The Anterior Subcutaneous Pelvic Ring Fixator: No Biomechanical Advantages Compared with External Fixation.

Martin C Jordan1, Anna-Christina Brems1, Timo Heintel1, Hendrik Jansen1, Stefanie Hoelscher-Doht1, Rainer H Meffert1.   

Abstract

BACKGROUND: Subcutaneous fixation of the anterior pelvic ring is an emerging surgical technique for trauma patients. The aim of this study was to biomechanically evaluate 2 internal fixation devices for stabilizing a disrupted pelvic ring and compare them with traditional external fixation.
METHODS: Thirty-six synthetic pelvises with a simulated unstable ring fracture (anteroposterior compression type III) were divided into 3 groups. Group A underwent fixation with a supra-acetabular external fixator; group B, with an internal fixator using the USS II polyaxial system; and group C, with an internal fixator using the Click'X polyaxial system. Biomechanical testing included measurement of peak-to-peak displacement at 300, 400, and 500 N; total displacement; plastic deformation; stiffness; and fracture-line displacement.
RESULTS: Statistical analysis of all measured parameters revealed no significant differences among the groups. However, vertical displacement of the preshaped connecting rod within the screw heads occurred as a result of inadequate stability of the internal fixation at the rod-screw interface.
CONCLUSIONS: Although internal fixator devices are placed close to the bone and should therefore maintain greater stiffness, our data did not support the hypothesis of superior stability. Special attention is required when using a curved connecting rod as the rod is easily displaced, placing relevant anatomical structures at risk. These devices require further refinement to avoid potential patient injury.

Entities:  

Mesh:

Year:  2019        PMID: 31577677     DOI: 10.2106/JBJS.18.01363

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  4 in total

1.  Biomechanical comparison of three minimally invasive fixations for unilateral pubic rami fractures.

Authors:  Yong Zhao; Yupeng Ma; Dexin Zou; Xiujiang Sun; Gong Cheng; Wei Lian; Shengjie Dong; Yuchi Zhao; Wenqing Qu; Hao Wu
Journal:  BMC Musculoskelet Disord       Date:  2020-09-04       Impact factor: 2.362

2.  Biomechanical evaluation of seven fixation methods to treat pubic symphysis diastasis using finite element analysis.

Authors:  Yi-Quan Zheng; Li-Li Chen; Jia-Zuo Shen; Bing Gao; Xiao-Chuan Huang
Journal:  J Orthop Surg Res       Date:  2022-03-28       Impact factor: 2.359

3.  Trans-obturator cable fixation of open book pelvic injuries.

Authors:  Martin C Jordan; Veronika Jäckle; Sebastian Scheidt; Fabian Gilbert; Stefanie Hölscher-Doht; Süleyman Ergün; Rainer H Meffert; Timo M Heintel
Journal:  Sci Rep       Date:  2021-06-29       Impact factor: 4.379

4.  The role of anterior supra-acetabular external fixator as definitive treatment for anterior ring fixation in unstable pelvic fractures.

Authors:  Cristián Barrientos-Mendoza; Julián Brañes; Rodrigo Wulf; Alex Kremer; Maximiliano Barahona; Sebastián León
Journal:  Eur J Trauma Emerg Surg       Date:  2021-06-07       Impact factor: 2.374