| Literature DB >> 31571822 |
Devin DeLuna1, Audrey Netzel2, Jamie Dietze1, Brett A Begley1, Jideofor K Ndulue1, Donny W Suh1.
Abstract
PURPOSE: To compare the accuracy and precision of a conventional intraocular syringe to the Suh precision syringe (SPS), a new injection device designed to decrease operator error and to reduce adverse outcomes by providing a more ergonomic technique for periocular and intraocular injections.Entities:
Keywords: Suh precision syringe; intraocular injection
Year: 2019 PMID: 31571822 PMCID: PMC6759204 DOI: 10.2147/OPTH.S211485
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Figure 1(A) The Suh precision syringe and (B) a conventional syringe.
Figure 2Demonstration of the ruler tool used to calculate the forward-retraction movement of needles.
Comparison between conventional and Suh precision syringe. Comparisons were made using Wilcoxon signed-rank test
| Variables | Conventional Syringe | Suh Precision Syringe | |
|---|---|---|---|
| Median (Q1; Q3) | Median (Q1; Q3) | ||
| Forward-retraction movementa (mm) | 2.85 (2.30; 3.10) | 2.35 (2.10; 3.10) | 0.09 |
| Variability in forward-retraction movementa between 1st and 2nd attempt | 0.60 (0.30; 1.50) | 0.50 (−1.00; 1.00) | 0.68 |
| Forward-retraction movementb (mm) | 4.46 (3.35; 4.62) | 3.53 (3.23; 4.42) | 0.04* |
| Water expelled from syringe (mL) | 0.05 (0.04; 0.05) | 0.05 (0.05; 0.05) | 0.28 |
Notes: Q1=25th percentile; Q3=75th percentile. aInjection into a 0.2 mL tube. bInjection into sheep eye. *Indicates statistical significance with P<0.05.
Figure 3Box plot comparing forward-retraction movement of a conventional syringe and the SPS during injection of water into a 0.2 mL PCR tube.
Figure 4Box plot comparing forward-retraction movement between a conventional syringe and the SPS while injecting water into a sheep eye.
Injections into a 0.2 mL PCR tube by a practicing pediatric ophthalmologist
| Variables | Conventional Syringe | Suh Precision Syringe |
|---|---|---|
| Average forward-retraction movement (mm) | 2.5 | 1.85 |
| Average water expelled from syringe (mL) | 0.031 | 0.052 |