | 96.0 | 86.4 | <0.2 | 200.0 | |
| Cats | ||||||||
| Glucose | 307 | 77.4 | 95.3 | 69.6 | 100.0 | - | - | |
| Protein | 276 | 87.3 | 67.3 | 99.5 | 49.1 | - | - | |
| pH | 245 | 70.4 | 100.0 | 70.4 | 100.0 | 6.0–7.5 | 6.0–7.5 | |
| Bilirubin | 102 | 69.6 | 100.0 | 69.6 | 100.0 | - | - | |
| Ketone | 102 | 80.6 | 100.0 | 80.6 | 100.0 | - | - | |
| Albumin | 99 | 84.8 | 81.1 | 93.5 | 18.9 | 1.0 | 10.0 | |
| UP/C | 119 | 90.3 | 94.3 | 65.4 | 98.9 | <0.2 | 200.0 | |
Fig. 1.Data for quantitative biochemical analysis of protein in dogs. The data are shown as box and whisker plots. Box whisker plots represent the 25th/75th percentile, median, and range. The white and gray boxes represent the quantitative data categorized by human and animal analyzers, respectively. *P<0.05 and **P<0.01 between animal and human analyzers in the same categories.
Fig. 2.Data for quantitative biochemical analysis of glucose and protein in cats. (A) glucose, (B) protein. The data are shown as box and whisker plots. Box whisker plots represent the 25th/75th percentile, median, and range. The white and gray boxes represent the quantitative data categorized for human and animal analyzers, respectively. *P<0.05 and **P<0.01 between animal and human analyzers in the same categories.