False negative diagnoses of critical congenital heart disease with screening neonatal pulse oximetry.

J I E Hoffman1.   

Abstract

BACKGROUND: False negative pulse oximeter results occur in new born infants with critical congenital heart disease who have an oximeter saturation ≥95%. Some of these infants have abnormal physical findings but others do not.
OBJECTIVES: To determine the causes of false negative results.
METHODS: Mathematical analysis of determinants of arterial oxygen saturation and discussion of oximeter bias.
RESULTS: False negative oximeter results are not rare; the sensitivity of pulse oximetry screening for critical congenital heart disease is about 80%. The high saturation may be due to a very small right-to-left shunt at the time of study, a relatively high cardiac output and mixed venous saturation, or to positive bias in oximeter readings. It may also be due to some critical congenital heart lesions that do not show desaturation at the time of testing.
CONCLUSIONS: A diagnosis of a normal heart based on a negative oximeter test is presumptive, and requires careful follow-up for 1-2 weeks after birth.

Entities:  

Keywords:  Mixed venous saturation; cardiac output; oximeter zzm321990bias; right-to-left shunt

Mesh:

Substances:

Year:  2020        PMID: 31594260     DOI: 10.3233/NPM-190297

Source DB:  PubMed          Journal:  J Neonatal Perinatal Med        ISSN: 1878-4429


  1 in total

1.  Prevalence of aortic coarctation in Bogotá-Colombia from 2001 to 2018. The diagnostic needs to improve.

Authors:  Daniel F Ardila; Angela G Rincón-Montana; Laura D García-Pérez; Gloria Gracia; Ignacio Zarante
Journal:  Arch Cardiol Mex       Date:  2022-04-04
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.