Greta Castellini1, Silvia Gianola1, Elena Stucovitz2, Irene Tramacere3, Giuseppe Banfi4,5, Lorenzo Moja1,6.
Abstract
We aimed to determine the accuracy and failure of OAK device, an automated screening, for the assessment of fall risk in a prospective cohort of healthy adults aged over 65 years. The algorithm for fall risk assessment of the centers for disease control and prevention (CDC) was used as reference standard. Of the 183 individuals recruited, the CDC algorithm classified 80 as being at moderate/high risk and 103 at low risk of falling. OAK device failure incidence was 4.9% (confidence interval [CI] upper limit 7.7%), below the preset threshold for futility-early termination of the study (i.e., not above 15%). The OAK device showed a sensitivity of 84% and a specificity of 67% (receiver operating characteristic [ROC] area 82%; 95% confidence interval [CI] 76-88%), not reaching the preplanned target sensitivity (not lower than 85%). Diagnostic accuracy was not far from the sensitivity levels similar to those obtained with other fall risk assessment. However, some limitations can be considered.ClinicalTrials.gov identifier: NCT02655796.Entities:
Mesh:
Year: 2019 PMID: 31574809 PMCID: PMC6775395 DOI: 10.1097/MD.0000000000017105
Source DB: PubMed Journal: Medicine (Baltimore) ISSN: 0025-7974 Impact factor: 1.817
Figure 1The OAK device. The OAK device includes stabilometric platforms (outlined in red), 4 antennas that generate the low intensity magnetic field where the subject exercises (outlined in yellow), 3 bars that detect the subject's body weight during exercise (outlined in green), and a VR monitor that displays the exercise.
Figure 2STARD flow diagram.
Participant characteristics.
Figure 3ROC analysis of fall-risk assessment determined with the OAK device. ROC = receiver operating characteristic analysis.