Literature DB >> 31580602

Recalibration of the NBS Glass Standards of Spectral Transmittance.

Harry J Keegan, John C Schleter, Marion A Belknap.   

Abstract

In 1934, Gibson, Walker, and Brown developed sets of four colored glass filters to serve as working standards of spectral transmittance for checking the reliability of spectrophotometers. Several sets of these glasses were measured carefully and reserved and designated as future reference standards. Duplicate standards evaluated by comparison with the reference standards are available by purchase to the public. The current set of reference standards was established in the years 1945 to 1947, and one of these reference standards (selenium-red) was recalibrated in 1952. This paper reports a recalibration, made in 1961 and 1962, of all four glasses (selenium-red, carbon-yellow, copper-green, cobalt-blue) on three spectrophotometers (Cary 14, Beckman DU, König-Martens). Except for the cobalt-blue standard, the values of spectral transmittance found differ from those previously assigned by amounts differing at some wavelengths by as much as or slightly more than the uncertainties estimated for the present values, though not by amounts exceeding the combined uncertainties of the present and previous determinations. The indicated changes for these three standards are fairly regular, however, and support the view that the selenium-red and carbon-yellow standards are changing chiefly by formation of a reflectance-reducing film on the surfaces. The indicated rate of upward drift is slow, and suggests that it takes about 10 years for the drift to exceed the assigned uncertainity.

Entities:  

Year:  1963        PMID: 31580602      PMCID: PMC5322774          DOI: 10.6028/jres.067A.056

Source DB:  PubMed          Journal:  J Res Natl Bur Stand A Phys Chem        ISSN: 0022-4332


1. Introduction

In 1934, Gibson, Walker, and Brown [1]2 described a set of four colored glass filters which they had developed to serve as standards of spectral transmittance to check the reliability of spectrophotometers. Several sets of these glasses were carefully measured at a number of wavelengths by using mercury, helium and incandescent sources and were reserved and designated as future reference standards. Duplicate standards evaluated by comparison with the reference standards were made available to the public by purchase. Although nearly nonselective filters are best suited to the detection of errors in the photometric scale, the filters chosen have spectral transmittances that vary markedly with wavelength and thus permit detection of slit-width errors and stray-energy errors as well as photometric-scale errors; thus a single selective filter may afford checks of the photometric scale over a large fraction of the whole scale [2, 3]. Table 1 identifies the glasses that have served as reference standards, and figure 1 shows the approximate spectral transmittances of these filters. The reports issued with the sets of duplicate standards gave the assigned values of transmittance at about 20 wavelengths between 390 and 750 nm 3 together with the estimated uncertainties and the temperature coefficients.
Table 1

Identification of the glasses used as reference standards

No.Designation1934–19401945–1947




1Selenium-redJena OG 3–8Jena OG 3–21
Jena OG 3–10
2Carbon-yellowCorningCorning
 HT Yellow α16 HT Yellow C-14
 HT Yellow α7
3Copper-greenJena BG 14–1Jena BG 14–19
Jena BG 14–10
4Cobalt-blueCorningComing
 G 55A α11 G 55A β9
 G 55A α3
Figure 1

Typical spectral transmittance curves of glass duplicate standards issued by the NBS for checking the reliability of spectrophotometers.

The four types are: 1. Selenium-red, 2. Carbon-yellow, 3. Copper-green, and 4. Cobalt-blue.

Table 2 gives for each wavelength the light source used and the estimated uncertainties and a multiple of the temperature coefficients of the reference standards expressed on the absorbance scale. Absorbance is defined as the logarithm of the reciprocal of the internal transmittance T Since internal transmittance for these glasses is closely equal to transmittance T multiplied by a constant (about 1/0.9) to correct for reflection losses [8], differences in absorbance are closely equal to differences in the logarithm of the reciprocal of transmittance T uncorrected for reflection losses. Accordingly the uncertainties U in absorbance entered in table 2 are computed from the uncertainties U in transmittance as : Similarly, the temperature coefficients δ of absorbance are computed from the relation : where θ is the temperature of the glass in degrees C, b is the thickness of the glass in mm, and −log10T25 is the absorbance measurement made at 25 °C. The temperature coefficient δ is thus the change in −log10T (closely equal to the change in −log10T) for a temperature change of 1 °C, and a thickness of 1 mm. These glass reference standards are measured at 25 °C and are used over a temperature range of ± 15 °C. For convenience, because the temperature coefficient is relatively small compared with the value of absorbance, a quantity d equal to 15 times the temperature coefficient is used to designate the change in absorbance for a temperature change of 15 °C, as follows: This relationship is considered valid for values of θ between 0 and 50 °C.
Table 2

Estimated uncertainty in the value of absorbance, Ua, and the change d, in absorbance for temperature change of 15° C, for the selected wavelengths for incandescent (Inc.) and line (Hg and He) sources used for the four reference standards

WavelengthSourceSelenium-redCarbon-yellowCopper-greenCobalt-blue










λ in nmUadUadUadUad
390    Inc…………±0.052−0.0005±0.0020.000  ±0.0020.000  
404.7Hg………….069−.0025.002.000  .002.000  
420    Inc………….046−.0025……………………
435.8Hg………….014−.0015.001.000  .002.000  
471.3He………….011+.001  .001.000  .004.000  
491.6Hg……………………………….005+.0005
501.6He………….006+.0035.002.000  .007+.001  
520    Inc………….004+.0035………….010+.0005
530    Inc………….003+.003  .002.000  …………
540    Inc……………………………….008−.003  
546.1Hg………….003+.0025.003.000  .008−.004  
560    Inc………….002+.002  .003.000  .007−.003  
570    Inc0.000……………………………………
578    Hg±.06  +0.17  .003+.0015.004−.0005.008−.004  
587.6He.04  +.11  .003+.0015……………………
600    Inc.008+.022.002+.001  .006−.0005.024−.005  
620    Inc.003+.002.002+.0005.005−.001  .017–.003  
640    Inc.002.000.003+.0005.007−.001  .023–.005  
660    Inc.002.000.003.000  ……………………
667.8He…………………….007−.001  .027+.004  
680    Inc……………………………….033+.006  
690    Inc.002.000.003.000  ………….026+.004  
706.5He……………………………….004+.002  
710    Inc…………………….012−.001  …………
720    Inc.002.000.003−.0005………….003+.0005
750    Inc.002.000.003−.0005.015−.0005002.000  
The current set of reference standards was measured in the years 1945 to 1947. A detailed analysis of the data obtained for one of the reference standards (cobalt-blue, Corning G 55A β9) was published by Gibson and Balcom [4] in 1947. The uncertainties, U, were estimated as the huge error (4.9 times the probable error) of the mean values of transmittance. In 1950, Gibson and Belknap [5] published a study of the permanence of samples of the carbon-yellow, the copper-green, and the cobalt-blue glasses exposed under clear glass to south skylight and sunlight for periods up to three years. Expressed in terms of transmittance change ΔT, these samples were found to change but slightly, excepting only the copper-green glass below 500 nm, where there appears to be a definite solarization effect. No such study was made for ordinary use of these glasses in the laboratory. It was assumed that if the glasses would change only slowly with extensive exposure to sunlight under glass, they would not change significantly in many years of ordinary use. A redetermination of the spectral transmittance of the current reference standard for selenium-red glass was carried out in 1952 and resulted in slight changes in the assigned values, but it could not be determined with certainty that the transmittance had changed. The 1946 and 1952 data are given in table 3.
Table 3

Previously adopted value of transmittance, T, and estimated uncertainty in the values of transmittance, U, at the respective wavelengths for the four reference standards

WavelengthSelenium-red (1952)*Carbon-yellow (1946)*Copper-green (1946)*Cobalt-blue (1946)*









λ in nmTUTUTUTU
390    …………0.025  ±0.003  0.862±0.0030.895  ±0.004  
404.7………….020  .003  .877.003.884  .005  
420    ………….019  .002  ……………………
435.8………….0240.0007.893.004.806  .003  
471.3………….081  .002  .894.004.612  .006  
491.6……………………………….344  .004  
501.6………….208  .003  .859.005.245  .004  
520    ………….316  .003  ………….091  .002  
530    ………….379  .003  .760.004…………
540    ……………………………….0308.0006
546.1………….479  .003  .671.004.0335.0006
560    ………….557  .003  .585.004.064  .001  
570    0.000  ……………………………………
578    .0042±0.0005.636  .003  .473.004.0273.0005
587.6.118  .01    .668  .003  ……………………
600    .550  .02    .699  .003  .350.003.0074.0002
620    .852  .007  .731  .004  .256.003.0100.0002
640    .904  .005  .747  .005  .187.003.0074.0002
660    .914  .005  .754  .005  ……………………
667.8…………………….122.002.034  .002  
680    ……………………………….14    .01    
690    .919  .005  .755  .005  ………….34    .02    
706.5……………………………….713  .007  
710    …………………….074.002…………
720    .918  .005  .748  .005  ………….845  .005  
750    .917  .005  .730  .005  .057.002.901  .005  

Date of adoption of values.

Meanwhile, interest in the duplicate standards was increasing. Table 4 shows the number of duplicate standards of each type of glass issued in the years 1933–1942, 1943–1946, and 1947–1962.
Table 4

Numbers of duplicate standards issued

YearsCalibration of duplicate standards in charge of:Selenium-redCarbon-yellowCopper-greenCobalt-blueTotal







1933–1942Mabel E. Brown and Geraldine W. Haupt.720101552
1943–1946Margaret M. Balcom and Lois A. Peterson.01171634
1947–1962Marion A. Belknap2016111399393





 Totals27192130130479
A set of these glass standards, issued in 1958 to the Frankford Arsenal, was submitted in October 1961 for recalibration. A routine comparison of them with the same reference standards used in the original calibration showed that the transmittances for each of the four glasses bore a significantly different relation to the corresponding reference standard than in 1958. It was presumed that this changed relation referred primarily to the duplicate standards, whose uses and possible exposure to radiant energy at the Frankford Arsenal were unspecified, rather than to the reference standards known to be irradiated only by incandescent-lamp light for short periods of time at infrequent intervals; and this presumption was subsequently proved to be correct. Nevertheless, check data obtained in 1952, 1953, 1959, and 1960 for the reference standards were reviewed. Although no single set of data proved that any of the reference standards had certainly changed, all four sets of data considered together showed some evidence of a slow continuing drift in spectral transmittance of the selenium-red, the carbon-yellow and the copper-green reference standards. The cobalt-blue glass alone showed no change in spectral transmittance. An extensive recalibration of all four reference standards was then undertaken. It is the purpose of the present paper to describe this recalibration, and to discuss its results.

2. Method of Spectrophotometry

The recalibration of the four reference standards of spectral transmittance was carried out by means of the Cary 14 recording spectrophotometer, the Beckman DU manual spectrophotometer [6], and the König-Martens visual spectrophotometer [7]. Table 5 lists the spectrophotometers used in the several calibrations of reference standards of spectral transmittance. The procedural details of the present recalibration are given below.
Table 5

Spectrophotometers used in the several calibrations of reference standards of spectral transmittance

SpectrophotometerDate of calibration


König-Martens visual1930–1940 1940–1950 1950–1960 1960–
Gibson photoelectric and thermoelectric.1930–1940 1940–1950
Hilger photographic sector-photometer1930–1940
Beckman DU photoelectric (“BQ-1”)1950–1960 1960–
Beckman DU photoelectric (“BQ-2”)1940–1950
Cary 14M recording photoelectric1960–

2.1. Cary Recording Spectrophotometer

Two sets of measurements of each of the four reference standards of spectral transmittance were made on the Cary recording photoelectric spectrophotometer (Model 14M, Serial No. 173) by means of the absorbance (−logT) scale. At the same time measurements were made of a clear borosilicate crown glass, 1.0 mm thick, of known index of refraction and Abbe value (nD= 1.517, v=64.5).4 This index and Abbe value are the same as those found for the borosilicate crown glass, 1.0 mm thick, whose spectral transmittances were known from a previous study [8]. These known spectral transmittances were used as a photometric-scale correction for transmittances of the reference standards greater than 40 percent. For each wavelength the ratio Rλ of the reference-standard transmittance to that of the borosilicate crown glass was computed from the absorbance readings. The value Tλ of the transmittance of the reference standard was computed as: where T0λ is the known spectral transmittance of the borosilicate crown glass [8]. For Tλ less than 40 percent the value corresponding to the reading of the absorbance scale was accepted without correction. The wavelength calibration of the instrument was carried out as described by Keegan, Schleter, and Judd [9]. All measurements refer to the specimen at 25 °C.

2.2. Beckman DU Spectrophotometer

The spectral transmittances of the reference standards were measured on the Beckman DU (“BQ-1”) spectrophotometer with incandescent source after calibration of wavelength scale at emission lines of mercury, helium, hydrogen, neon, and cesium. Frequent checks of the wavelength calibration were made by means of the mercury line at 546.1 nm. Each standard was measured in two positions in the sample holder for each set of readings. The holder was positioned so as to place the standard near the exit slit and away from the phototube. As recommended by Gibson and Balcom [4], the measurements were made with no lens over the exit slit. The spectral transmittance of each standard was measured at selected wavelengths by ratio to a blank beam set to 100 percent transmittance as a reference. The shift from the blue-sensitive to red-sensitive phototube was made at 620 nm. The selector switch was shifted to 0.1 for transmittance readings below 10 percent to give “times-10” readings at narrower slitwidths. A purple stray-energy filter was used for all measurements taken at 390 nm. The sample compartment was surrounded by a water jacket maintained at 25 °C by a constant-temperature water bath. The numbers of sets of readings made on the reference standards are as follows: The extra set of readings above 600 nm for the selenium-red standard was obtained by ratio of readings for the standard to those for a clear glass (Corning 9700).

2.3. König-Martens Spectrophotometer

The spectral transmittances of the four reference standards were measured on the König-Martens visual spectrophotometer [7] by using emission lines of a mercury source and a helium source and, at some wavelengths, by using an incandescent source. As explained on pages 464–465 of reference [5], it was early found advisable, because of the low luminosities of the Hg and He sources at 471.3, 491.6, 667.8, and 706.5 nm, to use the incandescent source at these wavelength settings, and this has been done in all later work on the König-Martens visual spectrophotometer. However, for the present standardization of the NBS reference standards, measurements were made with both incandescent and line sources at these four wavelengths. The wavelength scale was checked by settings on certain of the mercury and helium lines each time the instrument was used and each time the width of the ocular slit was changed. The widths of the collimator and ocular slits were 0.2 mm for the readings taken by means of the incandescent source, and were varied from 0.2 to 0.4, or 0.5 mm for the line sources. Accurately calibrated sector disks (transmittances approximately 1, 10, 50, and 60 percent) were used for measurement of low transmittances or for transmittances near to those of the sectors to increase the accuracy of the readings. Stray-energy filters were used at wavelengths near the two ends of the visible spectrum and at wavelengths of low transmittance. Each set of readings consisted of 20 readings of angle on the Martens photometer; first, 10 for the standard inserted in one beam, and second, 10 for it in the other beam. The transmittance was computed as the cotangent of the first angle multiplied by the tangent of the second. At least two sets of readings were taken for each standard at each selected wavelength between 404.7 and 660 nm and a few additional readings between 660 and 706.5 nm because of the added uncertainty of the settings in this region of low luminosity. All readings were made with the standards in a holder maintained at 25° C by means of a constant-temperature water bath.

3. Reduction of Data and Estimates of Uncertainty

The spectral transmittances assigned to the reference standards are weighted means of the transmittances found individually by the three spectrophotometers used. The weights assigned to the individual values of transmittance for a particular standard at a particular wavelength were based on the known relative capabilities of the three instruments. For example, the values obtained by means of the König-Martens visual spectrophotometer in spectral regions yielding observing fields of low luminance (such as near the extremes of the visible spectrum) were given less weight because of the resulting unreliability of the visual settings. On the other hand values of spectral transmittance obtained by means of an emission line in a spectral region of rapid variation of transmittance with wavelength were given more weight. In tables 6, 7, 8, and 9, are given for each of the four reference standards and for each of the three instruments at the respective wavelengths, the average value of the transmittance so measured, the assigned weight, the range in transmittance, the adopted weighted mean, and the estimated uncertainty in the value of transmittance.
Table 6

Spectral transmittance, T, of selenium-red reference standard Jena OG 3–21, 1.7 mm, as measured on indicated spectrophotometers; assigned weight, W, range, R, adopted weighted mean, and estimated uncertainty of value of transmittance, U

WavelengthKönig-Martens (1962)*Beckman DU (1961–1962)*Cary Model 14 (1962)*RangeAdopted weighted meanEstimated uncertainty










λ in nmTWTWTWRTU
390to570}0.000  ……0.000  ……0. 000……0.000  0.000  ……
578    .00414.00332.00344.0008.0037±0.0005
587.6.115  4.102  3.110  3.013  .110  .01    
600    .564  3.556  4.561  3.008  .560  .01    
620    .858  4.856  4.861  2.005  .857  .006  
640    .906  4.903  4.909  2.006  .905  .004  
660    .913  4.914  4.920  2.007  .915  .004  
690    .917  3.918  5.925  2.008  .919  .004  
720    .918  2.919  6.926  2.008  .920  .004  
750    .919  1.919  7.925  2.006  .920  .004  

Date of measurement.

Table 7

Spectral transmittance, T, of carbon-yellow reference standards Corning HT yellow C14, 2.0 mm, as measured on indicated spectrophotometers; assigned weight, W, range, R, adopted weighted mean and estimated uncertainty of transmittance, U

WavelengthKönig-Martens (1962)*Beckman DU (1961–1962)*Cary Model 14 (1962)*RangeAdopted weighted meanEstimated uncertainty










λ in nmTWTWTWRTU
390    --------------------0.025  50.025  50.000  0.025  ±0.003  
404.70.019  3.019  4.019  3.000  .019  .003  
420    --------------------.019  5.019  5.000  .019  .002  
435.8.02444.02342.02484.0014.0244.0008
471.3.081  2.080  4.082  4.002  .081  .002  
501.6.209  3.206  4.209  3.003  .208  .003  
520    .316  3.318  4.320  3.004  .318  .003  
530    .381  3.380  4.382  3.002  .381  .003  
546.1.480  3.480  4.484  3.004  .481  .003  
560    .558  3.560  4.562  3.004  .560  .003  
578    .634  3.639  4.642  3.008  .638  .004  
587.6.671  3.669  4.674  3.005  .671  .004  
600    .702  4.702  4.704  2.002  .702  .004  
620    .736  4.734  4.739  2.005  .736  .004  
640    .750  4.751  4.756  2.006  .752  .005  
660    .756  4.758  4.763  2.007  .758  .005  
690    .760  3.759  4.764  3.005  .761  .005  
720    --------------------.752  6.756  4.004  .754  .005  
750    --------------------.733  6.736  4.003  .734  .005  

Date of measurement.

Table 8

Spectral transmittance, T, of copper-green reference standard Jena BG 14–19, 2.0 mm, as measured on indicated spectrophotometers; assigned weight, W, range, R, adopted weighted mean, and estimated uncertainty of transmittance, U

WavelengthKönig-Martens (1962)*Beckman DU (1962)*Cary Model 14 (1962)*RangeAdopted weighted meanEstimated uncertainty










λ in nmTWTWTWRTU
390    …………0.86660.86440. 0020.865±0.003
404.70. 8762.8784.8784.002.878.003
435.8.8933.8914.8923.002.892.003
471.3.8932.8934.8924.001.893.003
501.6.8623.8614.8603.002.861.003
530    .7593.7654.7623.006.762.004
546.1.6723.6784.6773.006.676.004
560    .5933.5944.5893.005.592.004
578    .4773.4814.4763.005.478.004
600    .3483.3584.3523.010.353.005
620    .2583.2574.2583.001.258.003
640    .1873.1884.1873.001.187003
667.8.1233.1234.1223.001.123002
710    .0781.0755.0754.003.075.002
750    ………….0575.0575.000.057.002

Date of measurement.

Table 9

Spectral transmittance, T, of cobalt-blue reference standard Corning G 55 Aβ9, 3.0 mm, as measured on indicated spectrophotometers; assigned weight, W, range, R, adopted weighted mean, and estimated uncertainty of transmittance, U

WavelengthKönig-Martens (1962)*Beckman DU (1962)*Cary Model 14 (1962)*RangeAdopted weighted meanEstimated uncertainty










λ in nmTWTWTWRTU
390    --------------------0.894  60.892  40.002  0.893  ±0.004  
404.70.887  2.880  4.880  4.007  .881  .005  
435.8.808  3.808  4.805  3.003  .807  .003  
471.3.613  2.618  4.612  4.006  .615  .006  
491.6.344  2.351  4.342  4.009  .346  .004  
501.6.248  3.249  4.244  3.005  .247  .004  
520    .091  4.092  3.089  3.003  .091  .002  
540    .03144.03083.03103.0006.0311.0006
546.1.03394.03343.03443.0010.0339.0006
560    .064  4.064  3.065  3.001  .064  .001  
578    .02904.02813.02743.0016.0282.0008
600    .00734.00723.00733.0001.0073.0004
620    .01004.01043.01043.0004.0102.0004
640    .00756.00822.00742.0008.0076.0004
667.8.032  4.032  4.034  2.002  .032  .002  
680    .127  2.132  5. 1393.012  .133  .01    
690    .310  2.338  5.352  3.042  .337  .02    
706.5.708  2.703  4.714  4.011  .708  .007  
720    --------------------.844  7.852  3.008  .846  .005  
750    --------------------.902  7.909  3.007  .904  .005  

Date of measurement.

The year or years of measurement on the indicated spectrophotometers are shown in the parenthesis under the name of the instrument. The uncertainty of the assigned value of transmittance at a particular wavelength was made the same as that assigned previously (see table 3); with exceptional cases where it was agreed that a change was necessary. In these cases the uncertainty was estimated from the range of the individual values obtained on the three instruments averaged over a spectral region variable in extent from 0 to 50 nm and centered on that wavelength. Some smoothing of these values was resorted to whenever no reason was apparent for a rapid change in uncertainty with wavelength. In no case was the uncertainty allowed to be less than one-half the range of the spectrophotometric data. These ranges correspond roughly to estimates of three times the standard deviation of the adopted value [10]. In table 10 the change in transmittance between the present value and the previously adopted value of spectral transmittance is given for each of the four reference standards at the respective wavelengths.
Table 10

Change in spectral transmittance, ΔT, for the four reference standards between the present adopted weighted mean and the previously adopted values

Wave-lengthSelenium-redCarbon-yellowCopper-greenCobalt-blue





λ in nmΔT(1962–1952)ΔT(1962–1946)ΔT(1962–1946)ΔT(1962–1946)
390    ……0.000  +0.003−0.002  
404.7……−.001  +.001−.003  
420    …….000  …………
435.8……+.0004−.001+.001  
471.3…….000  −.001+.003  
491.6………………+.002  
501.6…….000  +.002+.002  
520    ……+.002  …….000  
530    ……+.002  +.002……
540    ………………+.0003
546.1……+.002  +.005+.0004
560    ……+.003  +.007.000  
570    0.000  ………………
578    −.0005+.002  +.005+.0009
587.6−.008  +.003  …………
600    +.010  +.003  +.003−.0001
620    +.005  +.005  +.002+.0002
640    +.001  +.005  .000+.0002
660    +.001  +.004  …………
667.8…………+.001−.002  
680    ………………−.003  
690    .000  +.006  ……−.002  
706.5………………−.005  
710    …………+.001……
720    +.002  +.006  ……+.001  
750    +.003  +.004  .000+.003  
Figures 2, 3, 4, and 5 show as circles these differences in spectral transmittance assigned by the present recalibration and that assigned by the previous calibration for the selenium-red, the carbon-yellow, the copper-green, and the cobalt-blue standards. They also show as broken lines centered about the abscissa the estimates of the uncertainties of the values of spectral transmittance resulting from the present recalibration.
Figure 2

Selenium-red reference standard.

Differences (shown as circles) between spectral transmittances assigned in 1962 and in 1952 compared to estimates (shown as broken lines) of the uncertainty of the 1962 recalibration.

Figure 3

Carbon-yellow reference standard.

Differences (shown as circles) between spectral transmittances assigned in 1962 and in 1946 compared to estimates (shown as broken lines) of the uncertainty of the 1962 recalibration.

Figure 4

Copper-green reference standard.

Differences (shown as circles) between spectral transmittances assigned in 1962 and in 1946 compared to estimates (shown as broken lines) of the uncertainty of the 1962 recalibration.

Figure 5

Cobalt-blue reference standard.

Differences (shown as circles) between spectral transmittances assigned in 1962 and in 1946 compared to estimates (shown as broken lines) of the uncertainty of the 1962 recalibration.

4. Discussion

It will be noted from figure 2 that three of the ten newly assigned values of spectral transmittance for the selenium-red reference standard differ from the previous assigned values by amounts approaching the estimated uncertainty. Because the previously assigned values are uncertain by about the same amount, this recalibration by itself fails to prove that the selenium-red standard has changed since 1952. The results of the two spectrophotometers (König-Martens and Beckman DU “BQ-1”) common to the 1952 determination and the present determination, however, both indicate that a change has occurred. Furthermore, comparison of the 1947 with the 1952 determination shows similar changes. It seems likely that indications of increases in transmittance between 620 and 750 nm correspond to formation of a reflectance-reducing film on the surfaces. Several of the duplicates of this reference standard show visible evidence of such films. The decreases in transmittance indicated near the cut-off of the transmittance curve may not be real. (See fig. 2, wavelengths 570 and 587.6 nm). In any case, there is no established cause for such a change. Figure 3 gives an indication of formation of a reflectance-reducing film on the surfaces of the carbon-yellow reference standard similar to that by figure 2 for the selenium-red standard. Note that the indicated discrepancies between the present and the previous assignments of spectral transmittance are roughly proportional to the spectral transmittances themselves, (see fig. 1), as would be expected from the formation of a reflectance-reducing film. Figure 4 indicates some increases (390, 405, 500–620 nm) and some decreases (436, 471 nm) in assigned values of spectral transmittance. This indicated pattern is not consistent with the hypothesis that a reflectance-reducing film has been forming on the copper-green reference standard. If the indicated changes are real, and this has not been proved, they must be ascribed to a cause other than film formation (such as a slow chemical change in the glass), or to a combination of film formation with other cause or causes unknown. Figure 5 gives no plausible suggestion that the cobalt-blue reference standard is changing even slowly. In table 10 the indicated changes are either small compared to the estimated uncertainty of the present determinations, or irregular. In summary, it may be stated that three of the reference standards of spectral transmittance are probably subject to a slight impermanence, the maximum changes in a 10-year period being of the order of magnitude of the uncertainties in their calibration. Owners of duplicate standards not subjected to unusually severe usage (exposure to high-energy particles, ultraviolet energy, high temperatures, or chemical fumes) may find in figures 2 to 5 a reasonably valid basis for revising the adopted transmittance for their duplicate standards. In the absence of information to the contrary, the changes may be assumed as a first approximation to be proportional with time, and the time for recently issued duplicates (1952 to 1961) should be counted from 1952 for the selenium-red standards, and from 1946 for the carbon-yellow, the copper-green, and the cobalt-blue standards.
Selenium-redCarbon-yellowCopper-greenCobalt-blue





BelowAboveFullFullFull
600nm600nmspectraspectraspectra
23732
  1 in total

1.  Transmission measurements with the Beckman quartz spectrophotometer.

Authors:  K S GIBSON; M M BALCOM
Journal:  J Res Natl Bur Stand (1934)       Date:  1947-06
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.