Impedance and conductivity of bovine myocardium during freezing and thawing at slow rates - implications for cardiac cryo-ablation.

G Fischer1, M Handler2, P R Johnston3, D Baumgarten4.   

Abstract

Increasing impedance during freezing might be a valuable marker for guiding cardiac cryo-ablation. We provide model based insights on how decreasing temperature below the freezing point of tissue relates to the percentage of frozen water. Furthermore, we provide experimental data for comparing this percentage with the increase in impedance. Measurements were performed on a bovine tissue sample at frequencies between 5 and 80 kHz. Slow cooling and heating rates were applied to minimize temperature gradients in the myocardial sample and to allow accurate assessment of the freezing point. Computer simulation was applied to link impedance with temperature dependent conductivities. The osmotic virial equation was used to estimate the percentage of frozen water. Measurements identified the freezing point at -0.6 ∘C. At -5 ∘C, impedance rose by more than a factor of ten compared to that at the freezing point and the percentage of frozen water was estimated as being 89%. At -49 ∘C impedance had increased by up to three orders of magnitude and ice formation was most pronounced in the extracellular space. Progressive ice formation in tissue is reflected by a large increase in impedance, and impedance increases monotonically with the percentage of frozen water. Its analysis allows for experimental assessment of factors relevant to cell death. Solid ice contributes to the rupture of the micro-vasculature, while phase shifts reflect concentration differences between extra- and intracellular space driving osmotic water transfer across cell membranes.
Copyright © 2019 IPEM. All rights reserved.

Entities:  

Keywords:  Cryo-ablation; Electrode polarization; Osmotic virial equation; Tissue impedance

Mesh:

Year:  2019        PMID: 31570217     DOI: 10.1016/j.medengphy.2019.09.017

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  2 in total

1.  A Lead Field Two-Domain Model for Longitudinal Neural Tracts-Analytical Framework and Implications for Signal Bandwidth.

Authors:  G Fischer; M Kofler; M Handler; D Baumgarten
Journal:  Comput Math Methods Med       Date:  2020-05-29       Impact factor: 2.238

2.  Antral lesion characterization of a new cryoballoon ablation system in terms of local impedance drop: The first reported case.

Authors:  Saverio Iacopino; Pasquale Filannino; Paolo Artale; Filippo Placentino; Francesca Pesce; Andrea Petretta
Journal:  HeartRhythm Case Rep       Date:  2020-12-29